A robotic needle-positioning and guidance system for CT-guided puncture: Ex vivo results

被引:44
作者
Kettenbach, Joachim [1 ]
Kara, Levent [2 ]
Toporek, Grzegorz [1 ]
Fuerst, Martin [3 ]
Kronreif, Gernot [3 ]
机构
[1] Univ Bern, ARTORG Ctr Biomed Engn, CH-3010 Bern, Switzerland
[2] Univ Hosp Bern, Inselspital, Inst Diagnost Intervent & Pediat Radiol, Bern, Switzerland
[3] Austrian Ctr Med Innovat & Technol, Wiener Neustadt, Austria
关键词
Computed tomography; helical; image guidance; interventional radiology; needle interventions; robotics; percutaneous biopsy; technology for surgical and interventional radiology; ASSISTED NAVIGATION SYSTEM; RADIATION-EXPOSURE; RESPIRATORY MOTION; BIOPSY; FLUOROSCOPY; INTERVENTIONS; IMAGE; PLACEMENT; ACCURACY; TESTS;
D O I
10.3109/13645706.2014.928641
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To test the feasibility of a robotic needle-guidance platform during CT-guided puncture ex vivo. Material and methods: Thin copper wires inserted into a torso phantom served as targets. The phantom was placed on a carbon plate and the robot-positioning unit (RPU) of the guidance platform (iSYS Medizintechnik GmbH, Kitzbuehel, Austria) was attached. Following CT imaging and automatic registration a double oblique trajectory was planned and the RPU was remotely moved into appropriate position and angulation. A 17G-puncture needle was then manually inserted until the preplanned depth, permanently guided by the RPU. The CT scan was repeated and the distance between the actual needle tip and the target was evaluated. Results: Automatic registration was successful in ten experiments and the median duration of an experiment was 9.6 (6.4-46.0) minutes. The angulation of the needle path in x-y and z-axis was within 15.6 degrees to 32.6 degrees, and -32.8 degrees to 3.2 degrees, respectively and the needle insertion depth was 92.8 +/- 14.4 mm. The Euclidean distance between the actual needle tip and the target was 2.3 +/- 0.8 (range, 0.9-3.7) mm. Conclusion: Automatic registration and accurate needle placement close to small targets was demonstrated. Study settings and torso phantom were very close to the clinical reality.
引用
收藏
页码:271 / 278
页数:8
相关论文
共 50 条
  • [31] 3D image-guided robotic needle positioning system for small animal interventions
    Bax, Jeffrey S.
    Waring, Christopher S. R.
    Sherebrin, Shi
    Stapleton, Shawn
    Hudson, Thomas J.
    Jaffray, David A.
    Lacefield, James C.
    Fenster, Aaron
    MEDICAL PHYSICS, 2013, 40 (01)
  • [32] Evaluation of a New CT-Guided Robotic System for Percutaneous Needle Insertion for Thermal Ablation of Liver Tumors: A Prospective Pilot Study
    Thierry de Baère
    Charles Roux
    Frédéric Deschamps
    Lambros Tselikas
    Boris Guiu
    CardioVascular and Interventional Radiology, 2022, 45 : 1701 - 1709
  • [33] Preliminary clinical experience with a dedicated interventional robotic system for CT-guided biopsies of lung lesions: a comparison with the conventional manual technique
    Anzidei, Michele
    Argiro, Renato
    Porfiri, Andrea
    Boni, Fabrizio
    Anile, Marco
    Zaccagna, Fulvio
    Vitolo, Domenico
    Saba, Luca
    Napoli, Alessandro
    Leonardi, Andrea
    Longo, Flavia
    Venuta, Federico
    Bezzi, Mario
    Catalano, Carlo
    EUROPEAN RADIOLOGY, 2015, 25 (05) : 1310 - 1316
  • [34] Results and complications of CT-guided transthoracic fine-needle aspiration biopsy of pulmonary lesions
    de Lima, Cristiano Dias
    Nunes, Rodolfo Acatauassu
    Saito, Eduardo Haruo
    Higa, Claudio
    Fernando Cardona, Zanier Jose
    dos Santos, Denise Barbosa
    JORNAL BRASILEIRO DE PNEUMOLOGIA, 2011, 37 (02) : 209 - 216
  • [35] CT-guided transthoracic needle biopsy of pulmonary nodules smaller than 20 mm: Results with an automated 20-gauge coaxial cutting needle
    Laurent, F
    Latrabe, V
    Vergier, B
    Montaudon, M
    Vernejoux, JM
    Dubrez, J
    CLINICAL RADIOLOGY, 2000, 55 (04) : 281 - 287
  • [36] Development of a robotic FD-CT-guided navigation system for needle placement - preliminary accuracy tests
    Tovar-Arriaga, Saul
    Tita, Ralf
    Carlos Pedraza-Ortega, Jesus
    Gorrostieta, Efren
    Kalender, Willi A.
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2011, 7 (02) : 225 - 236
  • [37] Percutaneous CT-guided fine-needle aspiration of pulmonary lesions: Results and complications in 409 patients
    Laspas, F.
    Roussakis, A.
    Efthimiadou, R.
    Papaioannou, D.
    Papadopoulos, S.
    Andreou, J.
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2008, 52 (05) : 458 - 462
  • [38] System for CT-guided needle placement in the thorax and abdomen: A design for clinical acceptability, applicability and usability
    Arnolli, Maarten M.
    Buijze, Martijn
    Franken, Michel
    de Jong, Koert P.
    Brouwer, Dannis M.
    Broeders, Ivo A. M. J.
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2018, 14 (01)
  • [39] Robotic CT-guided out-of-plane needle insertion: comparison of angle accuracy with manual insertion in phantom and measurement of distance accuracy in animals
    Toshiyuki Komaki
    Takao Hiraki
    Tetsushi Kamegawa
    Takayuki Matsuno
    Jun Sakurai
    Ryutaro Matsuura
    Takuya Yamaguchi
    Takanori Sasaki
    Toshiharu Mitsuhashi
    Soichiro Okamoto
    Mayu Uka
    Yusuke Matsui
    Toshihiro Iguchi
    Hideo Gobara
    Susumu Kanazawa
    European Radiology, 2020, 30 : 1342 - 1349
  • [40] Robotic CT-guided out-of-plane needle insertion: comparison of angle accuracy with manual insertion in phantom and measurement of distance accuracy in animals
    Komaki, Toshiyuki
    Hiraki, Takao
    Kamegawa, Tetsushi
    Matsuno, Takayuki
    Sakurai, Jun
    Matsuura, Ryutaro
    Yamaguchi, Takuya
    Sasaki, Takanori
    Mitsuhashi, Toshiharu
    Okamoto, Soichiro
    Uka, Mayu
    Matsui, Yusuke
    Iguchi, Toshihiro
    Gobara, Hideo
    Kanazawa, Susumu
    EUROPEAN RADIOLOGY, 2020, 30 (03) : 1342 - 1349