Piezoelectric Electrocardiographic Artifact in a Patient after Surgery with Bradycardia and Hypotension

被引:2
作者
Cheng, Richard [1 ]
Chakravarty, Tarun [1 ]
机构
[1] Cedars Sinai Heart Inst, Los Angeles, CA USA
关键词
equipment-related ECG artifact; piezoelectric; bradycardia; UNUSUAL ECG ARTIFACT; INFUSION-PUMP;
D O I
10.1111/anec.12154
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A 64-year-old man was evaluated by a rapid response team for altered mental status shortly after an uncomplicated surgery. He was found to be hypotensive and bradycardic, and an emergent electrocardiogram showed extra "P" wave complexes, ultimately found to be piezoelectric artifacts from a fluid infusion pump. Equipment-related artifacts have been known to mimic arrhythmias prompting unnecessary therapeutic interventions including antiarrhythmics and direct current shocks. Timely recognition of the unusual properties of the complexes resulted in the avoidance of atropine, epinephrine, or transcutaneous pacing in a rapid response team scenario.
引用
收藏
页码:598 / 600
页数:3
相关论文
共 9 条
[1]  
AGARWAL SK, 1980, HEART LUNG, V9, P1063
[2]   PSEUDO-ARRHYTHMIA DUE TO DEFECTIVE INFUSION PUMP AND ECG MONITOR [J].
CROKE, RP ;
BULCHANDANI, KV ;
JACOBS, WR ;
LOEB, HS .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1976, 235 (07) :705-706
[3]  
GRAHAM MM, 1981, J NUCL MED, V22, P660
[4]   ECG INFUSION ARTIFACT [J].
GROEGER, JS ;
MIODOWNIK, S ;
HOWLAND, WS .
CHEST, 1984, 85 (01) :143-143
[5]  
MEHARG JG, 1979, NEW ENGL J MED, V301, P165
[6]   Electrocardiographic artifact mimicking ventricular tachycardia during high-frequency oscillatory ventilation: A case report [J].
Patel, Santosh .
AMERICAN JOURNAL OF CRITICAL CARE, 2006, 15 (03) :310-311
[7]   ECG ARTIFACT PRODUCED BY CRYSTALLOID ADMINISTRATION THROUGH BLOOD FLUID WARMING SETS [J].
PAULSEN, AW ;
PRITCHARD, DG .
ANESTHESIOLOGY, 1988, 69 (05) :803-804
[8]   ELECTRICAL CURRENT LEAKAGE TRANSMITTED TO AN INFANT VIA AN IV CONTROLLER - UNUSUAL ECG ARTIFACT [J].
SAHN, DJ ;
VAUCHER, YE .
JOURNAL OF PEDIATRICS, 1976, 89 (02) :301-302
[9]   AN UNUSUAL CAUSE OF ELECTROCARDIOGRAM PACEMAKER ARTIFACT - FAULTY ELECTRICAL GROUNDING OF EQUIPMENT [J].
TRAVIS, KW ;
ROSE, RJ ;
QUILL, TJM ;
MCCLENACHAN, A .
ANESTHESIOLOGY, 1994, 80 (01) :232-232