Electromagnetic Design and Performance of a Conical Microwave Blackbody Target for Radiometer Calibration

被引:11
|
作者
Houtz, Derek A. [1 ,2 ]
Emery, William [2 ]
Gu, Dazhen [1 ]
Jacob, Karl [3 ]
Murk, Axel [3 ]
Walker, David K. [1 ]
Wylde, Richard J. [4 ]
机构
[1] NIST, RF Technol Div, Boulder, CO 80305 USA
[2] Univ Colorado, Dept Aerosp Engn & Sci, Boulder, CO 80309 USA
[3] Univ Bern, Inst Appl Phys, CH-3012 Bern, Switzerland
[4] Thomas Keating Ltd, Billinghurst RH14 9SH, England
来源
基金
瑞士国家科学基金会;
关键词
Blackbody; brightness-temperature standard; calibration source; calibration target; cold load; hot load; microwave absorber; permeability; permittivity; radiometer calibration; PERMITTIVITY; STANDARD;
D O I
10.1109/TGRS.2017.2694319
中图分类号
P3 [地球物理学]; P59 [地球化学];
学科分类号
0708 ; 070902 ;
摘要
A conical cavity has been designed and fabricated for use as a broadband passive microwave calibration source, or blackbody, at the National Institute of Standards and Technology. The blackbody will be used as a national primary standard for brightness temperature and will allow for the prelaunch calibration of spaceborne radiometers and calibration of ground-based systems to provide traceability among radiometric data. The conical geometry provides performance independent of polarization, minimizing reflections, and standing waves, thus having a high microwave emissivity. The conical blackbody has advantages over typical pyramidal array geometries, including reduced temperature gradients and excellent broadband electromagnetic performance over more than a frequency decade. The blackbody is designed for use between 18 and 230 GHz, at temperatures between 80 and 350 K, and is vacuum compatible. To approximate theoretical blackbody behavior, the design maximizes emissivity and thus minimizes reflectivity. A newly developed microwave absorber is demonstrated that uses cryogenically compatible, thermally conductive two-part epoxy with magnetic carbonyl iron (CBI) powder loading. We measured the complex permittivity and permeability properties for different CBI-loading percentages; the conical absorber is then designed and optimized with geometric optics and finite-element modeling, and finally, the reflectivity of the resulting fabricated structure is measured. We demonstrated normal incidence reflectivity considerably below -40 dB at all relevant remote sensing frequencies.
引用
收藏
页码:4586 / 4596
页数:11
相关论文
共 50 条
  • [21] RADIOMETRIC CHARACTERIZATION OF A WATER-BASED CONICAL BLACKBODY CALIBRATION TARGET FOR MILLIMETRE-WAVE REMOTE SENSING
    Jacob, K.
    Schroder, A.
    Murk, A.
    von Werra, L.
    Reinhard, F.
    Raisin, P.
    2018 IEEE 15TH SPECIALIST MEETING ON MICROWAVE RADIOMETRY AND REMOTE SENSING OF THE ENVIRONMENT (MICRORAD), 2018, : 143 - 146
  • [22] An incidence angle correction for a conical scanning microwave radiometer
    Kabore, JY
    Jones, WL
    Hood, R
    James, M
    LaFontaine, F
    Proceedings of the IEEE SoutheastCon 2004: EXCELLENCE IN ENGINEERING, SCIENCE, AND TECHNOLOGY, 2005, : 257 - 264
  • [23] Design and Implementation of a Ku-Band High-Precision Blackbody Calibration Target
    Liu, Jie
    Sun, Zhenlin
    Sun, Guangmin
    Li, Yu
    Cao, Tong
    Tang, Wenjie
    MICROMACHINES, 2023, 14 (01)
  • [24] Design, development and calibration of the MWR Microwave Radiometer on board Sentinel-3
    Palacios Lazaro, Miguel Angel
    Bergada Pujades, Marc
    Gonzalez Sola, Raquel
    Gamonal Coto, Ataulfo
    Garcia Fernandez, Jose Luis
    Camacho, Yolanda
    2014 11TH EUROPEAN RADAR CONFERENCE (EURAD), 2014, : 368 - 371
  • [25] Design, development and calibration of the MWR Microwave Radiometer on board Sentinel-3
    Palacios Lazaro, Miguel Angel
    Bergada Pujades, Marc
    Gonzalez Sola, Raquel
    Gamonal Coto, Ataulfo
    Garcia Fernandez, Jose Luis
    Camacho, Yolanda
    2014 44TH EUROPEAN MICROWAVE CONFERENCE (EUMC), 2014, : 1671 - 1674
  • [26] The Conical Microwave Imager/Sounder (CMIS): Next generation conical-scanning microwave radiometer for NPOESS
    Jewell, JJ
    Chauhan, NS
    Kunkee, DB
    SIXTH SYMPOSIUM ON INTEGRATED OBSERVING SYSTEMS, 2002, : 104 - 107
  • [27] Microwave Radiometer Instability Due to Infrequent Calibration
    Coakley, Kevin J.
    Splett, Jolene
    Walker, David
    Aksoy, Mustafa
    Racette, Paul
    IEEE JOURNAL OF SELECTED TOPICS IN APPLIED EARTH OBSERVATIONS AND REMOTE SENSING, 2020, 13 (13) : 3281 - 3290
  • [28] Calibration of Clock Scanning Microwave Interferometric Radiometer
    Zhang, Cheng
    Liu, Hao
    Wu, Ji
    2018 IEEE 15TH SPECIALIST MEETING ON MICROWAVE RADIOMETRY AND REMOTE SENSING OF THE ENVIRONMENT (MICRORAD), 2018, : 121 - 124
  • [29] Parametric optimization of microwave radiometer calibration load
    Hui, Zhang
    Plettemeier, Dirk
    Jungang, Miao
    Chun, Wu Bo
    Ming, Bai
    2008 ASIA-PACIFIC SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY AND 19TH INTERNATIONAL ZURICH SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY, VOLS 1 AND 2, 2008, : 839 - +
  • [30] Evaluation of disturbance in the antenna calibration of the microwave radiometer
    Li, Zhiping
    Peng, Yi
    Miao, Jungang
    Wang, Zhenrong
    2008 INTERNATIONAL CONFERENCE ON MICROWAVE AND MILLIMETER WAVE TECHNOLOGY PROCEEDINGS, VOLS 1-4, 2008, : 810 - 813