Arterial Input Function Placement for Accurate CT Perfusion Map Construction in Acute Stroke

被引:41
作者
Ferreira, Rafael M. [2 ]
Lev, Michael H. [2 ]
Goldmakher, Gregory V. [2 ]
Kamalian, Shahmir [2 ]
Schaefer, Pamela W. [2 ]
Furie, Karen L. [2 ]
Gonzalez, R. Gilberto [2 ]
Sanelli, Pina C. [1 ]
机构
[1] New York Presbyterian Hosp, Dept Radiol, Weill Cornell Med Coll, New York, NY 10065 USA
[2] Massachusetts Gen Hosp, Dept Radiol, Boston, MA 02114 USA
关键词
acute stroke; arterial input function; CT perfusion; deconvolution method; middle cerebral artery clot; postprocessing perfusion software; SINGULAR-VALUE DECOMPOSITION; COMPUTED-TOMOGRAPHY; CEREBRAL-ARTERY; MR PERFUSION; VALUES; ISCHEMIA; VOLUME; FLOW;
D O I
10.2214/AJR.09.2845
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The objective of our study was to evaluate the effect of varying arterial input function (AIF) placement on the qualitative and quantitative CT perfusion parameters. MATERIALS AND METHODS. Retrospective analysis of CT perfusion data was performed on 14 acute stroke patients with a proximal middle cerebral artery (MCA) clot. Cerebral blood flow (CBF), cerebral blood volume (CBV), and mean transit time (MTT) maps were constructed using a systematic method by varying only the AIF placement in four positions relative to the MCA clot including proximal and distal to the clot in the ipsilateral and contralateral hemispheres. Two postprocessing software programs were used to evaluate the effect of AIF placement on perfusion parameters using a delay-insensitive deconvolution method compared with a standard deconvolution method. RESULTS. One hundred sixty-eight CT perfusion maps were constructed for each software package. Both software programs generated a mean CBF at the infarct core of < 12 mL/100 g/min and a mean CBV of < 2 mL/100 g for AIF placement proximal to the clot in the ipsilateral hemisphere and proximal and distal to the clot in the contralateral hemisphere. For AIF placement distal to the clot in the ipsilateral hemisphere, the mean CBF significantly increased to 17.3 mL/100 g/min with delay-insensitive software and to 19.4 mL/100 g/min with standard software (p < 0.05). The mean MTT was significantly decreased for this AIF position. Furthermore, this AIF position yielded qualitatively different parametric maps, being most pronounced with MTT and CBF. Overall, CBV was least affected by AIF location. CONCLUSION. For postprocessing of accurate quantitative CT perfusion maps, laterality of the AIF location is less important than avoiding AIF placement distal to the clot as detected on CT angiography. This pitfall is less severe with deconvolution-based software programs using a delay-insensitive technique than with those using a standard deconvolution method.
引用
收藏
页码:1330 / 1336
页数:7
相关论文
共 22 条
  • [1] Effect of the arterial input function on the measured perfusion values and infarct volumetric in acute cerebral ischemia evaluated by perfusion computed tomography
    Bisdas, Sotirios
    Konstantinou, George N.
    Gurung, Jessen
    Lehnert, Thomas
    Donnerstag, Frank
    Becker, Hartmut
    Vogl, Thomas J.
    Koh, Tong San
    [J]. INVESTIGATIVE RADIOLOGY, 2007, 42 (03) : 147 - 156
  • [2] Calamante F, 2000, MAGN RESON MED, V44, P466, DOI 10.1002/1522-2594(200009)44:3<466::AID-MRM18>3.0.CO
  • [3] 2-M
  • [4] CT perfusion scanning with deconvolution analysis: Pilot study in patients with acute middle cerebral artery stroke
    Eastwood, JD
    Lev, MH
    Azhari, T
    Lee, TY
    Barboriak, DP
    Delong, DM
    Fitzek, C
    Herzau, M
    Wintermark, M
    Meuli, R
    Brazier, D
    Provenzale, JM
    [J]. RADIOLOGY, 2002, 222 (01) : 227 - 236
  • [5] Ibaraki Masanobu, 2005, Magn Reson Med Sci, V4, P27, DOI 10.2463/mrms.4.27
  • [6] Perfusion CT compared to H215O/O15O PET in patients with chronic cervical carotid artery occlusion
    Kamath, Amita
    Smith, Wade S.
    Powers, William J.
    Cianfoni, Alessandro
    Chien, Jeffrey D.
    Videen, Tom
    Lawton, Michael T.
    Finley, Bruce
    Dillon, William P.
    Wintermark, Max
    [J]. NEURORADIOLOGY, 2008, 50 (09) : 745 - 751
  • [7] Combined perfusion- and diffusion-weighted MR imaging in acute ischemic stroke during the 1st week: A longitudinal study
    Karonen, JO
    Liu, YW
    Vanninen, RL
    Ostergaard, L
    Partanen, PLK
    Vainio, PA
    Vanninen, EJ
    Nuutinen, J
    Roivainen, R
    Soimakallio, S
    Kuikka, IT
    Aronen, HJ
    [J]. RADIOLOGY, 2000, 217 (03) : 886 - 894
  • [8] User-defined vascular input function curves: Influence on mean perfusion parameter values and signal-to-noise ratio
    Kealey, SM
    Loving, VA
    Delong, DM
    Eastwood, JD
    [J]. RADIOLOGY, 2004, 231 (02) : 587 - 593
  • [9] MEASUREMENT OF REGIONAL CEREBRAL BLOOD-VOLUME BY COMPUTERIZED AXIAL TOMOGRAPHY
    LADURNER, G
    ZILKHA, E
    ILIFF, LD
    MARSHALL, J
    DUBOULAY, GH
    [J]. JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1976, 39 (02) : 152 - 158
  • [10] Functional CT: physiological models
    Lee, Ting-Yim
    [J]. TRENDS IN BIOTECHNOLOGY, 2002, 20 (08) : S3 - S10