A cine-EPID based method for jaw detection and quality assurance for tracking jaw in IMRT/VMAT treatments

被引:16
作者
Fuangrod, Todsaporn [1 ]
Rowshanfarzad, Pejman [2 ]
Greer, Peter B. [3 ,4 ]
Middleton, Richard H. [1 ]
机构
[1] Univ Newcastle, Fac Engn & Built Environm, Sch Elect Engn & Comp Sci, Callaghan, NSW 2308, Australia
[2] Univ Western Australia, Sch Phys, Crawley, WA 6009, Australia
[3] Univ Newcastle, Fac Sci & IT, Sch Math & Phys Sci, Callaghan, NSW 2308, Australia
[4] Calvary Mater Newcastle Hosp, Dept Radiat Oncol, Newcastle, NSW 2310, Australia
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2015年 / 31卷 / 01期
关键词
Real-time verification; Jaws; QA; EPID; RADIOCHROMIC FILM; DOSIMETRY;
D O I
10.1016/j.ejmp.2014.11.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new tool with the potential to verify and track jaw position during delivery has been developed. The method should be suitable for independent quality assurance for jaw position during jaw tracking dynamic IMRT and VMAT treatments. The jaw detection and tracking algorithm developed consists of five main steps. Firstly, the image is enhanced by removing a normalised predicted EPID image (that does not include the collimator transmission) from each cine EPID image. Then, using a histogram clustering technique a global intensity threshold level was determined. This threshold level was used to classify each pixel of the image as either under the jaws or under the MLC. Additionally, the collimator angle was automatically detected and used to rotate the image to vertical direction. Finally, this rotation allows the jaw positions to be determined using vertical and horizontal projection profiles. Nine IMRT fields (with static jaws) and a single VMAT clinical field (with dynamic jaws) were tested by determining the root mean square difference between planned and detected jaw positions. The test results give a detection accuracy of +/-1 mm RMS error for static jaw IMRT treatments and +/-1.5 mm RMS error for the dynamic jaw VMAT treatment. This method is designed for quality assurance and verification in modern radiation therapy; to detect the position of static jaws or verify the position of tracking jaws in more complex treatments. This method uses only information extracted from EPID images and it is therefore independent from the linear accelerator. (C) 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:16 / 36
页数:21
相关论文
共 10 条
[1]  
Alber M., 2008, Guidelines for the verification of IMRT
[2]  
Boyer AL, 2012, TREATMENT PLANNING R, P201
[3]   A comparison between radiochromic EBT2 film model and its predecessor EBT film model [J].
Carrasco, M. A. ;
Perucha, M. ;
Luis, F. J. ;
Baeza, M. ;
Herrador, M. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (04) :412-422
[4]   Radiochromic film dosimetry: Past, present, and future [J].
Devic, Slobodan .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2011, 27 (03) :122-134
[5]   An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan [J].
Lewis, David ;
Micke, Andre ;
Yu, Xiang ;
Chan, Maria F. .
MEDICAL PHYSICS, 2012, 39 (10) :6339-6350
[6]   Dosimetry tools and techniques for IMRT [J].
Low, Daniel A. ;
Moran, Jean M. ;
Dempsey, James F. ;
Dong, Lei ;
Oldham, Mark .
MEDICAL PHYSICS, 2011, 38 (03) :1313-1338
[7]   Enhanced dosimetry procedures and assessment for EBT2 radiochromic film [J].
Mayer, Rulon R. ;
Ma, Fuhwa ;
Chen, Yu ;
Miller, Rachel I. ;
Belard, Arnaud ;
McDonough, James ;
O'Connell, John J. .
MEDICAL PHYSICS, 2012, 39 (04) :2147-2155
[8]   Multichannel film dosimetry with nonuniformity correction [J].
Micke, Andre ;
Lewis, David F. ;
Yu, Xiang .
MEDICAL PHYSICS, 2011, 38 (05) :2523-2534
[9]  
Press W. H., 1997, The art of scientific computing, V1
[10]  
Smyth G. K., 2002, ENCY ENV, V3, P1405