Prospectively ECG-triggered high-pitch coronary CT angiography at 70 kVp with 30 mL contrast agent: An intraindividual comparison with sequential scanning at 120 kVp with 60 mL contrast agent

被引:22
作者
Wang, Wei [1 ]
Zhao, Yan E. [1 ]
Qi, Li [1 ]
Li, Xie [1 ]
Zhou, Chang Sheng [1 ]
Zhang, Long Jiang [1 ]
Lu, Guang Ming [1 ]
机构
[1] Nanjing Univ, Sch Med, Jinling Hosp, Dept Med Imaging, Nanjing 210002, Jiangsu, Peoples R China
关键词
Coronary CT angiography; Low tube voltage; Low contrast medium; High-pitch acquisition; Radiation dose; Diagnostic accuracy; COMPUTED-TOMOGRAPHY ANGIOGRAPHY; LOW TUBE VOLTAGE; IMAGE QUALITY; DIAGNOSTIC-ACCURACY; ARTERY-DISEASE; ITERATIVE RECONSTRUCTION; RADIATION-EXPOSURE; SPIRAL ACQUISITION; FEASIBILITY; OPTIMIZATION;
D O I
10.1016/j.ejrad.2017.02.020
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate image quality, radiation dose, and diagnostic efficiency of prospectively ECG-triggered high-pitch coronary CT angiography (CCTA) at 70 kVp with 30 mL contrast agent intra-individually compared with routine CCTA protocol. Materials and methods: One hundred and thirty eight patients with suspected coronary artery disease, body mass index (BMI) <= 25 kg/m(2) and heart rate (HR) <= 70 beats per minute (bpm) underwent prospectively ECG-triggered high-pitch CCTA at 70 kVp and 30 mL contrast agent (protocol A) and prospectively ECG-triggered sequential scanning at 120 kVp and 60 mL contrast medium (protocol B). Objective and subjective image quality, radiation doses, and diagnostic accuracy were evaluated and compared between the two protocols. Results: Higher CT attenuation, higher noise, lower signal-to-noise ratios (SNRs) and lower contrast-to noise ratios (CNRs) were found in protocol A than in protocol B (P<0.001). However, image quality of protocol A were diagnostic. In patients with BMI < 23 kg/m(2) or HR< 60 bpm, subjective image quality scores of some coronary arteries in protocol A were not significantly different from protocol B (P> 0.05). Effective dose in protocol A has reduced by 96.7% compared with protocol B (P<0.001). No significant differences were found for diagnostic accuracy between the two protocols on a per-segment (P=0.513), per-vessel (P=0.317) and per-patient (P=0.125) basis. Conclusions: Prospectively ECG-triggered high-pitch CCTA at 70 kVp with 30 mL contrast agent can reduce radiation dose but maintain image quality and high diagnostic accuracy in a selected, non-obese population. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:97 / 105
页数:9
相关论文
共 30 条
  • [1] Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition
    Achenbach, Stephan
    Marwan, Mohamed
    Ropers, Dieter
    Schepis, Tiziano
    Pflederer, Tobias
    Anders, Katharina
    Kuettner, Axel
    Daniel, Werner G.
    Uder, Michael
    Lell, Michael M.
    [J]. EUROPEAN HEART JOURNAL, 2010, 31 (03) : 340 - 346
  • [2] Low-dose, 128-slice, dual-source CT coronary angiography: accuracy and radiation dose of the high-pitch and the step-and-shoot mode
    Alkadhi, Hatem
    Stolzmann, Paul
    Desbiolles, Lotus
    Baumueller, Stephan
    Goetti, Robert
    Plass, Andre
    Scheffel, Hans
    Feuchtner, Gudrun
    Falk, Volkmar
    Marincek, Borut
    Leschka, Sebastian
    [J]. HEART, 2010, 96 (12) : 933 - 938
  • [3] American Association of Physicists in Medicine AAPM, 2011, AAPM report 204, P1
  • [4] Austen WG, 1975, CIRCULATION, V51, pS5
  • [5] Investigation of American Association of Physicists in Medicine Report 204 Size-specific Dose Estimates for Pediatric CT Implementation
    Brady, Samuel L.
    Kaufman, Robert A.
    [J]. RADIOLOGY, 2012, 265 (03) : 832 - 840
  • [6] CT coronary angiography at an ultra-low radiation dose (&lt;0.1 mSv): feasible and viable in times of constraint on healthcare costs
    Cademartiri, Filippo
    Maffei, Erica
    Arcadi, Teresa
    Catalano, Onofrio
    Midiri, Massimo
    [J]. EUROPEAN RADIOLOGY, 2013, 23 (03) : 607 - 613
  • [7] Radiation and contrast agent doses reductions by using 80-kV tube voltage in coronary computed tomographic angiography: A comparative study
    Cao, Jian-xin
    Wang, Yi-min
    Lu, Jin-guo
    Zhang, Yu
    Wang, Peng
    Yang, Cheng
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2014, 83 (02) : 309 - 314
  • [8] Submillisievert Median Radiation Dose for Coronary Angiography with a Second-Generation 320-Detector Row CT Scanner in 107 Consecutive Patients
    Chen, Marcus Y.
    Shanbhag, Sujata M.
    Arai, Andrew E.
    [J]. RADIOLOGY, 2013, 267 (01) : 76 - 85
  • [9] SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT
    Halliburton, Sandra S.
    Abbara, Suhny
    Chen, Marcus Y.
    Gentry, Ralph
    Mahesh, Mahadevappa
    Raff, Gilbert L.
    Shaw, Leslee J.
    Hausleiter, Joerg
    [J]. JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2011, 5 (04) : 198 - 224
  • [10] Image Quality and Radiation Exposure With Prospectively ECG-Triggered Axial Scanning for Coronary CT Angiography The Multicenter, Multivendor, Randomized PROTECTION-III Study
    Hausleiter, Jorg
    Meyer, Tanja S.
    Martuscelli, Eugenio
    Spagnolo, Pietro
    Yamamoto, Hiroaki
    Carrascosa, Patricia
    Anger, Thomas
    Lehmkuhl, Lukas
    Alkadhi, Hatem
    Martinoff, Stefan
    Hadamitzky, Martin
    Hein, Franziska
    Bischoff, Bernhard
    Kuse, Miriam
    Schoemig, Albert
    Achenbach, Stephan
    [J]. JACC-CARDIOVASCULAR IMAGING, 2012, 5 (05) : 484 - 493