Surface Alterations of Polished and Sandblasted and Acid-Etched Titanium Implants after Er:YAG, Carbon Dioxide, and Diode Laser Irradiation

被引:1
|
作者
Stuebinger, Stefan [1 ,2 ]
Etter, Christof [1 ]
Miskiewicz, Miroslaw [1 ]
Homann, Frank [3 ]
Saldamli, Belma [1 ,2 ]
Wieland, Marco [3 ]
Sader, Robert [2 ]
机构
[1] Univ Basel, Hightech Res Ctr Craniomaxillofacial Surg, Basel, Switzerland
[2] Goethe Univ Frankfurt, Med Ctr, Dept Oral Craniomaxillofacial & Facial Plast Surg, Frankfurt, Germany
[3] Inst Straumann, Basel, Switzerland
关键词
dental implants; peri-implantitis; scanning electron microscopy; titanium disk; ER-YAG LASER; NONSURGICAL TREATMENT; PERI-IMPLANTITIS; DENTAL IMPLANTS; GAALAS LASER; PERIIMPLANTITIS LESIONS; CLINICAL-EVALUATION; TEMPERATURE-CHANGES; BONE; CO2;
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: Laser treatment has become a popular method for resolving peri-implantitis, but the full range of its effects on implant surfaces is unknown. The purpose of the present investigation was to analyze the influence of different clinically applicable erbium:yttrium-aluminum-garnet (Er:YAG), carbon dioxide (CO2), and diode laser parameters on titanium surfaces that were either polished or sandblasted, large-grit, acid-etched (SLA). Materials and Methods: Six polished and six SLA titanium disks were irradiated at nine different power settings (n = 54 polished, 54 SLA) with Er: YAG, CO2, or diode lasers. The CO2 and diode lasers were used in continuous wave mode, and the Er: YAG laser was used in a pulsed manner. The surface of each disk was analyzed by scanning electron microscopy and confocal white light microscopy. Each disk was irradiated on six circular areas of 5 mm in diameter with the same specific laser setting for 10 seconds. Results: Within the chosen parameters, the CO2 and diode laser did not cause any visible surface alterations on either the polished or SLA disks. In contrast, both polished and SLA disks showed surface alterations when irradiated with the pulsed Er: YAG laser. The SLA surfaces showed alteration after 10 seconds of irradiation with Er: YAG laser at 300 mJ/10 Hz. The surfaces of the polished disks did not show alteration with the Er: YAG laser until they were irradiated at the higher energy of 500 mJ/10 Hz for 10 seconds. The results of confocal white light microscopy were in agreement with scanning electron micrographs. Conclusion: In contrast to continuous-wave diode and CO2 laser irradiation, pulsed Er: YAG laser irradiation caused distinct alterations with power settings beyond 300 mJ/10 Hz on the SLA surface and 500 mJ/10 Hz on the polished surface. Thus, it is only safe to use the Er: YAG laser for implant surface irradiation with settings no higher than 300 or 500 mJ/10 Hz. INT J ORAL MAXILLOFAC IMPLANTS 2010; 25: 104-111
引用
收藏
页码:104 / 111
页数:8
相关论文
共 50 条
  • [1] Alterations in Surface Roughness and Chemical Characteristics of Sandblasted and Acid-Etched Titanium Implants after Irradiation with Different Diode Lasers
    Kim, Hak-Ki
    Park, Su-Yeon
    Son, Keunbada
    Kim, Yong-Gun
    Yu, Won-Jae
    Lee, Kyu-Bok
    Lee, Jae-Mok
    APPLIED SCIENCES-BASEL, 2020, 10 (12):
  • [2] Biomechanical comparison of the sandblasted and acid-etched and the machined and acid-etched titanium surface for dental implants
    Li, DH
    Ferguson, SJ
    Beutler, T
    Cochran, DL
    Sittig, C
    Hirt, HP
    Buser, D
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2002, 60 (02): : 325 - 332
  • [3] Surface analysis of machined versus sandblasted and acid-etched titanium implants
    Orsini, G
    Assenza, B
    Scarano, A
    Piattelli, M
    Piattelli, A
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2000, 15 (06) : 779 - 784
  • [4] Surface analysis of machined and sandblasted and acid-etched (SLA) titanium implants
    Piattelli, M
    Iezzi, G
    Orsini, G
    Petrone, G
    Piattelli, M
    Scarano, A
    JOURNAL OF DENTAL RESEARCH, 2001, 80 (04) : 1245 - 1245
  • [5] The Effects of Erbium-Doped Yttrium Aluminum Garnet Laser (Er: YAG) Irradiation on Sandblasted and Acid-Etched (SLA) Titanium, an In Vitro Study
    Scarano, Antonio
    Lorusso, Felice
    Inchingolo, Francesco
    Postiglione, Francesca
    Petrini, Morena
    MATERIALS, 2020, 13 (18)
  • [6] Efficacy of plasma treatment for surface cleansing and osseointegration of sandblasted and acid-etched titanium implants
    Bae, Gang-Ho
    Cho, Won-Tak
    Lee, Jong-Ho
    Huh, Jung-Bo
    JOURNAL OF ADVANCED PROSTHODONTICS, 2024, 16 (03): : 189 - 199
  • [7] Early loading of non-submerged titanium implants with a sandblasted and acid-etched surface
    Bornstein, MM
    Schmid, B
    Belser, UC
    Lussi, A
    Buser, D
    CLINICAL ORAL IMPLANTS RESEARCH, 2005, 16 (06) : 631 - 638
  • [8] Bone response is rabbit to machined and sandblasted and acid-etched titanium implants
    Ripari, M
    Maggiore, C
    Scarano, A
    Piattelli, A
    JOURNAL OF DENTAL RESEARCH, 2002, 81 : B262 - B262
  • [9] Microleakage of composite restorations after Er. YAG laser irradiation versus acid-etched
    Ceballos, L
    Osorio, R
    Toledano, M
    Marshall, GW
    JOURNAL OF DENTAL RESEARCH, 2000, 79 : 536 - 536
  • [10] Surface characteristics and biocompatibility of sandblasted and acid-etched titanium surface modified by ultraviolet irradiation: An in vitro study
    Li, Shaobing
    Ni, Jia
    Liu, Xiangning
    Zhang, Xueyang
    Yin, Shiheng
    Rong, Mingdeng
    Guo, Zehong
    Zhou, Lei
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2012, 100B (06) : 1587 - 1598