The noise power spectrum in CT with direct fan beam reconstruction

被引:66
作者
Baek, Jongduk [1 ]
Pelc, Norbert J. [1 ,2 ,3 ]
机构
[1] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[2] Stanford Univ, Dept Elect Engn, Stanford, CA 94305 USA
[3] Stanford Univ, Dept Bioengn, Stanford, CA 94305 USA
关键词
computerised tomography; image reconstruction; medical image processing; noise; RAY COMPUTED-TOMOGRAPHY; IMAGE-RECONSTRUCTION; DETECTOR PERFORMANCE; RADIOGRAPHY SYSTEMS; PROPAGATION;
D O I
10.1118/1.3378673
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The noise power spectrum (NPS) is a useful metric for understanding the noise content in images. To examine some unique properties of the NPS of fan beam CT, the authors derived an analytical expression for the NPS of fan beam CT and validated it with computer simulations. The nonstationary noise behavior of fan beam CT was examined by analyzing local regions and the entire field-of-view (FOV). This was performed for cases with uniform as well as nonuniform noise across the detector cells and across views. The simulated NPS from the entire FOV and local regions showed good agreement with the analytically derived NPS. The analysis shows that whereas the NPS of a large FOV in parallel beam CT (using a ramp filter) is proportional to frequency, the NPS with direct fan beam FBP reconstruction shows a high frequency roll off. Even in small regions, the fan beam NPS can show a sharp transition (discontinuity) at high frequencies. These effects are due to the variable magnification and therefore are more pronounced as the fan angle increases. For cases with nonuniform noise, the NPS can show the directional dependence and additional effects.
引用
收藏
页码:2074 / 2081
页数:8
相关论文
共 29 条
[1]  
Alpert N M, 1982, IEEE Trans Med Imaging, V1, P142, DOI 10.1109/TMI.1982.4307561
[2]  
BARRETT H H, 1976, Computers in Biology and Medicine, V6, P307, DOI 10.1016/0010-4825(76)90068-8
[3]   OPTICAL TOMOGRAPHY - EXPERIMENTAL-VERIFICATION OF NOISE THEORY [J].
BENNETT, K ;
BYER, RL .
OPTICS LETTERS, 1984, 9 (07) :270-272
[4]   FAN-BEAM-TOMOGRAPHY NOISE THEORY [J].
BENNETT, KE ;
BYER, RL .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1986, 3 (05) :624-633
[5]   EXPERIMENTAL OPTICAL FAN BEAM TOMOGRAPHY [J].
BENNETT, KE ;
FARIS, GW ;
BYER, RL .
APPLIED OPTICS, 1984, 23 (16) :2678-2685
[6]   CT image reconstruction from fan-parallel data [J].
Besson, G .
MEDICAL PHYSICS, 1999, 26 (03) :415-426
[7]  
BUSHBERG JT, 2001, ESSENTIAL PHYS MED I, P238
[8]   ANALYSIS OF X-RAY COMPUTED-TOMOGRAPHY IMAGES USING THE NOISE POWER SPECTRUM AND AUTO-CORRELATION FUNCTION [J].
FAULKNER, K ;
MOORES, BM .
PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (11) :1343-1352
[9]  
HANSON K M, 1979, Medical Physics (Woodbury), V6, P441, DOI 10.1118/1.594534
[10]   A simple theorem relating noise anti patient dose in computed tomography [J].
Harpen, MD .
MEDICAL PHYSICS, 1999, 26 (11) :2231-2234