Strategies for accurate response assessment of radiochromic film using flatbed scanner for beam quality assurance

被引:3
作者
Zhang, Xu-Dong [1 ]
Liu, Yuan-Hao [1 ,2 ]
Tang, Xiao-Bin [1 ]
Hsiao, Ming-Chen [2 ]
Chen, Wei-Lin [2 ]
Geng, Chang-Ran [1 ]
Shao, Wen-Cheng [1 ]
Gong, Chun-Hui [1 ,3 ]
Bortolussi, Silva [3 ]
Shu, Di-Yun [1 ]
机构
[1] Nanjing Univ Aeronaut & Astronaut, Dept Nucl Sci & Technol, Nanjing 210016, Peoples R China
[2] Neuboron Medtech Ltd, Nanjing 211112, Jiangsu, Peoples R China
[3] Natl Inst Nucl Phys INFN, Unit Pavia, Via A Bassi 6, I-27100 Pavia, Italy
基金
中国国家自然科学基金;
关键词
Radiochromic film; Response assessment; Scanning setting; Interscan variability; Standard operating procedure; GAFCHROMIC FILM; RADIATION; EBT2; OPTIMIZATION; DOSIMETER; THERAPY; ARRAY;
D O I
10.1007/s41365-019-0685-0
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Radiochromic film is a useful tool for beam quality assurance, but accurate response assessment of the film is still a problem. In this study, the response uncertainties of HDV2 film were investigated using a flatbed scanner from both the scanning settings and interscan variability. Scanning settings are fixed conditions for scanning, including scanning resolution and focus setting. In this study, multipeak distributions of pixel values were found under some dots-per-inch values, which should be avoided, and the optimal setting of 2000 dpi without this problem was selected. By changing the focus setting, the relative standard deviation of pixel values was reduced by 36-50%. The influence of the interscan variability induced by three factors was investigated, including the outside illumination intensity, film homogeneity, and operating temperature. Scanning the film before and after irradiation at the same position was recommended. Moreover, the suitable operating temperature range for the scanner was found to be 15-24 degrees C, which results in stable film responses. Regarding the studied factors, correction methods and strategies were proposed, and the accurate response assessment of HDV2 film was realized. Finally, a standard operating procedure for response assessment of films was introduced. It can help other researchers study more scanners, films, and particle types.
引用
收藏
页数:9
相关论文
共 25 条
[21]  
Spears K.E., 2005, OPTICAL IMAGE SCANNE
[22]   A study of four radiochromic films currently used for (2D) radiation dosimetry [J].
Yao, Tiantian ;
Luthjens, Leonard H. ;
Gasparini, Alessia ;
Warman, John M. .
RADIATION PHYSICS AND CHEMISTRY, 2017, 133 :37-44
[23]   INFLUENCE OF NEUTRON SOURCES AND 10B CONCENTRATION ON BORON NEUTRON CAPTURE THERAPY FOR SHALLOW AND DEEPER NON-SMALL CELL LUNG CANCER [J].
Yu, Haiyan ;
Tang, Xiaobin ;
Shu, Diyun ;
Liu, Yuanhao ;
Geng, Changran ;
Gong, Chunhui ;
Hang, Shuang ;
Chen, Da .
HEALTH PHYSICS, 2017, 112 (03) :258-265
[24]   Study on the coloration response of a radiochromic film to MeV cluster ion beams [J].
Yuri, Yosuke ;
Narumi, Kazumasa ;
Chiba, Atsuya ;
Hirano, Yoshimi ;
Saitoh, Yuichi .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 872 :126-130
[25]   Characterization of a Gafchromic film for the two-dimensional profile measurement of low-energy heavy-ion beams [J].
Yuri, Yosuke ;
Narumi, Kazumasa ;
Yuyama, Takahiro .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 828 :15-21