Evaluation of four airway training manikins as patient simulators for the insertion of eight types of supraglottic airway devices

被引:113
作者
Jackson, K. M. [1 ]
Cook, T. M. [1 ]
机构
[1] Royal United Hosp, Bath BA1 3NG, Avon, England
关键词
D O I
10.1111/j.1365-2044.2007.04983.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We evaluated the performance of four currently available manikins: Airway Management Trainer (TM) (Ambu, UK), Airway Trainer (TM) (Laerdal, Norway), Airsim (TM) (Trucorp, Ireland), 'Bill 1'(TM) (VBM, Germany), with eight supraglottic air-way devices: Airway Management Device (TM), Cobra Perilaryngeal Airway, Combitube (TM), i-Gel, Laryngeal Tube (TM), Laryngeal Tube Disposable, Laryngeal Tube Suction II and Streamlined Liner of the Pharynx Airway. Ten anaesthetists inserted each supraglottic airway device twice into each manikin. Each insertion was scored and ranked. Manikin score and rank data showed statistically significant overall performance differences. Post hoc analysis showed the Trucorp manikin performed best, followed by the Laerdal manikin. No one manikin performed best for all individual supraglottic airway devices. The Trucorp manikin performed adequately for all supraglottic airway devices. Comparing supraglottic airway devices, i-Gel insertion was significantly the easiest. Our results show that manikin performance for supraglottic airway device insertion is unequal, which has implications for selecting manikins for supraglottic airway device training and for manikin studies assessing performance of supraglottic airway devices.
引用
收藏
页码:388 / 393
页数:6
相关论文
共 4 条
  • [1] [Anonymous], 2005, RESUSCITATION, V67, pS39
  • [2] New airways for resuscitation?
    Cook, T. M.
    Hommers, C.
    [J]. RESUSCITATION, 2006, 69 (03) : 371 - 387
  • [3] Novel airway devices: Spoilt for choice?
    Cook, TM
    [J]. ANAESTHESIA, 2003, 58 (02) : 107 - 110
  • [4] Evaluation of four airway training manikins as simulators for inserting the LMA Classic™
    Silsby, J.
    Jordan, G.
    Bayley, G.
    Cook, T. M.
    [J]. ANAESTHESIA, 2006, 61 (06) : 576 - 579