摘要
AIM: To evaluate intraocular pressure (IOP) measurements and fluctuations using the iCare ONE rebound tonometer (RT-ONE), during home monitoring, in diagnosed and suspected glaucoma patients. METHODS: A retrospective case series of consecutive patients with known glaucoma or glaucoma suspects who were followed-up and treated between January 2016 and January 2017. The study included 80 eyes of 40 patients with a mean age of 59.1 +/- 14.6y (range, 24-78). All patients have undergone 4-5d of IOP home monitoring with RT-ONE at morning, noon, afternoon, and night time. RESULTS: Baseline mean IOP, as measured in the clinic (8 a.m.-12 p.m.), was 17.4 +/- 5.1 mm Hg, compared to RT-ONE home monitoring mean 10P of 15.6 +/- 4.1 mm Hg (P=0.002). Mean IOP was significantly lower at noon, afternoon and night times compared to clinic measured IOP and morning measurements (P=0.005). IOP peak measured during home monitoring was significantly higher compared to the clinic measured IOP (21.3 +/- 5.6 mm Hg and 17.4 +/- 5.1 mm Hg, P<0.001). IOP peaks during home monitoring demonstrated a majority of 47 peaks during morning measurements, compared to 23 at noon, 19 at afternoon and only 12 at night (P<0.001). The home monitoring results led to treatment modification of 44 eyes (55%), treatment regime was insufficient for 40 (50%) eyes. CONCLUSION: Home monitoring IOP with RT-ONE can provide good assessment of mean IOP, IOP fluctuations and peaks throughout the hours of the day, which lead to an accurate treatment for glaucoma patients.
引用
收藏
页码:405 / 408
页数:4
相关论文
共 25 条
[1]   Clinical utility of Intraocular pressure monitoring outside of normal office hours in patients with glaucoma [J].
Barkana, Yaniv ;
Anis, Sarah ;
Liebmann, Jeffrey ;
Tello, Celso ;
Ritch, Robert .
ARCHIVES OF OPHTHALMOLOGY, 2006, 124 (06) :793-797
[2]   Variability of Intraocular Pressure Measurements in Observation Participants in the Ocular Hypertension Treatment Study [J].
Bhorade, Anjali M. ;
Gordon, Mae O. ;
Wilson, Brad ;
Weinrab, Robert N. ;
Kass, Michael A. .
OPHTHALMOLOGY, 2009, 116 (04) :717-724
[3]   Rebound tonometry: new opportunities and limitations of non-invasive determination of intraocular pressure [J].
Cervino, A. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2006, 90 (12) :1444-1446
[4]   Method for the noninvasive measurement of intraocular pressure in mice [J].
Danias, J ;
Kontiola, AI ;
Filippopoulos, T ;
Mittag, T .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (03) :1138-1141
[5]   DIURNAL INTRAOCULAR-PRESSURE VARIATIONS - AN ANALYSIS OF 690 DIURNAL CURVES [J].
DAVID, R ;
ZANGWILL, L ;
BRISCOE, D ;
DAGAN, M ;
YAGEV, R ;
YASSUR, Y .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1992, 76 (05) :280-283
[6]   Diurnal Intraocular Pressure Fluctuations with Self-tonometry in Glaucoma Patients and Suspects: A Clinical Trial [J].
Huang, Jessie ;
Katalinic, Paula ;
Kalloniatis, Michael ;
Hennessy, Michael P. ;
Zangerl, Barbara .
OPTOMETRY AND VISION SCIENCE, 2018, 95 (02) :88-95
[7]   24-hour monitoring of intraocular pressure in glaucoma management: A retrospective review [J].
Hughes, E ;
Spry, P ;
Diamond, J .
JOURNAL OF GLAUCOMA, 2003, 12 (03) :232-236
[8]  
Kass MA, 2002, ARCH OPHTHALMOL-CHIC, V120, P701
[9]   Asymmetry of diurnal intraocular pressure fluctuation between right and left eyes [J].
Kim, Mo Sae ;
Kim, Joon Mo ;
Park, Ki Ho ;
Choi, Chul Young .
ACTA OPHTHALMOLOGICA, 2011, 89 (04) :352-357
[10]   DIURNAL-VARIATION OF INTRAOCULAR-PRESSURE IN PRIMARY OPEN-ANGLE GLAUCOMA [J].
KITAZAWA, Y ;
HORIE, T .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1975, 79 (04) :557-566