Point-of-care testing for coagulation function: CoaguChek® XS System versus standard laboratory testing in pediatric patients with normal and abnormal coagulation function

被引:3
作者
Kako, Hiromi [1 ,2 ]
Raman, Vidya T. [1 ,2 ]
Tumin, Dmitry [1 ]
Rice, Julie [1 ]
Tobias, Joseph D. [1 ,2 ]
机构
[1] Nationwide Childrens Hosp, Dept Anesthesiol & Pain Med, 700 Childrens Dr, Columbus, OH 43205 USA
[2] Ohio State Univ, Dept Anesthesiol & Pain Med, 410 W 10th Ave, Columbus, OH 43210 USA
关键词
Point-of-care testing; Coagulation function; CoaguChek (R) XS System; International normalized ratio; INR;
D O I
10.1007/s00540-017-2317-3
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Purpose Intraoperative abnormalities of coagulation function may occur for various reasons. In most scenarios, treatment is directed by laboratory parameters. Unfortunately, standard laboratory testing may take 1-2 h. The purpose of the current study was to evaluate a point-of-care testing device (CoaguChek (R) XS System) in pediatric patients. Methods Patients ranging in age from 2 to 18 years, undergoing posterior spinal fusion (PSF) or cardiac surgery using cardiopulmonary bypass (CPB) were eligible for inclusion. After CPB and/or the surgical procedure, 2.8 ml of blood was obtained and simultaneously tested on both the standard laboratory apparatus and the CoaguChek (R) XS System. Results The study cohort consisted of 100 patients (50 PSF and 50 cardiac cases) with 13 cases excluded, leaving 87 patients (49 PSF and 38 cardiac cases) for analysis. In PSF cases, reference laboratory international normalized ratio (INR) ranged from 0.98 to 1.77 while CoaguChek (R) XS INR ranged from 1.0 to 1.3. The correlation coefficient was 0.69. The results of the Bland-Altman analysis showed a bias of 0.09, precision of 0.1, and 95% limits of agreement ranging from -0.11 to 0.28. In cardiac cases, reference INR ranged from 1.68 to 14.19, while CoaguChek (R) XS INR ranged from 1.4 to 7.9. The correlation coefficient was 0.35. The results of the Bland-Altman analysis showed a bias of -1.8, precision of 2.1, and 95% limits of agreement ranging from -6.0 to 2.4. Conclusions INR values obtained from CoaguChek (R) XS showed a moderate correlation with reference laboratory values within the normal range. However, in the presence of coagulopathy, the discrepancy was significantly greater, thereby making the CoaguChek (R) XS clinically unreliable.
引用
收藏
页码:345 / 350
页数:6
相关论文
共 6 条
  • [2] Agreement between Coaguchek XS and STA-R Evolution (Hepato Quick) INR results depends on the level of INR
    Biedermann, Joseph S.
    Leebeek, Frank W. G.
    Buhre, Peter N.
    de Lathouder, Sacha
    Barends, Jan P. F.
    de Maat, Moniek P. M.
    van der Meer, Felix J. M.
    Kruip, Marieke J. H. A.
    [J]. THROMBOSIS RESEARCH, 2015, 136 (03) : 652 - 657
  • [3] Horlocker TT, 2001, ANESTH ANALG, V93, P33
  • [4] Multidisciplinary Approach to the Challenge of Hemostasis
    Levy, Jerrold H.
    Dutton, Richard P.
    Hemphill, J. Claude, III
    Shander, Aryeh
    Cooper, David
    Paidas, Michael J.
    Kessler, Craig M.
    Holcomb, John B.
    Lawson, Jeffre H.
    [J]. ANESTHESIA AND ANALGESIA, 2010, 110 (02) : 354 - 364
  • [5] Comparison of the international normalized ratio levels obtained by portable coagulometer and laboratory in a clinic specializing in oral anticoagulation
    Meneghelo, Z. M.
    Barroso, C. M. Q.
    Liporace, I. L.
    Cora, A. P.
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2015, 37 (04) : 536 - 543
  • [6] Comparison of point-of-care testing (POCT): i-STAT® international normalized ratio (INR) vs reference laboratory INR in pediatric patients undergoing major surgery
    Spielmann, Nelly
    Mauch, Jacqueline Y.
    Madjdpour, Caveh
    Schmugge, Markus
    Albisetti, Manuela
    Weiss, Markus
    Haas, Thorsten
    [J]. PEDIATRIC ANESTHESIA, 2011, 21 (10) : 1041 - 1045