Evaluation of subcutaneous glucose monitoring systems under routine environmental conditions in patients with type 1 diabetes

被引:113
作者
Aberer, Felix [1 ]
Hajnsek, Martin [2 ]
Rumpler, Markus [2 ]
Zenz, Sabine [1 ]
Baumann, Petra M. [2 ]
Elsayed, Hesham [1 ]
Puffing, Adelheid [1 ]
Treiber, Gerlies [1 ]
Pieber, Thomas R. [1 ,2 ]
Sourij, Harald [1 ]
Mader, Julia K. [1 ]
机构
[1] Med Univ Graz, Dept Internal Med, Div Endocrinol & Diabetol, Auenbruggerpl 15, A-8036 Graz, Austria
[2] Joanneum Res GmbH, Hlth Inst Biomed & Hlth Sci, Graz, Austria
关键词
continuous glucose monitoring (CGM); glycaemic control; type; 1; diabetes; ACCURACY; SENSOR; PERFORMANCE; MANAGEMENT; PREGNANCY; EXERCISE; IMPROVE; WOMEN;
D O I
10.1111/dom.12907
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Continuous and flash glucose monitoring (GM) systems have been established in diabetes care. We compared the sensor performance of 3 commercially available GM systems. A total of 12 patients with type 1 diabetes were included in a single-centre, open-label study in which the sensor performance of the Abbott FreeStyle libre (Abbott), Dexcom G4 Platinum (Dexcom) and Medtronic MiniMed 640G (Medtronic) systems over 12 hours was compared during mimicked real-life conditions (meals, exercise, hypo-and hyperglycaemia). Sensor performance was determined by fulfilment of ISO 15197: 2013 criteria, calculating mean absolute relative difference (MARD), and was also illustrated using Parkes error grid and Bland-Altman plots. Sensor performance during changes in metabolic variables (lactate, betahydroxybutyrate, glucagon, non-esterified-fatty-acids) was determined by Spearman's rank correlation coefficient testing. The systems fulfilled ISO 15197: 2013 criteria by 73.2% (Abbott), 56.1% (Dexcom) and 52.0% (Medtronic). The MARDs +/- standard deviation in the entire glycaemic range were 13.2% +/- 10.9% (Abbott), 16.8% +/- 12.3% (Dexcom) and 21.4% +/- 17.6% (Medtronic), respectively. All sensors performed less accurately during hypoglycaemia and best during hyperglycaemia. We did not observe an influence of metabolic variables on sensor performance.
引用
收藏
页码:1051 / 1055
页数:5
相关论文
共 21 条
  • [1] [Anonymous], 2014, J DIABETES SCI TECHN
  • [2] The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System
    Bailey, Timothy
    Bode, Bruce W.
    Christiansen, Mark P.
    Klaff, Leslie J.
    Alva, Shridhara
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 (11) : 787 - 794
  • [3] Bailey Timothy S, 2015, J Diabetes Sci Technol, V9, P209, DOI 10.1177/1932296814559746
  • [4] Baysal Nihat, 2014, J Diabetes Sci Technol, V8, P1091, DOI 10.1177/1932296814553267
  • [5] Performance Comparison of the Medtronic Sof-Sensor and Enlite Glucose Sensors in Inpatient Studies of Individuals with Type 1 Diabetes
    Calhoun, Peter
    Lum, John
    Beck, Roy W.
    Kollman, Craig
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2013, 15 (09) : 758 - 761
  • [6] Real-Time Continuous Glucose Monitoring Significantly Reduces Severe Hypoglycemia in Hypoglycemia-Unaware Patients With Type 1 Diabetes
    Choudhary, Pratik
    Ramasamy, Sharmin
    Green, Louisa
    Gallen, Geraldine
    Pender, Siobhan
    Brackenridge, Anna
    Amiel, Stephanie A.
    Pickup, John C.
    [J]. DIABETES CARE, 2013, 36 (12) : 4160 - 4162
  • [7] Effect of the approach to insulin therapy on glycaemic fluctuations and autonomic tone in hospitalized patients with diabetes
    Dungan, K. M.
    Osei, K.
    Sagrilla, C.
    Binkley, P.
    [J]. DIABETES OBESITY & METABOLISM, 2013, 15 (06) : 558 - 563
  • [8] Pumpers, skypers, surfers and texters: technology to improve the management of diabetes in teenagers
    Harris, M. A.
    Hood, K. K.
    Mulvaney, S. A.
    [J]. DIABETES OBESITY & METABOLISM, 2012, 14 (11) : 967 - 972
  • [9] Accuracy of the Enlite 6-Day Glucose Sensor with Guardian and Veo Calibration Algorithms
    Keenan, Desmond Barry
    Mastrototaro, John Joseph
    Zisser, Howard
    Cooper, Kenneth A.
    Raghavendhar, Gautham
    Lee, Scott W.
    Yusi, Jonathan
    Bailey, Timothy S.
    Brazg, Ronald Leonard
    Shah, Rajiv V.
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2012, 14 (03) : 225 - 231
  • [10] Assessing Sensor Accuracy for Non-Adjunct Use of Continuous Glucose Monitoring
    Kovatchev, Boris P.
    Patek, Stephen D.
    Ortiz, Edward Andrew
    Breton, Marc D.
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 (03) : 177 - 186