Blood pressure-independent neurogenic effect on conductance and resistance vessels: a consideration for cuffless blood pressure measurement?

被引:5
|
作者
Cox, James [1 ]
Avolio, Alberto P. [1 ]
Louka, Kyrollos [1 ]
Shirbani, Fatemeh [1 ]
Tan, Isabella [1 ]
Butlin, Mark [1 ]
机构
[1] Macquarie Univ, Fac Med Hlth & Human Sci, Macquarie Med Sch, Sydney, NSW, Australia
来源
2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC) | 2021年
关键词
cuffless blood pressure; photoplethysmography; tonometry; pulse transit time; pulse arrival time; arterial stiffness; wearable devices; HEART-RATE; DEPENDENCY;
D O I
10.1109/EMBC46164.2021.9629936
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Pulse transit time (PTT) and pulse arrival time (PAT) are promising measures for cuffless arterial blood pressure (BP) estimation given the intrinsic arterial stiffness-BP relationship. However, arterial stiffness (and PTT) is altered by autonomically-driven smooth muscle tension changes, potentially independent of BP. This would limit PTT or PAT as accurate BP correlates, more so in resistance vessels than conductance arteries. Objective: To quantify if there is a measurable neurogenic effect on PAT measured using photoplethysmography (PPG) (path includes resistance vessels) and radial artery tonometry (path includes only conductance vessels) during physiologically induced BP changes. Methods: PATs were measured continuously in participants (n=15, 35 +/- 15 years, 9 male) using an electrocardiogram and, simultaneously, a Finometer (R) PRO finger sensor, a finger PPG sensor and radial artery tonometer during seated rest, cold pressor test, cycling and isometric handgrip (IHG) exercise. Delta BP/Delta PAT was calculated for each sensor and each condition. Results: All interventions significantly increased BP. A significant difference was observed in Delta BP/Delta PAT between cycling and both the cold pressor test and IHG exercise (p<0.05). Delta BP/Delta PAT did not differ whether measured via PPG or tonometry. Conclusions: Under the conditions tested, autonomic function does not have a BP-independent effect on PAT where the path includes resistance vessels (PPG signal), likely due to the speed of the wave and the short path length of resistance vessels. Autonomic function therefore does not limit the ability for use of PPG as a signal for potentially estimating BP without a cuff.
引用
收藏
页码:7485 / 7488
页数:4
相关论文
共 50 条
  • [21] Cuffless Blood Pressure Measurement Using Smartwatches: A Large-Scale Validation Study
    Liu, Zeng-Ding
    Li, Ye
    Zhang, Yuan-Ting
    Zeng, Jia
    Chen, Zu-Xian
    Cui, Zhi-Wei
    Liu, Ji-Kui
    Miao, Fen
    IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2023, 27 (09) : 4216 - 4227
  • [22] Blood Pressure Continuous Measurement through a Wearable Device: Development and Validation of a Cuffless Method
    De Marchi, Beatrice
    Frigerio, Mattia
    De Nadai, Silvia
    Longinotti-Buitoni, Gianluigi
    Aliverti, Andrea
    SENSORS, 2021, 21 (21)
  • [23] Cuffless Blood Pressure Monitoring Promises and Challenges
    Pandit, Jay A.
    Lores, Enrique
    Batlle, Daniel
    CLINICAL JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2020, 15 (10): : 1531 - 1538
  • [24] Cuffless blood pressure monitoring using hydrostatic pressure changes
    Shaltis, Phillip A.
    Reisner, Andrew T.
    Asada, H. Harry
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2008, 55 (06) : 1775 - 1777
  • [25] The reality and serendipity of cuffless blood pressure monitoring
    Alberto Avolio
    Hypertension Research, 2023, 46 : 1609 - 1611
  • [26] The reality and serendipity of cuffless blood pressure monitoring
    Avolio, Alberto
    HYPERTENSION RESEARCH, 2023, 46 (06) : 1609 - 1611
  • [27] Cuffless Single-Site Photoplethysmography for Blood Pressure Monitoring
    Hosanee, Manish
    Chan, Gabriel
    Welykholowa, Kaylie
    Cooper, Rachel
    Kyriacou, Panayiotis A.
    Zheng, Dingchang
    Allen, John
    Abbott, Derek
    Menon, Carlo
    Lovell, Nigel H.
    Howard, Newton
    Chan, Wee-Shian
    Lim, Kenneth
    Fletcher, Richard
    Ward, Rabab
    Elgendi, Mohamed
    JOURNAL OF CLINICAL MEDICINE, 2020, 9 (03)
  • [28] Pulse transit time technique for cuffless unobtrusive blood pressure measurement: from theory to algorithm
    Xiaorong Ding
    Yuan-Ting Zhang
    Biomedical Engineering Letters, 2019, 9 : 37 - 52
  • [29] Continual Learning for Cuffless Blood Pressure Measurement using PPG and ECG Signals
    Zhang, Chunlin
    Shen, Zhan
    Ding, Xiaorong
    2023 45TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY, EMBC, 2023,
  • [30] Validation of Cuffless Blood Pressure Monitoring Using Wearable Device
    Baek, JinHyeok
    Kim, Junho
    Kim, Namho
    Lee, Daeho
    Park, Sung-Min
    PROCEEDINGS OF TENCON 2018 - 2018 IEEE REGION 10 CONFERENCE, 2018, : 0416 - 0419