A robotic needle-positioning and guidance system for CT-guided puncture: Ex vivo results

被引:45
作者
Kettenbach, Joachim [1 ]
Kara, Levent [2 ]
Toporek, Grzegorz [1 ]
Fuerst, Martin [3 ]
Kronreif, Gernot [3 ]
机构
[1] Univ Bern, ARTORG Ctr Biomed Engn, CH-3010 Bern, Switzerland
[2] Univ Hosp Bern, Inselspital, Inst Diagnost Intervent & Pediat Radiol, Bern, Switzerland
[3] Austrian Ctr Med Innovat & Technol, Wiener Neustadt, Austria
关键词
Computed tomography; helical; image guidance; interventional radiology; needle interventions; robotics; percutaneous biopsy; technology for surgical and interventional radiology; ASSISTED NAVIGATION SYSTEM; RADIATION-EXPOSURE; RESPIRATORY MOTION; BIOPSY; FLUOROSCOPY; INTERVENTIONS; IMAGE; PLACEMENT; ACCURACY; TESTS;
D O I
10.3109/13645706.2014.928641
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To test the feasibility of a robotic needle-guidance platform during CT-guided puncture ex vivo. Material and methods: Thin copper wires inserted into a torso phantom served as targets. The phantom was placed on a carbon plate and the robot-positioning unit (RPU) of the guidance platform (iSYS Medizintechnik GmbH, Kitzbuehel, Austria) was attached. Following CT imaging and automatic registration a double oblique trajectory was planned and the RPU was remotely moved into appropriate position and angulation. A 17G-puncture needle was then manually inserted until the preplanned depth, permanently guided by the RPU. The CT scan was repeated and the distance between the actual needle tip and the target was evaluated. Results: Automatic registration was successful in ten experiments and the median duration of an experiment was 9.6 (6.4-46.0) minutes. The angulation of the needle path in x-y and z-axis was within 15.6 degrees to 32.6 degrees, and -32.8 degrees to 3.2 degrees, respectively and the needle insertion depth was 92.8 +/- 14.4 mm. The Euclidean distance between the actual needle tip and the target was 2.3 +/- 0.8 (range, 0.9-3.7) mm. Conclusion: Automatic registration and accurate needle placement close to small targets was demonstrated. Study settings and torso phantom were very close to the clinical reality.
引用
收藏
页码:271 / 278
页数:8
相关论文
共 50 条
  • [41] Implementation of an electromagnetic tracking system for accurate intrahepatic puncture needle guidance: Accuracy results in an in vitro model
    Levy, Elliot B.
    Tang, Jonathan
    Lindisch, David
    Glossop, Neil
    Banovac, Filip
    Cleary, Kevin
    ACADEMIC RADIOLOGY, 2007, 14 (03) : 344 - 354
  • [42] Cervical Selective Nerve Root Block: Three-dimensional Puncture Planning With Dyna-CT Is Superior to Conventional CT-guidance in an Ex Vivo Model
    Schneider, Sebastian
    Kerl, Hans ulrich
    Freundt, Miriam
    Herrmann, Katrin E.
    Groden, Christoph
    Maros, MaTe E.
    Mohamed, Sherif A.
    IN VIVO, 2025, 39 (02): : 713 - 723
  • [43] Factors that contribute to false-negative results in CT-guided transthoracic lung core-needle biopsy
    Sun, Daqian
    Yao, Tianxiao
    Wang, Lixin
    Ma, Shengmei
    Gao, Min
    Li, Jingshuo
    Li, Chunhai
    Xu, Lisheng
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2023, 19 (01) : 86 - 91
  • [44] Puncture Cube Patient-Mounted Navigation System versus Freehand Method for CT-Guided Needle Placement: Study on a Neoprene Covered Elliptical Cylinder Gelatin Phantom
    Wessendorf, Joel
    Scheschenja, Michael
    Bastian, Moritz B.
    Gjini, Tefta
    Viniol, Simon
    Owczarek, Andreas
    Koenig, Alexander M.
    Jedelska, Jarmila
    Mahnken, Andreas H.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2024, 47 (11) : 1527 - 1531
  • [45] Robotic-Assisted Real-Time Image-Guided: From System Development to Ex Vivo Experiment
    Farouk, Z. I. Bibi
    Jiang, Shan
    Yang, Zhiyong
    Umar, Abubakar
    ARABIAN JOURNAL FOR SCIENCE AND ENGINEERING, 2023, 48 (03) : 3709 - 3722
  • [46] Improved Thermal Sensitivity Using Virtual Monochromatic Imaging Derived from Photon Counting Detector CT Data Sets: Ex Vivo Results of CT-Guided Cryoablation in Porcine Liver
    Josua A. Decker
    Franka Risch
    Florian Schwarz
    Christian Scheurig-Muenkler
    Thomas J. Kroencke
    CardioVascular and Interventional Radiology, 2023, 46 : 1385 - 1393
  • [47] Improved Thermal Sensitivity Using Virtual Monochromatic Imaging Derived from Photon Counting Detector CT Data Sets: Ex Vivo Results of CT-Guided Cryoablation in Porcine Liver
    Decker, Josua A.
    Risch, Franka
    Schwarz, Florian
    Scheurig-Muenkler, Christian
    Kroencke, Thomas J.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2023, 46 (10) : 1385 - 1393
  • [48] Evaluation of a Robotic Assistance-System For Percutaneous Computed Tomography-Guided (CT-Guided) Facet Joint Injection: A Phantom Study
    Beyer, Lukas Philipp
    Michalik, Katharina
    Niessen, Christoph
    da Silva, Natascha Platz Batista
    Wiesinger, Isabell
    Stroszczynski, Christian
    Wiggermann, Philipp
    MEDICAL SCIENCE MONITOR, 2016, 22 : 3334 - 3339
  • [49] Laser-guided percutaneous kidney access with the Uro Dyna-CT: first experience of three-dimensional puncture planning with an ex vivo model
    Ritter, M.
    Rassweiler, M. -C.
    Haecker, A.
    Michel, M. S.
    WORLD JOURNAL OF UROLOGY, 2013, 31 (05) : 1147 - 1151
  • [50] Computer-assisted navigation system for interventional CT-guided procedures: Results of phantom and clinical studies
    Meier-Meitinger, M.
    Nagel, M.
    Kalender, W.
    Bautz, W. A.
    Baum, U.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2008, 180 (04): : 310 - 317