Prospectively ECG-triggered high-pitch coronary CT angiography at 70 kVp with 30 mL contrast agent: An intraindividual comparison with sequential scanning at 120 kVp with 60 mL contrast agent

被引:22
作者
Wang, Wei [1 ]
Zhao, Yan E. [1 ]
Qi, Li [1 ]
Li, Xie [1 ]
Zhou, Chang Sheng [1 ]
Zhang, Long Jiang [1 ]
Lu, Guang Ming [1 ]
机构
[1] Nanjing Univ, Sch Med, Jinling Hosp, Dept Med Imaging, Nanjing 210002, Jiangsu, Peoples R China
关键词
Coronary CT angiography; Low tube voltage; Low contrast medium; High-pitch acquisition; Radiation dose; Diagnostic accuracy; COMPUTED-TOMOGRAPHY ANGIOGRAPHY; LOW TUBE VOLTAGE; IMAGE QUALITY; DIAGNOSTIC-ACCURACY; ARTERY-DISEASE; ITERATIVE RECONSTRUCTION; RADIATION-EXPOSURE; SPIRAL ACQUISITION; FEASIBILITY; OPTIMIZATION;
D O I
10.1016/j.ejrad.2017.02.020
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate image quality, radiation dose, and diagnostic efficiency of prospectively ECG-triggered high-pitch coronary CT angiography (CCTA) at 70 kVp with 30 mL contrast agent intra-individually compared with routine CCTA protocol. Materials and methods: One hundred and thirty eight patients with suspected coronary artery disease, body mass index (BMI) <= 25 kg/m(2) and heart rate (HR) <= 70 beats per minute (bpm) underwent prospectively ECG-triggered high-pitch CCTA at 70 kVp and 30 mL contrast agent (protocol A) and prospectively ECG-triggered sequential scanning at 120 kVp and 60 mL contrast medium (protocol B). Objective and subjective image quality, radiation doses, and diagnostic accuracy were evaluated and compared between the two protocols. Results: Higher CT attenuation, higher noise, lower signal-to-noise ratios (SNRs) and lower contrast-to noise ratios (CNRs) were found in protocol A than in protocol B (P<0.001). However, image quality of protocol A were diagnostic. In patients with BMI < 23 kg/m(2) or HR< 60 bpm, subjective image quality scores of some coronary arteries in protocol A were not significantly different from protocol B (P> 0.05). Effective dose in protocol A has reduced by 96.7% compared with protocol B (P<0.001). No significant differences were found for diagnostic accuracy between the two protocols on a per-segment (P=0.513), per-vessel (P=0.317) and per-patient (P=0.125) basis. Conclusions: Prospectively ECG-triggered high-pitch CCTA at 70 kVp with 30 mL contrast agent can reduce radiation dose but maintain image quality and high diagnostic accuracy in a selected, non-obese population. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:97 / 105
页数:9
相关论文
共 30 条
  • [21] Low contrast- and low radiation dose protocol for cardiac CT of thin adults at 256-row CT: usefulness of low tube voltage scans and the hybrid iterative reconstruction algorithm
    Nakaura, Takeshi
    Kidoh, Masafumi
    Sakaino, Naritsugu
    Utsunomiya, Daisuke
    Oda, Seitaro
    Kawahara, Tetsuya
    Harada, Kazunori
    Yamashita, Yasuyuki
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2013, 29 (04) : 913 - 923
  • [22] Coronary Computed Tomographic Angiography at Low Concentration of Contrast Agent and Low Tube Voltage in Patients with Obesity: A Feasibility Study
    Pan, Yu-Ning
    Li, Ai-Jing
    Chen, Xiao-Min
    Wang, Jian
    Ren, Da -Wei
    Huang, Qiu-Li
    [J]. ACADEMIC RADIOLOGY, 2016, 23 (04) : 438 - 445
  • [23] PROMISE of Coronary CT Angiography: Precise and Accurate Diagnosis and Prognosis in Coronary Artery Disease
    Thomas, Dustin M.
    Branch, Kelley R.
    Cury, Ricardo C.
    [J]. SOUTHERN MEDICAL JOURNAL, 2016, 109 (04) : 242 - 247
  • [24] Image quality and dose performance of 80 kV low dose scan protocol in high-pitch spiral coronary CT angiography: feasibility study
    Wang, Dan
    Hu, Xiu-hua
    Zhang, Shi-zheng
    Wu, Run-ze
    Xie, Su-su
    Chen, Bin
    Zhang, Qiao-wei
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2012, 28 (02) : 415 - 423
  • [25] Contrast media volume optimization in high-pitch dual-source CT coronary angiography: feasibility study
    Yang, Wen Jie
    Chen, Ke Min
    Liu, Bo
    Pang, Li Fang
    Zhang, Huan
    Pan, Zi Lai
    Yan, Fu Hua
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2013, 29 (01) : 245 - 252
  • [26] REPLY: The Ethics of Publishing Dual Exposure Scans Involving Ionizing Radiation When Validated Alternatives Exist
    Yin, Wei-Hua
    Lu, Bin
    Meinel, Felix G.
    Schoepf, U. Joseph
    [J]. JACC-CARDIOVASCULAR IMAGING, 2014, 7 (09) : 964 - 965
  • [27] Iterative Reconstruction to Preserve Image Quality and Diagnostic Accuracy at Reduced Radiation Dose in Coronary CT Angiography An Intraindividual Comparison
    Yin, Wei-Hua
    Lu, Bin
    Li, Nan
    Han, Lei
    Hou, Zhi-Hui
    Wu, Run-Ze
    Wu, Yong-Jian
    Niu, Hong-Xia
    Jiang, Shi-Liang
    Krazinski, Aleksander W.
    Ebersberger, Ullrich
    Meinel, Felix G.
    Schoepf, U. Joseph
    [J]. JACC-CARDIOVASCULAR IMAGING, 2013, 6 (12) : 1239 - 1249
  • [28] Image quality, radiation dose, and diagnostic accuracy of prospectively ECG-triggered high-pitch coronary CT angiography at 70 kVp in a clinical setting: comparison with invasive coronary angiography
    Zhang, Long Jiang
    Wang, Yining
    Schoepf, U. Joseph
    Meinel, Felix G.
    Bayer, Richard R., II
    Qi, Li
    Cao, Jian
    Zhou, Chang Sheng
    Zhao, Yan E.
    Li, Xie
    Bin Gong, Jian
    Jin, Zhengyu
    Lu, Guang Ming
    [J]. EUROPEAN RADIOLOGY, 2016, 26 (03) : 797 - 806
  • [29] High-Pitch Coronary CT Angiography at 70 kVp With Low Contrast Medium Volume Comparison of 80 and 100 kVp High-Pitch Protocols
    Zhang, Long Jiang
    Qi, Li
    De Cecco, Carlo N.
    Zhou, Chang Sheng
    Spearman, James V.
    Schoepf, U. Joseph
    Lu, Guang Ming
    [J]. MEDICINE, 2014, 93 (22)
  • [30] Feasibility of prospectively ECG-triggered high-pitch coronary CT angiography with 30 mL iodinated contrast agent at 70 kVp: initial experience
    Zhang, Long Jiang
    Qi, Li
    Wang, Jing
    Tang, Chun Xiang
    Zhou, Chang Sheng
    Ji, Xue Man
    Spearman, James V.
    De Cecco, Carlo Nicola
    Meinel, Felix G.
    Schoepf, U. Joseph
    Lu, Guang Ming
    [J]. EUROPEAN RADIOLOGY, 2014, 24 (07) : 1537 - 1546