Calibration transfer of near infrared spectrometers for the assessment of plasma ethanol precipitation process

被引:10
作者
Sun, Zhongyu [1 ,2 ]
Wang, Jiayue [1 ,2 ]
Nie, Lei [3 ]
Li, Lian [4 ]
Cao, Dawei [5 ]
Fan, Jiajin [5 ]
Wang, Haiyan [1 ,2 ]
Liu, Ruichen [1 ,2 ]
Zhang, Yinran [6 ]
Zang, Hengchang [1 ,2 ]
机构
[1] Shandong Univ, Sch Pharmaceut Sci, Wenhuaxi Rd 44, Jinan 250012, Shandong, Peoples R China
[2] Shandong Univ, Natl Glycoengn Res Ctr, Wenhuaxi Rd 44, Jinan 250012, Shandong, Peoples R China
[3] Shandong Univ, Sch Pharmaceut Sci & Pharmaceut Anal, Wenhuaxi Rd 44, Jinan 250012, Shandong, Peoples R China
[4] Shandong Univ, Sch Basic Med Sci, Wenhuaxi Rd 44, Jinan 250012, Shandong, Peoples R China
[5] Shandong Taibang Biol Prod Ltd Co, East Hushan Rd 14, Tai An 271000, Shandong, Peoples R China
[6] Ohio State Univ, Dept Chem & Biochem, Newman & Wolfram Lab, 100 West 18th Ave, Columbus, OH 43210 USA
基金
美国国家科学基金会;
关键词
Calibration transfer; Plasma ethanol precipitation; Near infrared spectroscopy (NIRS); Piecewise reverse standardization (PRS); Rank sample set partitioning based on joint x-y distance (Rank-SPXY); Process analytical technology (PAT); MULTIVARIATE CALIBRATION; CORRELATION-COEFFICIENT; ALCOHOL PRECIPITATION; FOURIER-TRANSFORM; NIR SPECTRA; MICRO-NIR; STANDARDIZATION; QUALITY; SPECTROSCOPY; INSTRUMENTS;
D O I
10.1016/j.chemolab.2018.08.012
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Plasma ethanol precipitation process is an important step in the biopharmacy production process, and near infrared spectroscopy (NIBS) is an effective process analytical technology (PAT) tool in detecting the key quality attribute: the total protein (TP) content. But establishing a robust model is pretty complex. The aim of this study was to explore the calibration transfer strategy between master and slave spectrometers with the biopharmacy. Firstly, 8 batches of ethanol precipitation process were experimented. The changing spectra were collected by two spectrometers, reference values were measured by biuret method. Then, calibration transfer based on the PRS algorithm was carried out. Finally, before establishment of the optimal TP content partial least squares regression (PLSR) model of calibration transfer, rank sample set partitioning based on joint x-y distance (Rank-SPXY) and spectral pretreatment methods were selected after comparison. The determination coefficients of prediction (R-p(2)) and root mean squares error of prediction (RMSEP) were 0.934 and 0.8395 g/L, respectively. The results showed that Rank-SPXY-PRS was an effective calibration transfer strategy with different NIBS spectrometers. To some extent, the predictive capability of PLSR model was improved after calibration transfer, and it could provide reference for the further research in biopharmacy production process.
引用
收藏
页码:64 / 71
页数:8
相关论文
共 59 条
  • [11] Portable near-infrared instruments: Application for quality control of polymorphs in pharmaceutical raw materials and calibration transfer
    da Silva, Vitor Hugo
    da Silva, Janson Jose
    Pereira, Claudete Fernandes
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2017, 134 : 287 - 294
  • [12] Raman spectroscopy as a process analytical technology (PAT) tool for the in-line monitoring and understanding of a powder blending process
    De Beer, T. R. M.
    Bodson, C.
    Dejaegher, B.
    Walczak, B.
    Vercruysse, P.
    Burggraeve, A.
    Lemos, A.
    Delattre, L.
    Heyden, Y. Vander
    Remon, J. P.
    Vervaete, C.
    Baeyens, W. R. G.
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2008, 48 (03) : 772 - 779
  • [13] THE PLASMA PROTEINS AND THEIR FRACTIONATION
    EDSALL, JT
    [J]. ERGEBNISSE DER PHYSIOLOGIE BIOLOGISCHEN CHEMIE UND EXPERIMENTELLEN PHARMAKOLOGIE, 1950, 46 : 308 - 353
  • [14] Determination of terpenoid contents of aromatic plants using NIRS
    Ercioglu, Elif
    Velioglu, Hasan Murat
    Boyaci, Ismail Hakki
    [J]. TALANTA, 2018, 178 : 716 - 721
  • [15] Qualitative near infrared spectroscopy, a new tool to recognize past vegetation signature in soil organic matter
    Ertlen, Damien
    Schwartz, Dominique
    Brunet, Didier
    Trendel, Jean-Michel
    Adam, Pierre
    Schaeffer, Philippe
    [J]. SOIL BIOLOGY & BIOCHEMISTRY, 2015, 82 : 127 - 134
  • [16] The dynamics of contract plasma fractionation
    Farrugia, Albert
    Scaramuccia, Daniela
    [J]. BIOLOGICALS, 2017, 46 : 159 - 167
  • [17] Feng B., 2014, ANAL INSTRUM, V3
  • [18] Folch-Fortuny A., 2017, J CHEMOMETR, V31
  • [19] A method for calibration and validation subset partitioning
    Galvao, RKH
    Araujo, MCU
    José, GE
    Pontes, MJC
    Silva, EC
    Saldanha, TCB
    [J]. TALANTA, 2005, 67 (04) : 736 - 740