The multi-module, multi-resolution system (M3R):: A novel small-animal SPECT system

被引:40
作者
Hesterman, Jacob Y.
Kupinski, Matthew A.
Furenlid, Lars R.
Wilson, Donald W.
Barrett, Harrison H.
机构
[1] Univ Arizona, Coll Opt Sci, Tucson, AZ 85721 USA
[2] Univ Arizona, Dept Radiol, Tucson, AZ 85721 USA
[3] Univ Arizona, Dept Radiol, Tucson, AZ 85724 USA
关键词
small-animal SPECT; hardware optimization; pinhole apertures; multi-pinhole apertures;
D O I
10.1118/1.2432071
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We have designed and built an inexpensive, high-resolution, tomographic imaging system, dubbed the multi-module, multi-resolution system, or (MR)-R-3. Slots machined into the system shielding allow for the interchange of pinhole plates, enabling the system to operate over a wide range of magnifications and with virtually any desired pinhole configuration. The flexibility of the system allows system optimization for specific imaging tasks and also allows for modifications necessary due to improved detectors, electronics, and knowledge of system construction (e.g., system sensitivity optimization). We provide an overview of M3R, focusing primarily on system design and construction, aperture construction, and calibration methods. Reconstruction algorithms will be described and reconstructed images presented. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:987 / 993
页数:7
相关论文
共 19 条
  • [1] Barrett HH., 2004, FDN IMAGE SCI
  • [2] Beekman FJ, 2005, J NUCL MED, V46, P1194
  • [3] Chen J. C., 1997, Biomedical Engineering, Applications Basis Communications, V9, P20
  • [4] Optimizing lens-coupled digital radiographic imaging systems based on model observers' performance
    Chen, LY
    Barrett, HH
    [J]. MEDICAL IMAGING 2003: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2003, 5034 : 63 - 70
  • [5] Calibration of scintillation cameras and pinhole SPECT Imaging systems
    Chen, YC
    Furenlid, LR
    Wilson, DW
    Barrett, HH
    [J]. Small Animal SPECT Imaging, 2005, : 195 - 201
  • [6] Application of task-based measures of image quality to optimization and evaluation of three-dimensional reconstruction-based compensation methods in myocardial perfusion SPECT
    Frey, EC
    Gilland, KL
    Tsui, BMW
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (09) : 1040 - 1050
  • [7] A multipinhole small animal SPECT system with submillimeter spatial resolution
    Funk, Tobias
    Despres, Philippe
    Barber, William C.
    Shah, Kanai S.
    Hasegawa, Bruce H.
    [J]. MEDICAL PHYSICS, 2006, 33 (05) : 1259 - 1268
  • [8] FastSPECT II: A second-generation high-resolution dynamic SPECT imager
    Furenlid, LR
    Wilson, DW
    Chen, YC
    Kim, H
    Pietraski, PJ
    Crawford, MJ
    Barrett, HH
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (03) : 631 - 635
  • [9] Optimizing a multiple-pinhole SPECT system using the ideal observer
    Gross, K
    Kupinski, MA
    Peterson, T
    Clarkson, E
    [J]. MEDICAL IMAGING 2003: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2003, 5034 : 314 - 322
  • [10] *HC STARCK CO, KUL DAT SHEET