Microwave and digital imaging technology reduce turnaround times for diagnostic electron microscopy

被引:35
作者
Giberson, RT
Austin, RL
Charlesworth, J
Adamson, G
Herrera, GA
机构
[1] Ted Pella Inc, Redding, CA 96003 USA
[2] Louisiana State Univ, Hlth Sci Ctr, Shreveport, LA 71105 USA
[3] Mayo Clin, EM Core Facil, Rochester, MN USA
[4] Univ Calif Davis, Sch Med, Davis, CA 95616 USA
关键词
diagnostic; digital imaging; electron microscopy; microwave; turnaround time;
D O I
10.1080/01913120309937
中图分类号
TH742 [显微镜];
学科分类号
摘要
The contributions of microwave methods and digital imaging techniques, when taken together, can reduce routine specimen processing and evaluation for diagnostic electron microscopy to a time frame never thought possible. Significant improvements in both technologies over the last 5 years led the authors to evaluate their combined attributes as the most likely candidate to provide a realistic solution in the reduction of turnaround times for diagnostic electron microscopy. For diagnostic electron microscopy to compete favorably with immunohistochemistry and other ancillary diagnostic techniques, it must improve its turnaround time. To evaluate this hypothesis the microwave-assisted processing results of over 2,000 diagnostic cases were evaluated as was a digital image administration system used for the acquisition and dissemination of diagnostic results. The incorporation of both technologies resulted in turnaround times being reduced to 4 h or less.
引用
收藏
页码:187 / 196
页数:10
相关论文
共 16 条
[1]  
Austin RL, 2001, MICROWAVE TECHNIQUES AND PROTOCOLS, P37, DOI 10.1007/978-1-59259-128-2_4
[3]  
CARSON FL, 1973, AM J CLIN PATHOL, V59, P365
[4]  
DARDICK I, 1994, SOC ULTRASTRUCT PATH, V7, P7
[5]  
EM D, 2002, STATE ART DESKTOP VI
[6]   FORMALDEHYDE FIXATION [J].
FOX, CH ;
JOHNSON, FB ;
WHITING, J ;
ROLLER, PP .
JOURNAL OF HISTOCHEMISTRY & CYTOCHEMISTRY, 1985, 33 (08) :845-853
[7]   Four-hour processing of clinical/diagnostic specimens for electron microscopy using microwave technique [J].
Giberson, RT ;
Demaree, RS ;
Nordhausen, RW .
JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 1997, 9 (01) :61-67
[8]  
Giberson RT, 2001, MICROWAVE TECHNIQUES AND PROTOCOLS, P191, DOI 10.1007/978-1-59259-128-2_16
[9]  
Giberson RT, 2001, MICROWAVE TECHNIQUES AND PROTOCOLS, P13, DOI 10.1007/978-1-59259-128-2_2
[10]  
GIBERSON RT, 2001, MICROSC MICROANAL S2, V7, P1192