Measurement of gamma knife collimator factors using optically stimulated luminescence (OSL) dosimeter

被引:2
作者
Tong, S.
Yang, J.
Yahnke, C.
Beatty, R.
King, S.
Erdman, M.
机构
[1] Penn State Hershey Med Ctr, Hershey, PA USA
[2] Landauer, Glenwood, IL USA
关键词
D O I
10.1118/1.2760949
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Optically stimulated luminescence (OSL) technology is considered a breakthrough in radiation detection. The advantages of OSL technology includes: good linearity and energy response, non‐destructive readout, cost effective, easy to use, etc. It has been widely used in personnel and environmental radiation monitoring, and more recently it finds applications in radiation dose measurement in radiation therapy. The purpose of this study is to evaluate the Landauer microStar dosimetry system and its application in measurement of Leksell Gamma Knife collimator factors. Method and Materials: Landauer OSL Dot dosimeter, aluminum oxide crystals [formula omitted], was placed in the center of the Elekta standard 80 cm spherical phantom to be irradiated in the Gamma Knife Co‐60 beams. A calibration curve of doses vs. dosimeter readouts was established by irradiating the dosimeters in the 18 mm helmet with a series of known doses. The dosimeters were then mounted in other helmets and irradiated in order to measure the collimator factors. Multiple readouts on the dosimeters were carried out to evaluate consistency and reproducibility. Results: The dose calibration curve demonstrated a linear relationship between irradiated doses and dosimeter readouts. Least‐square fit of measured data shows a straight line with a slope of 1039 counts/cGy, and a correlation coefficient of 0.9998. Collimator factors for 14 mm and 8 mm helmets measured with the OSL dosimeters are 0.979±0.025 and 0.942±0.021, respectively. These values are in good agreement with the values provided by the Gamma Knife manufacturer. Measurement of the 4 mm collimator factor requires taking into account of the dosimeter dimensions and applying further corrections. More measurement results are to be presented. Conclusion: Preliminary result has shown that radiation dose measurement using the OSL dosimetry system is accurate and reproducible. The OSL dosimeter can be used to measure or verify Gamma Knife collimator factors. Research sponsored by Landauer. © 2007, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:2467 / 2468
页数:2
相关论文
共 50 条
[31]   Changes in the optically stimulated luminescence (OSL) sensitivity of single grains of quartz during the measurement of natural OSL: Implications for the reliability of optical ages [J].
Chauhan, Naveen ;
Singhvi, A. K. .
QUATERNARY GEOCHRONOLOGY, 2019, 53
[32]   Multi-Slice CT Dose Evaluation Using Optically Stimulated Luminescence (OSL) System [J].
Ruan, C. ;
Clouse, W. ;
Yukihara, E. ;
Gasparian, P. ;
Ahmad, S. .
MEDICAL PHYSICS, 2009, 36 (06)
[33]   New aspects of a BeO-based optically stimulated luminescence dosimeter [J].
Sommer, Marian ;
Freudenberg, Robert ;
Henniger, Juergen .
RADIATION MEASUREMENTS, 2007, 42 (4-5) :617-620
[34]   Investigation of regeneration effect of blue luminescence in NaCl using variable delay optically stimulated luminescence (VD-OSL) [J].
Biernacka, Magdalena ;
Mandowski, Arkadiusz .
RADIATION MEASUREMENTS, 2013, 56 :31-35
[35]   Investigation of backscatter factor in medical radiography using anthropomorphic phantom by optically stimulated luminescence dosimeter [J].
Arimoto, Akihiro ;
Asada, Yasuki .
BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2021, 7 (06)
[36]   Measurement of Gamma Knife® helmet factors using MOSFETs [J].
Kurjewicz, Laryssa ;
Berndt, Anita .
MEDICAL PHYSICS, 2007, 34 (03) :1007-1012
[37]   A simple and effective method for validation and measurement of collimator output factors for Leksell Gamma Knife® Perfexion™ [J].
Ma, Lijun ;
Kjall, Per ;
Novotny, Josef, Jr. ;
Nordstrom, Hakan ;
Johansson, Jonas ;
Verhey, Lynn .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (12) :3897-3907
[38]   Validation of the Commissioning of an Optically Stimulated Luminescence (OSL) System for Remote Dosimetry Audits [J].
Aguirre, J. ;
Alvarez, P. ;
Amador, C. ;
Tailor, A. ;
Followill, D. ;
Ibbott, G. .
MEDICAL PHYSICS, 2010, 37 (06)
[39]   Optically Stimulated Luminescence (OSL) Dosimeters Can Be Used for Remote Dosimetry Services [J].
Homnick, J. ;
Ibbott, G. ;
Springer, A. ;
Aguirre, J. .
MEDICAL PHYSICS, 2008, 35 (06)
[40]   Optically stimulated luminescence from BeO ceramics: An LM-OSL study [J].
Bulur, E. ;
Yeltik, A. .
RADIATION MEASUREMENTS, 2010, 45 (01) :29-34