Validation of an analytical method to determine sulfamides in kidney by HPLC-DAD and PARAFAC2 with first-order derivative chromatograms

被引:39
作者
Garcia, I.
Ortiz, M. C.
Sarabia, L.
Aldama, J. M.
机构
[1] Fac Sci, Dept Chem, Burgos 09001, Spain
[2] Fac Sci, Dept Math & Computat, Burgos 09001, Spain
[3] Inst Ciencias Salud Anal Fis Quim, Talavera De La Reina, Toledo, Spain
关键词
PARAFAC2; sulfonamides; first-order derivative; three-way models; baseline drift; decision limit CC alpha; capability of detection CC beta; decision; 2002/657/EC;
D O I
10.1016/j.aca.2007.01.054
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Six sulfamides were extracted from kidney and analysed by high-performance liquid chromatography with diode array detection (HPLC-DAD): sulfadiazine, sulfamethazine, sulfamethoxypyridazine, sulfamethoxazole, sulfadimethoxine and sulfaquinoxaline. Two main difficulties arose in identifying and quantifying the analytes. Firstly, the chromatographic peaks of the matrix interferences overlapped with those of the analytes. The uniqueness property of PARAFAC2 solved this problem. Secondly, the gradient elution caused a baseline drift. The first-order derivative of the chromatograms minimized its effect. The analytical method was validated. As the performance criteria detailed in the European Decision 2002/657/EC are based on specific signals, this paper generalizes those criteria for higher-order and non-specific signals. In this sense the proposed methodology is general and can be applied to any chromatographic method (HPLC or GC) with a detector that provide a multivariate signal (MS, DAD, EC, etc.). (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:222 / 234
页数:13
相关论文
共 36 条
[11]   METHODS FOR ASSAY OF TRIMETHOPRIM AND SULFADIAZINE IN BROILER TISSUES USING LIQUID-CHROMATOGRAPHY [J].
DAGORN, M ;
DELMAS, JM .
ANALYTICA CHIMICA ACTA, 1994, 285 (03) :353-358
[12]   Comparison of porcine urine and bile as matrices to screen for the residues of two sulfonamides using a semi-automated enzyme immunoassay [J].
Fodey, TL ;
Crooks, SRH ;
Elliott, CT ;
McCaughey, WJ .
ANALYST, 1997, 122 (02) :165-168
[13]   Robustness of the extraction step when parallel factor analysis (PARAFAC) is used to quantify sulfonamides in kidney by high performance liquid chromatography-diode array detection (HPLC-DAD) [J].
García, I ;
Sarabia, L ;
Ortiz, MC ;
Aldama, JM .
ANALYST, 2004, 129 (08) :766-771
[14]   Detection capability of tetracyclines analysed by a fluorescence technique:: comparison between bilinear and trilinear partial least squares models [J].
García, I ;
Sarabia, LA ;
Ortiz, MC .
ANALYTICA CHIMICA ACTA, 2004, 501 (02) :193-203
[15]  
GARCIA I, 2005, THESIS U BURGOS SPAI
[16]  
Gehring TA, 1997, J AOAC INT, V80, P751
[17]  
HARSHMAN RA, 1984, RES METHODS MULTIMOD, pCH6
[18]  
HARSHMAN RA, 1972, UCLA WORKING PAPERS, V22, P33
[19]  
Inczedy J., 1998, IUPAC Compendium of Analytical Nomenclature, V3rd
[20]  
International Organisation for Standardisation, 1994, 5725 ISO 1