Safety of optical coherence tomography in daily practice: a comparison with intravascular ultrasound

被引:66
作者
van der Sijde, Johannes N. [1 ]
Karanasos, Antonios [1 ]
van Ditzhuijzen, Nienke S. [1 ]
Okamura, Takayuki [2 ]
van Geuns, Robert-Jan [1 ]
Valgimigli, Marco [1 ]
Ligthart, Jurgen M. R. [1 ]
Witberg, Karen T. [1 ]
Wemelsfelder, Saskia [1 ]
Fam, Jiang Ming [1 ,3 ]
Zhang, BuChun [1 ,4 ]
Diletti, Roberto [1 ]
de Jaegere, Peter P. [1 ]
van Mieghem, Nicolas M. [1 ]
van Soest, Gijs [1 ]
Zijlstra, Felix [1 ]
van Domburg, Ron T. [1 ]
Regar, Evelyn [1 ]
机构
[1] Erasmus MC, Thoraxctr, Bd 585,S Gravendijkwal 230, NL-3015 CE Rotterdam, Netherlands
[2] Yamaguchi Univ, Dept Med & Clin Sci, Grad Sch Med, Yamaguchi, Japan
[3] Natl Heart Ctr Singapore, Singapore, Singapore
[4] Xuzhou Med Coll, Dept Cardiol, Affiliated Hosp, Xuzhou, Jiangsu, Peoples R China
关键词
Optical coherence tomography; Intravascular ultrasound; Safety; Adverse events; INTRACORONARY ULTRASOUND; FEASIBILITY; MULTICENTER; CATHETER; OCT;
D O I
10.1093/ehjci/jew037
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims Previous studies have reported the safety and feasibility of both time-domain optical coherence tomography (TDOCT) and Fourier-domain OCT (FD-OCT) in highly selected patients and clinical settings. However, the generalizability of these data is limited, and data in unselected patient populations reflecting a routine cathlab practice are lacking. We compared safety of intracoronary FD-OCT imaging to intravascular ultrasound (IVUS) imaging in a large real-world series of consecutive patients who underwent invasive imaging during coronary catheterization in our centre. Methods and results This is a prospective, single-centre registry of patients scheduled for coronary angiography or intervention undergoing intracoronary imaging with FD-OCT or IVUS between April 2008 and December 2013. Intra-procedural and major inhospital adverse events that could be possibly related to invasive imaging were registered routinely by the operator as part of our clinical report and prospectively recorded in our database. These events were retrospectively individually adjudicated by an independent safety committee. Between April 2008 and December 2013, 13 418 diagnostic or interventional coronary catheterization procedures were performed. Of these, 1142 procedures used OCT and 2476 procedures used IVUS. Invasive imaging-related complications were rare, did not differ between the two imaging methods (OCT: n = 7, 0.6%; IVUS: n = 12, 0.5%; P = 0.6), and were self-limiting after retrieval of the imaging catheter or easily treatable in the catheterization laboratory. No major adverse events, prolongation of hospital stay, or permanent patient harm was observed. Conclusion FD-OCT is safe in an unselected and heterogeneous group of patients with varying clinical settings.
引用
收藏
页码:467 / 474
页数:8
相关论文
共 19 条
[1]   A multicentre evaluation of the safety of intracoronary optical coherence tomography [J].
Barlis, Peter ;
Gonzalo, Nieves ;
Di Mario, Carlo ;
Prati, Francesco ;
Buellesfeld, Lutz ;
Rieber, Johannes ;
Dalby, Miles C. ;
Ferrante, Giuseppe ;
Cera, Maria ;
Grube, Eberhard ;
Serruys, Patrick W. ;
Regar, Evelyn .
EUROINTERVENTION, 2009, 5 (01) :90-95
[2]  
Batkoff BW, 1996, CATHETER CARDIO DIAG, V38, P238, DOI 10.1002/(SICI)1097-0304(199607)38:3<238::AID-CCD3>3.0.CO
[3]  
2-9
[4]  
Dobarro D, 2010, J INVASIVE CARDIOL, V22, P385
[5]   THE SAFETY OF INTRACORONARY ULTRASOUND - A MULTICENTER SURVEY OF 2207 EXAMINATIONS [J].
HAUSMANN, D ;
ERBEL, R ;
ALIBELLICHEMARIN, MJ ;
BOKSCH, W ;
CARACCIOLO, E ;
COHN, JM ;
CULP, SC ;
DANIEL, WG ;
DESCHEERDER, I ;
DIMARIO, C ;
FERGUSON, JJ ;
FITZGERALD, PJ ;
FRIEDRICH, G ;
GE, JB ;
GORGE, G ;
HANRATH, P ;
HODGSON, JM ;
ISNER, JM ;
JAIN, S ;
MAIERRUDOLPH, W ;
MOONEY, M ;
MOSES, JW ;
MUDRA, H ;
PINTO, FJ ;
SMALLING, RW ;
TALLEY, JD ;
TOBIS, JM ;
WALTER, PD ;
WEIDINGER, F ;
WERNER, GS ;
YEUNG, AC ;
YOCK, PG .
CIRCULATION, 1995, 91 (03) :623-630
[6]   Safety and feasibility of frequency domain optical coherence tomography to guide decision making in percutaneous coronary intervention [J].
Imola, Fabrizio ;
Mallus, Maria Teresa ;
Ramazzotti, Vito ;
Manzoli, Alessandro ;
Pappalardo, Alessandro ;
Di Giorgio, Alessandro ;
Albertucci, Mario ;
Prati, Francesco .
EUROINTERVENTION, 2010, 6 (05) :575-581
[7]   Optical Coherence Tomography: Potential Clinical Applications [J].
Antonios Karanasos ;
Jurgen Ligthart ;
Karen Witberg ;
Gijs van Soest ;
Nico Bruining ;
Evelyn Regar .
Current Cardiovascular Imaging Reports, 2012, 5 (4) :206-220
[8]   Periprocedural myocardial injury and right bundle branch block during coronary optical coherence tomography in an acute coronary syndrome patient with severe coronary ectasia [J].
Koyama, Kohei ;
Yoneyama, Kihei ;
Mitarai, Takanobu ;
Kobayashi, Yasuyuki ;
Saito, Mitsuo ;
Oono, Toshio ;
Kuwata, Shingo ;
Ishibashi, Yuki ;
Kongoji, Ken ;
Akashi, Yoshihiro J. ;
Harada, Tomoo .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2014, 177 (03) :1113-1115
[9]   Feasibility and safety of frequency-domain optical coherence tomography for coronary artery evaluation: a single-center study [J].
Lehtinen, Tuomas ;
Nammas, Wail ;
Airaksinen, Juhani K. E. ;
Karjalainen, Pasi P. .
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2013, 29 (05) :997-1005
[10]  
Prati Francesco, 2007, EuroIntervention, V3, P365, DOI 10.4244/EIJV3I3A66