Clinical and radiological Comparison of treatment of atlantoaxial instability by posterior C1-C2 transarticular screw fixation or C1 lateral mass-C2 pedicle screw fixation

被引:59
作者
Lee, Sun-Ho [1 ]
Kim, Eun Sang [1 ]
Sung, Joo-Kyung [2 ]
Park, Yeun-Mook [2 ]
Eoh, Whan [1 ]
机构
[1] Sungkyunkwan Univ, Sch Med, Dept Neurosurg, Samsung Med Ctr, Seoul 135710, South Korea
[2] Kyungpook Natl Univ, Sch Med, Dept Neurosurg, Taegu, South Korea
关键词
Atlantoaxial; Fixation; Harms technique; Instability; Transarticular screw; POLYAXIAL SCREW; TECHNICAL NOTE; FUSION; COMPLEX; STABILIZATION; ARTHRODESIS; SUITABILITY; SUBLUXATION; EXPERIENCE; GRAFT;
D O I
10.1016/j.jocn.2009.10.008
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
We compared the clinical and radiological results of posterior atlantoaxial fixation surgery using transarticular screws to those using a polyaxial screw-rod system in 55 patients with symptomatic atlantoaxial instability Patients underwent posterior C1-C2 fixation 28 patients (group 1) underwent C1-C2 transarticular screw fixation and 27 patients (group 2) underwent C1 lateral mass-C2 pedicle screw fixation Patients were followed-up for at least 24 months The clinical and radiological results were evaluated in the early postoperative period and at 3, 6, 12 and 24 months after surgery Long-term postoperative stability and bone fusion were examined After surgery, 93% of patients in group 1 and 96% of patients in group 2 were free of neck pain The solid fusion rates were 82% for group 1 patients and 96% for group 2 patients at 12 months (p < 0.092) In group 1, three patients showed fibrous union Four patients had hardware failure due to a screw malposition (one in group 1) and pseudoarthrodesis (two in group 1 and one in group 2) One patient in group 1 had cerebrospinal fluid leakage One patient in group 2 had occipital neuralgia One vertebral artery injury occurred during the screw placement in group 1 and another in group 2 during the muscle dissection C1-C2 transarticular screw fixation and C1 lateral mass-C2 pedicle screw fixation both produced excellent results for stabilization of the atlantoaxial complex, but the radiological outcome tended to be superior in C1 lateral mass-C2 pedicle screw fixation (C) 2009 Elsevier Ltd All rights reserved
引用
收藏
页码:886 / 892
页数:7
相关论文
共 50 条
[21]   Atlantoaxial arthrodesis using C1-C2 transarticular screw fixation in a case of Morquio syndrome [J].
Arvind G. Kulkarni ;
Siddharth M. Shah .
Indian Journal of Orthopaedics, 2011, 45 :470-472
[22]   Morphometric Characteristics of C1-C2 Vertebrae to Perform Anterior Transarticular Crossing Screw Fixation [J].
Tun, Kagan ;
Cemil, Berker ;
Gokce, Emre Cemal .
TURKISH NEUROSURGERY, 2018, 28 (02) :270-274
[23]   Application of Unilateral C2 Translaminar Screw in the Treatment for Atlantoaxial Instability as an Alternative or Salvage of Pedicle Screw Fixation [J].
Du, Shiyao ;
Ni, Bin ;
Lu, Xuhua ;
Xie, Ning ;
Guo, Xiang ;
Guo, Qunfeng ;
Yang, Jun ;
Chen, Fei .
WORLD NEUROSURGERY, 2017, 97 :86-92
[24]   The height for screw index (HSI) predicts the development of C2 nerve dysfunction associated with C1 lateral mass screw fixation for atlantoaxial instability [J].
Huang, Da-Geng ;
Hao, Ding-Jun ;
Jiang, Yong-Hong ;
Cheng, Yue ;
Pan, Jun-Wei ;
Qiang, Yong-Qian ;
He, Bao-Rong ;
Qian, Li-Xiong .
EUROPEAN SPINE JOURNAL, 2014, 23 (05) :1092-1098
[25]   C1-C2 Pedicle Screw Fixation for Treatment of Old Odontoid Fractures [J].
Qi, Lei ;
Li, Mu ;
Zhang, Shuai ;
Si, Haipeng ;
Xue, Jingsong .
ORTHOPEDICS, 2015, 38 (02) :94-100
[26]   Bilateral C1 laminar hooks combined with C2 pedicle screws fixation for treatment of C1-C2 instability not suitable for placement of transarticular screws [J].
Ni, Bin ;
Zhu, Zhuangchen ;
Zhou, Fengjin ;
Guo, Qunfeng ;
Yang, Jian ;
Liu, Jun ;
Wang, Fei .
EUROPEAN SPINE JOURNAL, 2010, 19 (08) :1378-1382
[27]   Safety Issues and Neurological Improvement following C1-C2 Fusion using C1 Lateral Mass and C2 Pedicle Screw in Atlantoaxial Instability [J].
Kwan, M. K. ;
Chan, C. Y. W. ;
Kwan, T. C. C. ;
Gashi, Y. N. ;
Saw, L. B. .
MALAYSIAN ORTHOPAEDIC JOURNAL, 2010, 4 (02) :17-22
[28]   Posterior C1-C2 screw-rod fixation and autograft fusion for the treatment of os odontoideum with C1-C2 instability [J].
Huang, Da-Geng ;
Wang, Tao ;
Hao, Ding-Jun ;
He, Bao-Rong ;
Liu, Tuan-Jiang ;
Ma, Xiao-Wen ;
Yu, Cheng-Cheng ;
Feng, Hang ;
Zhao, Song-Chum ;
Hui, Hua .
CLINICAL NEUROLOGY AND NEUROSURGERY, 2017, 163 :71-75
[29]   Anatomic suitability of the C1-C2 complex for pedicle screw fixation [J].
Resnick, DK ;
Lapsiwala, S ;
Trost, GR .
SPINE, 2002, 27 (14) :1494-1498
[30]   C1 Posterior Arch Crossing Screw Fixation for Atlantoaxial Joint Instability [J].
Jin, Guo-Xin ;
Wang, Huan ;
Li, Lei ;
Cui, Shao-Qian ;
Duan, Jing-Zhu .
SPINE, 2013, 38 (22) :E1397-E1404