Validation of a liquid chromatography-mass spectrometric assay for tacrolimus in liver biopsies after hepatic transplantation: Correlation with histopathologic staging of rejection

被引:59
作者
Capron, Arnaud
Lerut, Jan
Verbaandert, Catherine
Mathys, Jules
Ciccarelli, Olga
Vanbinst, Roger
Roggen, Francine
De Reyck, Chantal
Lemaire, Julien
Wallemacq, Pierre E.
机构
[1] Univ Catholique Louvain, Dept Clin Chem, Lab Toxicol & Therapeut Drug Monitoring, Univ Hosp St Luc, B-1200 Brussels, Belgium
[2] Univ Catholique Louvain, Dept Abdominal Transplantat, Univ Hosp St Luc, B-1200 Brussels, Belgium
[3] Univ Catholique Louvain, Dept Pathol, Univ Hosp St Luc, B-1200 Brussels, Belgium
关键词
liver transplantation; drug monitoring; tacrolimus; tissue concentrations; graft rejection;
D O I
10.1097/FTD.0b013e31805c73f1
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The aims of this work were both to validate a sensitive and specific method to quantify tacrolimus (TAC) in liver biopsies after hepatic transplantation and to evaluate the predictive value of either tissue or blood TAC concentrations for rejection in 146 adult patients under a TAC-based immunosuppression. Trough blood levels were monitored daily during the hospital stay by immunoassay. Liver biopsies were routinely performed at day 7 posttransplantation. The tissue assay was developed by liquid chromatography-mass spectrometry. The limit of quantification was 5 pg/mg, with intra- and interassay precision ranging from 3.9% to 14.3% and 4.7% to 15.9%, respectively. The extraction efficiency was approximately 80%. TAC found in liver biopsies ranged from less than 5 up to 387 pg/mg. Blood TAC levels ranged from 2.7 to 19.3 ng/mL. Tissue levels displayed excellent correlation with liver histopathologic BANFF rejection score, whereas blood levels did not. Clinically significant rejections (BANFF scores 6) were characterized by mean TAC tissue and blood concentration of 13.1 pg/mg and 7.6 ng/mL, respectively, whereas these mean values became, respectively, 74.9 pg/mg (P < 0.05) and 7.1 ng/mL (not significant) for nonclinically significant rejection episodes (BANFF < 6). In this study, hepatic tissue TAC concentrations were distributed in a wider range and displayed a significantly better correlation with the severity of the organ rejection than predose blood levels. A tissue TAC concentration less than 30 pg/mg is 89% sensitive and 98% specific to discriminate clinically significant cellular rejection. Further studies are required to better understand the factors affecting TAC distribution within liver tissue (such as carrier proteins and cytochrome genetic polymorphism, liver function, age, hepatic blood flow, type of organ transplanted time posttransplantation) and to define its value in the treatment of liver allograft rejection.
引用
收藏
页码:340 / 348
页数:9
相关论文
共 63 条
[1]   Ion suppression in mass spectrometry [J].
Annesley, TM .
CLINICAL CHEMISTRY, 2003, 49 (07) :1041-1044
[2]  
BACKMAN L, 1994, TRANSPLANTATION, V57, P519
[3]  
BACKMAN L, 1995, TRANSPLANT P, V27, P1124
[4]   Cyclosporine lymphocyte level and lymphocyte count: New guidelines for tailoring immunosuppressive therapy [J].
Barbari, A ;
Stephan, A ;
Masri, M ;
Mourad, N ;
Kamel, G ;
Kilani, H ;
Karam, A ;
Daya, IA .
TRANSPLANTATION PROCEEDINGS, 2003, 35 (07) :2742-2744
[5]  
Barbari A G, 2006, Exp Clin Transplant, V4, P400
[6]  
Böttiger Y, 1999, BRIT J CLIN PHARMACO, V48, P445
[7]   Rapid simultaneous quantification of immunosuppressants in transplant patients by turbulent flow chromatography combined with tandem mass spectrometry [J].
Ceglarek, U ;
Lembcke, J ;
Fiedler, GM ;
Werner, M ;
Witzigmann, H ;
Hauss, JP ;
Thiery, J .
CLINICA CHIMICA ACTA, 2004, 346 (02) :181-190
[8]   Simultaneous determination of three isomeric metabolites of tacrolimus (FK506) in human whole blood and plasma using high performance liquid chromatography-tandem mass spectrometry [J].
Chen, YL ;
Hirabayashi, H ;
Akhtar, S ;
Pelzer, M ;
Kobayashi, M .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2006, 830 (02) :330-341
[9]   Automated, fast and sensitive quantification of drugs in blood by liquid chromatography-mass spectrometry with on-line extraction: immunosuppressants [J].
Christians, U ;
Jacobsen, W ;
Serkova, N ;
Benet, LZ ;
Vidal, C ;
Sewing, KF ;
Manns, MP ;
Kirchner, GI .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2000, 748 (01) :41-53
[10]   Active drug transport of immunosuppressants - New insights for pharmacokinetics and pharmacodynamics [J].
Christians, U ;
Strom, T ;
Zhang, YL ;
Steudel, W ;
Schmitz, V ;
Trump, S ;
Haschke, M .
THERAPEUTIC DRUG MONITORING, 2006, 28 (01) :39-44