Electromagnetic Design and Performance of a Conical Microwave Blackbody Target for Radiometer Calibration

被引:11
|
作者
Houtz, Derek A. [1 ,2 ]
Emery, William [2 ]
Gu, Dazhen [1 ]
Jacob, Karl [3 ]
Murk, Axel [3 ]
Walker, David K. [1 ]
Wylde, Richard J. [4 ]
机构
[1] NIST, RF Technol Div, Boulder, CO 80305 USA
[2] Univ Colorado, Dept Aerosp Engn & Sci, Boulder, CO 80309 USA
[3] Univ Bern, Inst Appl Phys, CH-3012 Bern, Switzerland
[4] Thomas Keating Ltd, Billinghurst RH14 9SH, England
来源
基金
瑞士国家科学基金会;
关键词
Blackbody; brightness-temperature standard; calibration source; calibration target; cold load; hot load; microwave absorber; permeability; permittivity; radiometer calibration; PERMITTIVITY; STANDARD;
D O I
10.1109/TGRS.2017.2694319
中图分类号
P3 [地球物理学]; P59 [地球化学];
学科分类号
0708 ; 070902 ;
摘要
A conical cavity has been designed and fabricated for use as a broadband passive microwave calibration source, or blackbody, at the National Institute of Standards and Technology. The blackbody will be used as a national primary standard for brightness temperature and will allow for the prelaunch calibration of spaceborne radiometers and calibration of ground-based systems to provide traceability among radiometric data. The conical geometry provides performance independent of polarization, minimizing reflections, and standing waves, thus having a high microwave emissivity. The conical blackbody has advantages over typical pyramidal array geometries, including reduced temperature gradients and excellent broadband electromagnetic performance over more than a frequency decade. The blackbody is designed for use between 18 and 230 GHz, at temperatures between 80 and 350 K, and is vacuum compatible. To approximate theoretical blackbody behavior, the design maximizes emissivity and thus minimizes reflectivity. A newly developed microwave absorber is demonstrated that uses cryogenically compatible, thermally conductive two-part epoxy with magnetic carbonyl iron (CBI) powder loading. We measured the complex permittivity and permeability properties for different CBI-loading percentages; the conical absorber is then designed and optimized with geometric optics and finite-element modeling, and finally, the reflectivity of the resulting fabricated structure is measured. We demonstrated normal incidence reflectivity considerably below -40 dB at all relevant remote sensing frequencies.
引用
收藏
页码:4586 / 4596
页数:11
相关论文
共 50 条
  • [41] SIMULATIONS TO CHARACTERIZE A PASSIVE MICROWAVE BLACKBODY DESIGN
    Houtz, Derek A.
    Walker, David K.
    Gu, Dazhen
    2015 IEEE INTERNATIONAL GEOSCIENCE AND REMOTE SENSING SYMPOSIUM (IGARSS), 2015, : 3485 - 3488
  • [42] Ground-Based Microwave Radiometer Calibration: an Overview
    Miacci, Marcelo
    Angelis, Carlos Frederico
    JOURNAL OF AEROSPACE TECHNOLOGY AND MANAGEMENT, 2018, 10
  • [43] Characterization and correction of a drift in calibration of the TOPEX microwave radiometer
    Ruf, CS
    IEEE TRANSACTIONS ON GEOSCIENCE AND REMOTE SENSING, 2002, 40 (02): : 509 - 511
  • [44] Calibration procedure of a microwave total-power radiometer
    Al-Ansari, K
    Garcia, P
    Riera, JM
    Benarroch, A
    Fernandez, D
    Fernandez, L
    IEEE MICROWAVE AND WIRELESS COMPONENTS LETTERS, 2002, 12 (03) : 93 - 95
  • [45] SMAP MICROWAVE RADIOMETER CALIBRATION REVISIT APPROACHES AND PERFORMAMNCE
    Peng, Jinzheng
    Misra, Sidharth
    Piepmeier, Jeffrey
    Yueh, Simon
    Mohammed, Priscilla
    Dinnat, Emmanuel
    Le Vine, David
    Meissner, Thomas
    IGARSS 2020 - 2020 IEEE INTERNATIONAL GEOSCIENCE AND REMOTE SENSING SYMPOSIUM, 2020, : 3330 - 3333
  • [46] BIRMM Fully Polarimetric Calibration Standard for Microwave Radiometer
    Chen Ling
    Cheng Junjie
    Sun Xiaoning
    Zhai Hong
    Xing Xiaoyun
    Cheng Chunyue
    2019 INTERNATIONAL CONFERENCE ON MICROWAVE AND MILLIMETER WAVE TECHNOLOGY (ICMMT 2019), 2019,
  • [47] A New Linearly Polarized Microwave Radiometer Calibration Source
    Wang, Xinbiao
    Zeng, Chang
    Li, Jing
    Jiang, Jingshan
    ADVANCED MATERIALS IN MICROWAVES AND OPTICS, 2012, 500 : 341 - 347
  • [48] In-flight calibration/validation of ENVISAT microwave radiometer
    Tran, N
    Obligis, E
    Eymard, L
    Féménias, P
    IGARSS 2003: IEEE INTERNATIONAL GEOSCIENCE AND REMOTE SENSING SYMPOSIUM, VOLS I - VII, PROCEEDINGS: LEARNING FROM EARTH'S SHAPES AND SIZES, 2003, : 2993 - 2995
  • [49] The development and calibration of the ENVISAT-1 microwave radiometer
    Guijarro, J
    Graf, G
    IGARSS '98 - 1998 INTERNATIONAL GEOSCIENCE AND REMOTE SENSING SYMPOSIUM, PROCEEDINGS VOLS 1-5: SENSING AND MANAGING THE ENVIRONMENT, 1998, : 262 - 264
  • [50] Ground calibration of multifrequency scanning microwave radiometer (MSMR)
    Misra, T
    Jha, AM
    Putrevu, D
    Rao, J
    Dave, DB
    Rana, SS
    IEEE TRANSACTIONS ON GEOSCIENCE AND REMOTE SENSING, 2002, 40 (02): : 504 - 508