Subcutaneous implantable cardioverter-defibrillator: Initial experience

被引:4
|
作者
Galvao, Pedro [1 ]
Cavaco, Diogo [1 ]
Adragao, Pedro [1 ]
Costa, Francisco [1 ]
Carmo, Pedro [1 ]
Morgado, Francisco [1 ]
Bernardo, Ricardo [1 ]
Nunes, Manuela [2 ]
Abecasis, Miguel [3 ]
Neves, Jose [3 ]
Mendes, Miguel [1 ]
机构
[1] Hosp Santa Cruz, Serv Cardiol, Carnaxide, Portugal
[2] Hosp Santa Cruz, Serv Anestesiol, Carnaxide, Portugal
[3] Hosp Santa Cruz, Serv Cirurgia Cardiotorac, Carnaxide, Portugal
关键词
Ventricular tachycardia; Ventricular fibrillation; Implantable cardioverter-defibrillator; INAPPROPRIATE THERAPY; DETECTION ALGORITHMS; PREVENTION; GUIDELINES;
D O I
10.1016/j.repc.2014.01.018
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Implantable cardioverter-defibrillators (ICDs) are important tools in the prevention of sudden death, but implantation requires transvenous access, which is associated with complications. Subcutaneous implantable cardioverter-defibrillators (S-ICDs) may prevent some of these complications. Aim: To evaluate the therapeutics and complications associated with S-ICD systems. Methods: S-ICD implantation was planned in 23 patients, for whom the indications were vascular access problems, increased risk of infection or young patients with long predicted follow-up. The population consisted of four patients with ischemic heart disease, three of them on hemodialysis (two with subclavian vein thrombosis), five with left ventricular noncompaction, four with Brugada syndrome, three with arrhythmogenic right ventricular cardiomyopathy, one with transposition of the great vessels, two with dilated cardiomyopathy and four with hypertrophic cardiomyopathy. Results: S-ICDs were implanted in 21 patients, two having failed to fulfil the initial screening criteria. Mean implantation time was 77 minutes, with no complications. Defibrillation tests were performed, and in one patient the generator had to be repositioned to obtain an acceptable threshold. In a mean follow-up of 14 months, 10 patients had S-ICD shocks, which were appropriate in half of them; one developed infection, one needed early replacement due to loss of telemetry and one patient died of noncardiac cause. Conclusions: S-ICD implantation can be performed by cardiologists with a high success rate. Initial experience appears favorable, but further studies are needed with longer follow-up times to assess the safety and efficacy of this strategy compared to conventional devices. (C) 2013 Sociedade Portuguesa de Cardiologia. Published by Elsevier Espana, S.L.U. All rights reserved.
引用
收藏
页码:511 / 517
页数:7
相关论文
共 50 条
  • [31] Pocket infection as a complication of a subcutaneous implantable cardioverter-defibrillator
    Calvagna, Giuseppe Mario
    Ceresa, Fabrizio
    Patane, Salvatore
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2014, 177 (02) : 616 - 618
  • [32] Shock Without Wires: A Look at Subcutaneous Implantable Cardioverter-Defibrillator Compared to Transvenous Implantable Cardioverter-Defibrillator for Ventricular Arrhythmias
    Raja, Joel
    Guice, Kenneth
    Oberoi, Mansi
    Whitwell, Samantha
    Asbeutah, Abdul Aziz
    Russell, Alex
    Khouzam, Rami N.
    CURRENT PROBLEMS IN CARDIOLOGY, 2022, 47 (09)
  • [33] The implantable cardioverter-defibrillator
    Raj, SR
    Sheldon, RS
    PROGRESS IN CARDIOVASCULAR DISEASES, 2001, 44 (03) : 169 - 194
  • [34] The implantable cardioverter-defibrillator: Background and own experience
    Meininghaus, DG
    Siebels, J
    Langes, K
    Duckeck, W
    HERZ KREISLAUF, 2000, 32 (02): : 54 - 59
  • [36] The implantable cardioverter-defibrillator
    Greene, HL
    CLINICAL CARDIOLOGY, 2000, 23 (05) : 315 - 326
  • [37] CLINICAL-EXPERIENCE WITH THE IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    MIROWSKI, M
    REID, PR
    MOWER, MM
    WATKINS, L
    PLATIA, EV
    GRIFFITH, LSC
    JUANTEGUY, JM
    ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1984, 427 (JUN) : 297 - 306
  • [38] The implantable cardioverter-defibrillator
    Groh, WJ
    Foreman, LD
    Zipes, DP
    AMERICAN FAMILY PHYSICIAN, 1998, 57 (02) : 310 - 312
  • [39] CLINICAL-EXPERIENCE WITH THE IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    WILLIAMS, JM
    ROCK, DT
    PABST, SJ
    GRILL, CR
    DEANTONIO, HJ
    MAHMUD, R
    CHITWOOD, WR
    ANNALS OF THORACIC SURGERY, 1994, 58 (04): : 1297 - 1303
  • [40] The Subcutaneous Implantable Defibrillator: A New Technology That Raises an Existential Question for the Implantable Cardioverter-Defibrillator
    Saxon, Leslie A.
    CIRCULATION, 2013, 128 (09) : 938 - 940