Compression forces generated by Mini bone screws - a comparative study done on bone model

被引:31
作者
Adla, DN [1 ]
Kitsis, C
Miles, AW
机构
[1] Univ Wales Coll Cardiff, Leicester Gen Hosp, Dept Orthopaed Engn, Leicester, Leics, England
[2] Queens Hosp, Dept Orthopaed, Burton Upon Trent DE13 0RB, England
[3] Univ Bath, Ctr Orthopaed Biomech, Dept Mech Engn, Bath BA2 7AY, Avon, England
来源
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED | 2005年 / 36卷 / 01期
关键词
mini bone screws; load cell; compression forces; simulated bone model; experimental study;
D O I
10.1016/j.injury.2004.03.026
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
The compressive forces generated by the AO/ASIF 3.0 mm cannulated cancellous and 2.0 mm cortical screws, Mini-Acutrak and Herbert/Whipple small bone cannulated screws were measured in the laboratory with the use of simulated cancellous bone and a load cell washer as a means of quantifying their fixation capabilities. The Herbert/Whipple screw and the Mini-Acutrak screw were found to have nearly identical compression capabilities and provided more compression than the cortical screw. The AO/ASIF cannulated screw when used with a support screw demonstrated a compressive capacity twice that of the 2.0 mm cortical screw and higher than the headless Mini-Acutrak and Herbert/Whipple screws. The Mini-Acutrak screw produced about 70% of compression of the cancellous screw in spite of having a diameter almost half that of the cancellous screw. The Herbert/Whipple screw in spite of its larger size compared to the Mini-Acutrak produced almost the same amount of compression. (C) 2004 Elsevier Ltd. All rights reserved.
引用
收藏
页码:65 / 70
页数:6
相关论文
共 23 条
[1]  
[Anonymous], 1979, Manual of Internal Fixation
[2]   BIOMECHANICAL COMPARISON OF FIXATION DEVICES IN EXPERIMENTAL SCAPHOID OSTEOTOMIES [J].
CARTER, FM ;
ZIMMERMAN, MC ;
DIPAOLA, DM ;
MACKESSY, RP ;
PARSONS, JR .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1991, 16A (05) :907-912
[3]   Internal fixation of scaphoid fractures with an AO mini-fragment lag screw, using temporary interoperative AO mini external fixation [J].
Christodoulou, L ;
Chamberlain, ST .
JOURNAL OF HAND SURGERY-BRITISH AND EUROPEAN VOLUME, 1999, 24B (06) :676-678
[4]   Internal fixation of scaphoid non-union: a comparative study of three methods [J].
Christodoulou, LS ;
Kitsis, CK ;
Chamberlain, ST .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2001, 32 (08) :625-630
[5]  
CONRAD G, 1993, MANUAL SURG TECHNIQU
[6]   PERCUTANEOUS PINNING OF SYMPTOMATIC SCAPHOID NONUNIONS [J].
COSIO, MQ ;
CAMP, RA .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1986, 11A (03) :350-355
[7]   Effect of bone quality on the forces generated by compression screws [J].
Faran, KJ ;
Ichioka, N ;
Trzeciak, MA ;
Han, S ;
Medige, J ;
Moy, OJ .
JOURNAL OF BIOMECHANICS, 1999, 32 (08) :861-864
[8]  
GOH JC, 1986, ACTA ORTHOP SCAND, V60, P465
[9]  
Haddad F S, 1998, Chir Main, V17, P119
[10]  
INOUE G, 1991, Journal of Orthopaedic Surgical Techniques, V6, P1