Evaluation of a filling system for binary pediatric solutions

被引:4
作者
Combeau, D
Rey, JB
Fontan, JE
Arnaud, P
Brion, F
机构
[1] Hop Robert Debre AP HP, Serv Pharm, F-75019 Paris, France
[2] Lab Toxicopharmacol, F-75019 Paris, France
来源
PHARMACY WORLD & SCIENCE | 1998年 / 20卷 / 01期
关键词
automatic filling system; parenteral nutrition; pediatric; pharmacy; volumetric accuracy;
D O I
10.1023/A:1008640426705
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
The aim of this study was to assess the use of an automatic filling system (Siframix(TM) M31 and M32 system) to prepare pediatric parenteral nutrition. volumetric accuracy was measured for each siframix(TM) system loads cells (< 5% for 5 ml with the Siframix(TM) M32 and < 5% for 9 ml with the Siframix(TM) M31) with sterile water for injection, The minimal 20 ml of flushing sterile water of the common tubulure of the Siframix(TM) M32 (p=0.211 for 20 ml and p=0.75 for 500 ml), the use of viscous solutions (70 % dextrose) on the Siframix(TM) M31 (p=0.28 for 20 ml and p=0.12 for 500 ml) IX and the use of a special tubulure for using E.V.A. Luer-lock bags (p=0.89 for 20 mi and p=0.103 for 500 mi) do not modify the accuracy. Changing bags or bottles during the filling operation modify the accuracy (p=0.004 for 20 ml and p=0.009 for 500 ml). A flushing operation is necessary to lower the risk of electrolytic pollution for the filling of little bags. The filling speed for each module as also measured (the maximal filling speed was five liters per minute). The Siframix(TM) system allows one to prepare pediatric parenteral nutrition bags when volumes are above 4 ml and with adapted source solutions in terms of concentration and conditioning volumes.
引用
收藏
页码:32 / 37
页数:6
相关论文
共 12 条
  • [1] CHOUX C, 1990, ACT PHARM BIOL CLIN, V5, P461
  • [2] DAVIS NM, 1992, HOSP PHARM, V27, P488
  • [3] AUTOMATED COMPOUNDER FOR ADDING INGREDIENTS TO PARENTERAL NUTRIENT BASE SOLUTIONS
    DICKSON, LB
    SOMANI, SM
    HERRMANN, G
    ABRAMOWITZ, PW
    [J]. AMERICAN JOURNAL OF HOSPITAL PHARMACY, 1993, 50 (04): : 678 - 682
  • [4] CALCIUM AND PHOSPHORUS COMPATIBILITY IN PARENTERAL-NUTRITION SOLUTIONS FOR NEONATES
    EGGERT, LD
    RUSHO, WJ
    MACKAY, MW
    CHAN, GM
    [J]. AMERICAN JOURNAL OF HOSPITAL PHARMACY, 1982, 39 (01): : 49 - 53
  • [5] GURA KM, 1993, 50 ASHP ANN M JUN
  • [6] HEIDE PE, 1993, KRANKENHAUSPHARMAZIE, V11, P505
  • [7] MAROUN C, 1992, 27 ASHP MID CLIN M D
  • [8] ENSURING ACCURACY IN THE USE OF AUTOMATIC COMPOUNDERS
    MURPHY, C
    [J]. AMERICAN JOURNAL OF HOSPITAL PHARMACY, 1993, 50 (01): : 60 - 60
  • [9] EFFECT OF COMPUTER SOFTWARE ON TIME REQUIRED TO PREPARE PARENTERAL NUTRIENT SOLUTIONS
    SEIDEL, AM
    WOLLER, TW
    SOMANI, S
    ABRAMOWITZ, PW
    [J]. AMERICAN JOURNAL OF HOSPITAL PHARMACY, 1991, 48 (02): : 270 - 275
  • [10] SPECIFIC GRAVITY-BASED DETERMINATION OF DEXTROSE CONTENT OF TOTAL PARENTERAL NUTRIENT SOLUTIONS FOR NEONATES
    SILVERBERG, JM
    WEBB, B
    PAWLAK, R
    [J]. AMERICAN JOURNAL OF HOSPITAL PHARMACY, 1993, 50 (10): : 2090 - 2091