CrCoMo alloys are generally known to be highly resistant to corrosion-induced failures and to spontaneously form passive oxide films when immersed in physiological environments. The stresses resulting from the use of the prosthesis can cause fracture and abrasion of the oxide film that covers the surface. This can result in crevice corrosion, with the corresponding decrease in pH leading to severe attack. Protective coatings deposited onto the metals would minimise the release of metallic ions from the substrate and the incorporation of bioactive particles would enable the adhesion to the bone tissue. This work describes the electrochemical behavior of CrCoMo alloys (F75) as cast, covered by a hybrid silica coating obtained by sol-gel. The silica sol-gel coating contains bioactive glass ceramic and vitreous particles of the system CaO-SiO2-P2O5 previously obtained by fusion, crystallization and milling. The performance of the alloy with and without coating is evaluated in simulated body fluid (SBF) pH 7.31 at 37 degreesC. As cobalt base alloys are susceptible to crevice attack, coated pieces are also tested in SBF acid solution (pH 0.7) in order to simulate the acidic media and the high chloride concentrations that develop when a crevice is formed. The electrochemical behavior has been evaluated by potenciodynamic polarization and electrochemical impedance spectroscopy. It was found that for both pH conditions the samples coated with double layer of glass ceramic particles showed lower passive current densities than those with monolayers. The coating improves the protection potential of the base material in the neutral solution. Besides, in the acidic solution, all the samples remain passive but the protection potential shifts to potentials more active than those measured in the neutral SBF. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:1810 / 1814
页数:5
相关论文
共 10 条
[1]   CHROMIUM CONCENTRATIONS IN SERUM, BLOOD CLOT AND URINE FROM PATIENTS FOLLOWING TOTAL HIP-ARTHROPLASTY [J].
BARTOLOZZI, A ;
BLACK, J .
BIOMATERIALS, 1985, 6 (01) :2-8
[2]  
Hack H.P., 1987, METALS HDB CORROSION, VVolume 13, P234
[3]  
HAGINS H, 1988, J BONE JT SURG A, V71, P347
[4]   BIOCERAMICS - FROM CONCEPT TO CLINIC [J].
HENCH, LL .
JOURNAL OF THE AMERICAN CERAMIC SOCIETY, 1991, 74 (07) :1487-1510
[5]   Corrosion of metal orthopaedic implants [J].
Jacobs, JJ ;
Gilbert, JL ;
Urban, RM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1998, 80A (02) :268-282
[6]   SOLUTIONS ABLE TO REPRODUCE INVIVO SURFACE-STRUCTURE CHANGES IN BIOACTIVE GLASS-CERAMIC A-W3 [J].
KOKUBO, T ;
KUSHITANI, H ;
SAKKA, S ;
KITSUGI, T ;
YAMAMURO, T .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1990, 24 (06) :721-734
[7]   Human monocyte/macrophage response to cobalt-chromium corrosion products and titanium particles in patients with total joint replacements [J].
Lee, SH ;
Brennan, FR ;
Jacobs, JJ ;
Urban, RM ;
Ragasa, DR ;
Giant, TT .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1997, 15 (01) :40-49
[8]  
Li YS, 1999, J RAMAN SPECTROSC, V30, P97, DOI 10.1002/(SICI)1097-4555(199902)30:2<97::AID-JRS352>3.0.CO
[9]  
2-X
[10]   RELEASE OF COBALT AND NICKEL FROM A NEW TOTAL FINGER JOINT PROSTHESIS MADE OF VITALLIUM [J].
WOODMAN, JL ;
BLACK, J ;
NUNAMAKER, DM .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1983, 17 (04) :655-668