The effect of acoustic velocity on phantom measurements

被引:38
作者
Goldstein, L [1 ]
机构
[1] Wayne State Univ, Sch Med, Dept Radiol, Detroit, MI USA
关键词
acoustic velocity; ultrasound phantom measurements; urethane rubber; ultrasound quality assurance;
D O I
10.1016/S0301-5629(00)00248-9
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Urethane rubber ultrasound (US) phantoms have a much lower acoustic velocity (1430-1450 m/s) than the accepted soft tissue average of 1540 m/s, Two important questions arise: can the rod positions in these rubber phantoms be adjusted so that they may be used to test equipment distance measurement accuracy for all types of multielement transducers, and can they be used to measure beam focus (using the spread of the rod blur patterns)? These questions were addressed for linear-, phased-, convex- and vector-array transducers. Theoretical predictions for the different transducers' distance measurement errors agreed with careful measurements obtained with a specially designed array of stainless-steel rods immersed in paraffin oil (1447 m/s), The conclusions of this study are that phantoms with acoustic velocities different from 1540 mis cannot be used to check distance measurement accuracies of all the types of real-time transducers, nor to predict a transducer's focusing performance in clinical scans. (C) 2000 World Federation for Ultrasound in Medicine & Biology.