Calibration transfer of near infrared spectrometers for the assessment of plasma ethanol precipitation process

被引:10
作者
Sun, Zhongyu [1 ,2 ]
Wang, Jiayue [1 ,2 ]
Nie, Lei [3 ]
Li, Lian [4 ]
Cao, Dawei [5 ]
Fan, Jiajin [5 ]
Wang, Haiyan [1 ,2 ]
Liu, Ruichen [1 ,2 ]
Zhang, Yinran [6 ]
Zang, Hengchang [1 ,2 ]
机构
[1] Shandong Univ, Sch Pharmaceut Sci, Wenhuaxi Rd 44, Jinan 250012, Shandong, Peoples R China
[2] Shandong Univ, Natl Glycoengn Res Ctr, Wenhuaxi Rd 44, Jinan 250012, Shandong, Peoples R China
[3] Shandong Univ, Sch Pharmaceut Sci & Pharmaceut Anal, Wenhuaxi Rd 44, Jinan 250012, Shandong, Peoples R China
[4] Shandong Univ, Sch Basic Med Sci, Wenhuaxi Rd 44, Jinan 250012, Shandong, Peoples R China
[5] Shandong Taibang Biol Prod Ltd Co, East Hushan Rd 14, Tai An 271000, Shandong, Peoples R China
[6] Ohio State Univ, Dept Chem & Biochem, Newman & Wolfram Lab, 100 West 18th Ave, Columbus, OH 43210 USA
基金
美国国家科学基金会;
关键词
Calibration transfer; Plasma ethanol precipitation; Near infrared spectroscopy (NIRS); Piecewise reverse standardization (PRS); Rank sample set partitioning based on joint x-y distance (Rank-SPXY); Process analytical technology (PAT); MULTIVARIATE CALIBRATION; CORRELATION-COEFFICIENT; ALCOHOL PRECIPITATION; FOURIER-TRANSFORM; NIR SPECTRA; MICRO-NIR; STANDARDIZATION; QUALITY; SPECTROSCOPY; INSTRUMENTS;
D O I
10.1016/j.chemolab.2018.08.012
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Plasma ethanol precipitation process is an important step in the biopharmacy production process, and near infrared spectroscopy (NIBS) is an effective process analytical technology (PAT) tool in detecting the key quality attribute: the total protein (TP) content. But establishing a robust model is pretty complex. The aim of this study was to explore the calibration transfer strategy between master and slave spectrometers with the biopharmacy. Firstly, 8 batches of ethanol precipitation process were experimented. The changing spectra were collected by two spectrometers, reference values were measured by biuret method. Then, calibration transfer based on the PRS algorithm was carried out. Finally, before establishment of the optimal TP content partial least squares regression (PLSR) model of calibration transfer, rank sample set partitioning based on joint x-y distance (Rank-SPXY) and spectral pretreatment methods were selected after comparison. The determination coefficients of prediction (R-p(2)) and root mean squares error of prediction (RMSEP) were 0.934 and 0.8395 g/L, respectively. The results showed that Rank-SPXY-PRS was an effective calibration transfer strategy with different NIBS spectrometers. To some extent, the predictive capability of PLSR model was improved after calibration transfer, and it could provide reference for the further research in biopharmacy production process.
引用
收藏
页码:64 / 71
页数:8
相关论文
共 59 条
  • [1] BILLER C, 2014, NIR NEWS, V25, P16, DOI DOI 10.1255/NIRN.1471
  • [2] Brito ALB, 2017, ANAL METHODS-UK, V9, P3184, DOI [10.1039/C7AY00391A, 10.1039/c7ay00391a]
  • [3] Standardisation of near-infrared spectrometric instruments: A review
    Bouveresse, E
    Massart, DL
    [J]. VIBRATIONAL SPECTROSCOPY, 1996, 11 (01) : 3 - 15
  • [4] CALIBRATION TRANSFER ACROSS NEAR-INFRARED SPECTROMETRIC INSTRUMENTS USING SHENKS ALGORITHM - EFFECTS OF DIFFERENT STANDARDIZATION SAMPLES
    BOUVERESSE, E
    MASSART, DL
    DARDENNE, P
    [J]. ANALYTICA CHIMICA ACTA, 1994, 297 (03) : 405 - 416
  • [5] Issues in development of NIR micro spectrometer system for on-line process monitoring of milk product
    Brennan, D
    Alderman, J
    Sattler, L
    O'Connor, B
    O'Mathuna, C
    [J]. MEASUREMENT, 2003, 33 (01) : 67 - 74
  • [6] Calibration transfer of a Raman spectroscopic quantification method from at-line to in-line assessment of liquid detergent compositions
    Brouckaert, D.
    Uyttersprot, J. -S.
    Broeckx, W.
    De Beer, T.
    [J]. ANALYTICA CHIMICA ACTA, 2017, 971 : 14 - 25
  • [7] Chen YiYun Chen YiYun, 2017, Transactions of the Chinese Society of Agricultural Engineering, V33, P107
  • [8] Cho S, 2005, B KOREAN CHEM SOC, V26, P115
  • [9] Clermont TR, 2001, SPECTROSCOPY, V16, P28
  • [10] Standardization from a benchtop to a handheld NIR spectrometer using mathematically mixed NIR spectra to determine fuel quality parameters
    da Silva, Neirivaldo Cavalcante
    Cavalcanti, Claudia Jessica
    Honorato, Fernanda Araujo
    Amigo, Jose Manuel
    Pimentel, Maria Fernanda
    [J]. ANALYTICA CHIMICA ACTA, 2017, 954 : 32 - 42