Validity and reliability of bioelectrical impedance analysis to estimate body fat percentage against air displacement plethysmography and dual-energy X-ray absorptiometry

被引:63
作者
von Hurst, Pamela R. [1 ]
Walsh, Daniel C. I. [2 ]
Conlon, Cathryn A. [1 ]
Ingram, Michelle [1 ]
Kruger, Rozanne [1 ]
Stonehouse, Welma [1 ,3 ]
机构
[1] Massey Univ, Coll Hlth, Human Nutr & Dietet, Auckland 0745, New Zealand
[2] Massey Univ, Inst Nat & Math Sci, Auckland 0745, New Zealand
[3] CSIRO Anim Food & Hlth Sci, Adelaide, SA, Australia
关键词
bioelectrical impedance; BodPod; body composition; body fat percentage; DXA; obesity; VALIDATION; CHILDREN; DEXA; MEN;
D O I
10.1111/1747-0080.12172
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Air displacement plethysmography (ADP) and dual-energy X-ray absorptiometry (DXA) are well-regarded methods for predicting body fat percentage (BF%). Bioelectrical impedance analysis (BIA) also predicts BF% and has distinct advantages in research settings. Aim: To assess the validity of BIA against ADP and DXA to measure BF%, and to test the reliability of each method. Methods: Adults (n = 166) with a wide range of body mass index (19-38 kg/m(2)) were tested twice during a 5-day period. ADP was conducted in a BodPod (Life Measurement Inc, Concord, CA, USA); DXA measurements on a QDR Discovery A (Hologic) and BIA measurements used the InBody 230 (Biospace Ltd., Seoul, Korea). Agreement between measurements was analysed using t-tests, effect size, linear regression models and method of triads (estimating true value). Results: BIA showed excellent relative agreement to the estimated true value (rho = 0.97 (0.96, 0.98)) and to ADP (R-2 = 0.88) and DXA (R-2 = 0.92), but wide limits of agreement (-4.25 to 8.37%). BIA underestimated BF% by 2%, across all values. DXA showed excellent relative agreement to the estimated true value (rho = 0.97 (0.96, 0.98)) and with ADP (R-2 = 0.92), good absolute agreement but wide limits of agreement (-6.13 to 6.91%) and under- and overestimation at high and low BF% levels, respectively. All methods showed excellent reliability with repeat measurements differing by less than 0.2% with very small 95% CIs. Conclusions: BIA may be a valid method in research and population samples. All three methods showed excellent reliability.
引用
收藏
页码:197 / 204
页数:8
相关论文
共 50 条
  • [41] Air Displacement Plethysmography, Dual-Energy X-ray Absorptiometry, and Total Body Water to Evaluate Body Composition in Preschool-Age Children
    Crook, Tina A.
    Armbya, Narain
    Cleves, Mario A.
    Badger, Thomas M.
    Andres, Aline
    JOURNAL OF THE ACADEMY OF NUTRITION AND DIETETICS, 2012, 112 (12) : 1993 - 1998
  • [42] Body Composition in Heavy Smokers: Comparison of Segmental Bioelectrical Impedance Analysis and Dual-Energy X-Ray Absorptiometry
    Rom, O.
    Reznick, A. Z.
    Keidar, Z.
    Karkabi, K.
    Aizenbud, D.
    BODY METABOLISM AND EXERCISE, 2015, 840 : 1 - 11
  • [43] A comparison of dual energy X-ray absorptiometry and two bioelectrical impedance analyzers to measure body fat percentage and fat-free mass index in a group of Mexican young women
    del Consuelo Velazquez-Alva, Maria
    Irigoyen-Camacho, Maria Esther
    Huerta-Huerta, Raquel
    Delgadillo-Velazquez, Jaime
    NUTRICION HOSPITALARIA, 2014, 29 (05) : 1038 - 1046
  • [44] Underestimation of percentage fat mass measured by bioelectrical impedance analysis compared to dual energy X-ray absorptiometry method in obese children
    Eisenkölbl, J
    Kartasurya, M
    Widhalm, K
    EUROPEAN JOURNAL OF CLINICAL NUTRITION, 2001, 55 (06) : 423 - 429
  • [45] Measurement of percentage of body fat in 411 children and adolescents: A comparison of dual-energy X-ray absorptiometry with a four-compartment model
    Sopher, AB
    Thornton, JC
    Wang, J
    Pierson, RN
    Heymsfield, SB
    Horlick, M
    PEDIATRICS, 2004, 113 (05) : 1285 - 1290
  • [46] Underestimation of percentage fat mass measured by bioelectrical impedance analysis compared to dual energy X-ray absorptiometry method in obese children
    J Eisenkölbl
    M Kartasurya
    K Widhalm
    European Journal of Clinical Nutrition, 2001, 55 : 423 - 429
  • [47] A comparison of dual-energy X-ray absorptiometry, air displacement plethysmography and A-mode ultrasound toassess body composition in college-age adults
    Johnson, Kelly E.
    Miller, Brian
    Gibson, Ann L.
    McLain, Trisha A.
    Juvancic-Heltzel, Judith A.
    Kappler, Rachele M.
    Otterstetter, Ronald
    CLINICAL PHYSIOLOGY AND FUNCTIONAL IMAGING, 2017, 37 (06) : 646 - 654
  • [48] A Comparative Study of High-Frequency Bioelectrical Impedance Analysis and Dual-Energy X-ray Absorptiometry for Estimating Body Composition
    Yi, Youngseok
    Baek, Ji Yeon
    Lee, Eunju
    Jung, Hee-Won
    Jang, Il-Young
    LIFE-BASEL, 2022, 12 (07):
  • [49] Comparison of Body Mass Index, Skinfold Thickness, and Bioelectrical Impedance Analysis With Dual-Energy X-Ray Absorptiometry in Hemodialysis Patients
    de Abreu, Aline Miroski
    Wilvert, Luana Cristina
    Wazlawik, Elisabeth
    NUTRITION IN CLINICAL PRACTICE, 2020, 35 (06) : 1021 - 1028
  • [50] Practical but Inaccurate? A-Mode Ultrasound and Bioelectrical Impedance Underestimate Body Fat Percentage Compared to Dual-Energy X-ray Absorptiometry in Male College Students
    Olinto, Markus
    Lins, Victor Cesar
    Rocha, Gabriel
    Dourado, Marco Aurelio
    Dutra, Maurilio
    JOURNAL OF FUNCTIONAL MORPHOLOGY AND KINESIOLOGY, 2024, 9 (03)