Comparison of anterior segment and lens biometric measurements in patients with cataract

被引:47
作者
Fukuda, Shinichi [1 ]
Ueno, Yuta [1 ]
Fujita, Akari [1 ]
Mori, Haruhiro [1 ]
Tasaki, Kuniharu [1 ]
Murakami, Tomoya [1 ]
Beheregaray, Simone [2 ]
Oshika, Tetsuro [1 ]
机构
[1] Univ Tsukuba, Fac Med, Dept Ophthalmol, 1-1-1 Tennoudai, Tsukuba, Ibaraki 3058575, Japan
[2] Queen Elizabeth Hosp, Dept Ophthalmol, Adelaide, SA, Australia
基金
日本学术振兴会;
关键词
Deep-range swept-source anterior segment optical coherence tomography; CASIA2; Lens biometry; Anterior segment biometry; Cataract; OPTICAL COHERENCE TOMOGRAPHY; CHAMBER DEPTH MEASUREMENTS; AGING HUMAN LENS; CRYSTALLINE LENS; CORNEAL THICKNESS; ACCOMMODATION; REPRODUCIBILITY; REPEATABILITY; CURVATURE; EYE;
D O I
10.1007/s00417-019-04482-0
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose This study was performed to compare anterior ocular biometric measurements of deep-range swept-source anterior segment optical coherence tomography (AS-OCT) (CASIA2) versus short-range swept-source AS-OCT (CASIA SS-1000), ultrasonography (AL-4000), and Scheimpflug camera analysis (Pentacam and EAS-1000) in patients with cataract. Methods One hundred eighty-five eyes of 128 participants with mild refractive error or cataract were examined. The central corneal thickness (CCT), aqueous depth (AQD), and lens thickness were obtained. The repeatability of CASIA2 measurements was assessed. Results In patients with cataract, the CCT, AQD, lens thickness, and lens anterior curvature by CASIA2 showed high intraclass correlation coefficients (ICCs) of > 0.99. Conversely, measurements of the posterior part of the lens such as lens posterior curvature showed lower ICCs. The ICCs were higher in healthy young participants than in patients with cataract. The ICCs tended to be lower in patients with mild than dense cataract. There was no statistically significant difference in the CCT and AQD between the CASIA2 and CASIA SS-1000 or in the lens thickness measurements between the CASIA2 and AL-4000 and between the CASIA2 and EAS-1000. There was a significant linear correlation in the biometric measurements between the CASIA2 and the other instruments. Conclusion We evaluated the biometric measurements of the anterior eye segment by the CASIA2. The CASIA2 yielded results comparable with those of the CASIA SS-1000, ultrasonography, and Scheimpflug camera. However, mild cataract decreased the repeatability of measurements of the posterior part of the lens.
引用
收藏
页码:137 / 146
页数:10
相关论文
共 44 条
[1]   Measurement of the internal diameter and depth of the anterior chamber: IOLMaster versus anterior chamber optical coherence tomographer [J].
Baikoff, G ;
Jodai, HJ ;
Bourgeon, G .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (09) :1722-1728
[2]   COMPARING METHODS OF MEASUREMENT - WHY PLOTTING DIFFERENCE AGAINST STANDARD METHOD IS MISLEADING [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1995, 346 (8982) :1085-1087
[3]   CHANGE IN LENS CURVATURE WITH AGE [J].
BROWN, N .
EXPERIMENTAL EYE RESEARCH, 1974, 19 (02) :175-183
[4]   A simple description of age-related changes in crystalline lens thickness [J].
Carmen Garcia-Domene, Mari ;
Amparo Diez-Ajenjo, Ma ;
Gracia, Victor ;
Felipe, Adelina ;
Artigas, Jose M. .
EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2011, 21 (05) :597-603
[5]   Agreement of Anterior Segment Parameters Obtained From Swept-Source Fourier-Domain and Time-Domain Anterior Segment Optical Coherence Tomography [J].
Chansangpetch, Sunee ;
Nguyen, Anwell ;
Mora, Marta ;
Badr, Mai ;
He, Mingguang ;
Porco, Travis C. ;
Lin, Shan C. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2018, 59 (03) :1554-1561
[6]  
Devereux JG, 2000, ARCH OPHTHALMOL-CHIC, V118, P257
[7]   Human corneal thickness and its impact on intraocular pressure measures: A review and meta-analysis approach [J].
Doughty, MJ ;
Zaman, ML .
SURVEY OF OPHTHALMOLOGY, 2000, 44 (05) :367-408
[8]   The thickness of the aging human lens obtained from corrected scheimpflug images [J].
Dubbelman, M ;
Van der Heijde, GL ;
Weeber, HA .
OPTOMETRY AND VISION SCIENCE, 2001, 78 (06) :411-416
[9]   The shape of the aging human lens: curvature, equivalent refractive index and the lens paradox [J].
Dubbelman, M ;
Van der Heijde, GL .
VISION RESEARCH, 2001, 41 (14) :1867-1877
[10]   Comparative study of anterior segment measurement with Pentacam and anterior segment optical coherence tomography [J].
Fu, Jing ;
Wang, Xiaozhen ;
Li, Shuning ;
Wu, Gewei ;
Wang, Ningli .
CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 2010, 45 (06) :627-631