Dosimetric precision of an ion beam tracking system

被引:30
作者
Bert, Christoph [1 ]
Gemmel, Alexander [1 ]
Saito, Nami [1 ]
Chaudhri, Naved [1 ]
Schardt, Dieter [1 ]
Durante, Marco [1 ]
Kraft, Gerhard [1 ]
Rietzel, Eike [1 ,2 ]
机构
[1] GSI Helmholtzzentrum Schwerionenforsch GmbH, Biophys Abt, D-64291 Darmstadt, Germany
[2] Siemens AG, Healthcare Sector, Imaging & Therapy, Particle Therapy, D-91052 Erlangen, Germany
关键词
DOSE OPTIMIZATION; MOVING TARGETS; RADIOTHERAPY; ACCURACY; IRRADIATION; MOTION; THERAPY; VERIFICATION; FEASIBILITY; FIELDS;
D O I
10.1186/1748-717X-5-61
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Scanned ion beam therapy of intra-fractionally moving tumors requires motion mitigation. GSI proposed beam tracking and performed several experimental studies to analyse the dosimetric precision of the system for scanned carbon beams. Methods: A beam tracking system has been developed and integrated in the scanned carbon ion beam therapy unit at GSI. The system adapts pencil beam positions and beam energy according to target motion. Motion compensation performance of the beam tracking system was assessed by measurements with radiographic films, a range telescope, a 3D array of 24 ionization chambers, and cell samples for biological dosimetry. Measurements were performed for stationary detectors and moving detectors using the beam tracking system. Results: All detector systems showed comparable data for a moving setup when using beam tracking and the corresponding stationary setup. Within the target volume the mean relative differences of ionization chamber measurements were 0.3% (1.5% standard deviation, 3.7% maximum). Film responses demonstrated preserved lateral dose gradients. Measurements with the range telescope showed agreement of Bragg peak depth under motion induced range variations. Cell survival experiments showed a mean relative difference of -5% (-3%) between measurements and calculations within the target volume for beam tracking (stationary) measurements. Conclusions: The beam tracking system has been successfully integrated. Full functionality has been validated dosimetrically in experiments with several detector types including biological cell systems.
引用
收藏
页数:9
相关论文
共 28 条
[1]   Accuracy of a wireless localization system for radiotherapy [J].
Balter, JM ;
Wright, JN ;
Newell, LJ ;
Friemel, B ;
Dimmer, S ;
Cheng, Y ;
Wong, J ;
Vertatschitsch, E ;
Mate, TP .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 61 (03) :933-937
[2]   Quantification of interplay effects of scanned particle beams and moving targets [J].
Bert, Christoph ;
Groezinger, Sven O. ;
Rietzel, Eike .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (09) :2253-2265
[3]   4D treatment planning for scanned ion beams [J].
Bert, Christoph ;
Rietzel, Eike .
RADIATION ONCOLOGY, 2007, 2 (1)
[4]   Target motion tracking with a scanned particle beam [J].
Bert, Christoph ;
Saito, Nami ;
Schmidt, Alexander ;
Chaudhri, Naved ;
Schardt, Dieter ;
Rietzel, Eike .
MEDICAL PHYSICS, 2007, 34 (12) :4768-4771
[5]   Ion-optical studies for a range adaptation method in ion beam therapy using a static wedge degrader combined with magnetic beam deflection [J].
Chaudhri, Naved ;
Saito, Nami ;
Bert, Christoph ;
Franczak, Bernhard ;
Steidl, Peter ;
Durante, Marco ;
Rietzel, Eike ;
Schardt, Dieter .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (12) :3499-3513
[6]  
Debus J, 2000, STRAHLENTHER ONKOL, V176, P211, DOI 10.1007/s000660050002
[7]   Accuracy of the local effect model for the prediction of biologic effects of carbon ion beams in vitro and in vivo [J].
Elsaesser, Thilo ;
Kraemer, Michael ;
Scholz, Michael .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (03) :866-872
[8]   Biological dose optimization with multiple ion fields [J].
Gemmel, A. ;
Hasch, B. ;
Ellerbrock, M. ;
Weyrather, W. K. ;
Kraemer, M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (23) :6991-7012
[9]   Motion compensation with a scanned ion beam: a technical feasibility study [J].
Groezinger, Sven Oliver ;
Bert, Christoph ;
Haberer, Thomas ;
Kraft, Gerhard ;
Rietzel, Eike .
RADIATION ONCOLOGY, 2008, 3 (1)
[10]   Is adaptive treatment planning required for stereotactic radiotherapy of stage I non-small-cell lung cancer? [J].
Haasbeek, Cornelis J. A. ;
Lagerwaard, Frank J. ;
Cuijpers, Johan P. ;
Slotman, Ben J. ;
Senan, Suresh .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (05) :1370-1374