Intraocular Pressure Measurements in Standing Position with a Rebound Tonometer

被引:14
|
作者
De Bernardo, Maddalena [1 ]
Borrelli, Maria [2 ]
Cembalo, Giovanni [1 ]
Rosa, Nicola [1 ]
机构
[1] Univ Salerno, Dept Med Surg & Dent, I-84081 Baronissi, Italy
[2] Heinrich Heine Univ Dusseldorf, Dept Ophthalmol, D-40204 Dusseldorf, Germany
来源
MEDICINA-LITHUANIA | 2019年 / 55卷 / 10期
关键词
intraocular pressure; Icare PRO; standing position; sitting position; supine position; EPISCLERAL VENOUS-PRESSURE; BODY POSITION; APPLANATION TONOMETRY; GLAUCOMA PROGRESSION; RISK-FACTOR; FLUCTUATION; PNEUMATONOMETER; DIFFERENCE; BLOOD; HEAD;
D O I
10.3390/medicina55100701
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and Objectives: It has been established that body position can play an important role in intraocular pressure (IOP) fluctuation. IOP has been previously shown to increase significantly when lying down, relative to sitting; this type of investigation has not been extensively reported for the standing (ST) position. Therefore, this study aims to look for eventual significant IOP changes while ST, sitting, and lying down. Materials and Methods: An Icare PRO was used to measure the IOP of 120 eyes of 60 healthy individuals, with age ranging from 21 to 55 years (mean 29.22 +/- 9.12 years), in sitting, supine and ST positions; IOP was measured again, 5 min after standing (ST-5m). Results: Mean IOP difference between sitting and ST position was 0.39 +/- 1.93 mmHg (95% CI: 0.04 to 0.74 mmHg) (p = 0.027); between sitting and ST-5m, it was -0.48 +/- 1.79 mmHg (95% CI: -0.8 to -0.16 mmHg) (p = 0.004); between the sitting and supine position, it was -1.16 +/- 1.9 mmHg (95% CI: -1.5 to -0.82 mmHg) (p < 0.001); between the supine and ST position, it was 1.55 +/- 2.04 mmHg (95% CI: 1.18 to 1.92 mmHg) (p < 0.001); between supine and ST-5m, it was 0.68 +/- 1.87 mmHg (95% CI: 0.34 to 1.02 mmHg) (p < 0.001); and between ST-5m and ST, it was 0.94 +/- 1.95 mmHg (95% CI: 0.58 to 1.29 mmHg) (p < 0.001). Mean axial eye length was 24.45 mm (95% CI: 24.22 to 24.69 mm), and mean central corneal thickness was 535.30 mu m (95% CI: 529.44 to 541.19 mu m). Conclusion: Increased IOP in the ST-5m position suggests that IOP measurements should be performed in this position too. The detection of higher IOP values in the ST-5m position than in the sitting one, may explain the presence of glaucoma damage or progression in apparently normal-tension or compensated patients.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Effect of Head and Body Position on Intraocular Pressure
    Malihi, Mehrdad
    Sit, Arthur J.
    OPHTHALMOLOGY, 2012, 119 (05) : 987 - 991
  • [32] Comparing the intraocular pressure values obtained with a rebound tonometer (TONOVET Plus) and an indentation tonometer (IOPvet) in dogs with and without ocular disease
    Michaud, Bertrand
    Lesne, Fabien
    VETERINARY RECORD, 2024, 194 (12) : no
  • [33] Comparison of intraocular pressure measurements with the digital tonometer TGDc-01 'PRA' and the Goldmann applanation tonometer
    Meyer, MW
    Gockeln, R
    Hoy, L
    Meyer, A
    Erb, C
    OPHTHALMIC RESEARCH, 2004, 36 (05) : 250 - 254
  • [34] Intraocular Pressure Values using Icare® Rebound Tonometer and Correlation with Postconceptional Age in Premature Infants
    Monteiro, Catarina
    Vivas, Maria
    Almeida, Julio
    Ramalho, Mario
    Mota, Mafalda
    Teixeira, Susana
    JOURNAL OF OPHTHALMIC & VISION RESEARCH, 2023, 18 (03) : 267 - 271
  • [35] Accuracy of Measurements With the iCare HOME Rebound Tonometer
    Termuehlen, Julia
    Mihailovic, Natasa
    Alnawaiseh, Maged
    Dietlein, Thomas S.
    Rosentreter, Andre
    JOURNAL OF GLAUCOMA, 2016, 25 (06) : 533 - 538
  • [36] Non-invasive determination of intraocular pressure in the rat eye. Comparison of an electronic tonometer (TonoPen), and a rebound (impact probe) tonometer
    David Goldblum
    Antti Kontiola
    Thom Mittag
    Bin Chen
    John Danias
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2002, 240 : 942 - 946
  • [37] Intraocular Pressure Measurement with Pneumatonometry and a Tonometer Tip Cover
    Ferguson, Tanner J.
    Knier, Catherine G.
    Chowdhury, Uttio Roy
    Monson, Kjerseten J.
    Greenwood, Michael
    Swan, Russell J.
    Gorham, Richard
    Berdahl, John P.
    Fautsch, Michael P.
    OPHTHALMOLOGY AND THERAPY, 2020, 9 (01) : 127 - 137
  • [38] A Dynamic Tonometer for Measuring Intraocular Pressure
    Dikin V.I.
    Ivanyschev C.V.
    Kornev N.P.
    Mikheev A.A.
    Solomakha V.N.
    Biomedical Engineering, 2013, 47 (03) : 130 - 133
  • [39] Intraocular pressure measurement using ICare rebound tonometer in different positions of eye and different locations on cornea
    Wongwanwatana, Sirada
    Treesit, Isaraporn
    Funarunart, Panrapee
    Iemsomboon, Wallop
    Choontanom, Raveewan
    MEDICINE, 2023, 102 (36) : E34874
  • [40] Head Position and Intraocular Pressure in the Lateral Decubitus Position
    Seo, Hyejin
    Yoo, Chungkwon
    Lee, Tae-Eun
    Lin, Shan
    Kim, Yong Yeon
    OPTOMETRY AND VISION SCIENCE, 2015, 92 (01) : 95 - 101