An experimental study on real-time analysis of two-phase peristaltic slug flows in dialysis machines

被引:6
作者
Mirzaei, Mohsen [1 ]
Parrany, Ahmad Mahdian [1 ]
机构
[1] Vali E Asr Univ Rafsanjan, Dept Mech Engn, Rafsanjan, Iran
关键词
Two-phase slug flow; Peristaltic flow; Pulsatile flow; Measurement technique; Image processing; Hemodialysis; VERTICAL PIPES; REGIME; HYDRODYNAMICS;
D O I
10.1016/j.flowmeasinst.2021.101941
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Two-phase flows appear in many industrial and biomedical applications. One of the most vital biomedical applications of two-phase flows is in hemodialysis machines due to air embolism and heparin injection. Since these flows have a very complex and intermittent nature, studying their dynamics is a very challenging and fundamental problem. The purpose of this article is to present an experimental study on the dynamics of two-phase peristaltic slug flows. The measurement strategy is based on the image processing technology. The characteristic parameters of the two-phase pulsatile slug flows, including the slug length, as well as the translational velocity and frequency of the slug motion, are measured, and the effect of the liquid flow rate and liquid superficial velocity is investigated. The results show that the average and maximum slug velocities, and also the dominant amplitude of the slug velocity increase with the flow rate and liquid superficial velocity, while it is not possible to clearly predict a correlation between the liquid superficial velocity and the slug length. The measurement strategy presented in this article can be used in the control and alarm systems of smart dialysis machines.
引用
收藏
页数:9
相关论文
共 23 条
  • [1] Dynamics of a Taylor bubble in steady and pulsatile co-current flow of Newtonian and shear-thinning liquids in a vertical tube
    Abishek, S.
    King, A. J. C.
    Narayanaswamy, R.
    [J]. INTERNATIONAL JOURNAL OF MULTIPHASE FLOW, 2015, 74 : 148 - 164
  • [2] [Anonymous], 2011, DIGITAL IMAGE PROCES
  • [3] Azar A.T., 2013, MODELING CONTROL DIA, V1
  • [4] Dias R., 2012, SINGLE 2 PHASE FLOWS
  • [5] Image processing techniques for high-speed videometry in horizontal two-phase slug flows
    do Amaral, C. E. F.
    Alves, R. F.
    da Silva, M. J.
    Arruda, L. V. R.
    Dorini, L.
    Morales, R. E. M.
    Pipa, D. R.
    [J]. FLOW MEASUREMENT AND INSTRUMENTATION, 2013, 33 : 257 - 264
  • [6] FABRE J, 1992, ANNU REV FLUID MECH, V24, P21, DOI 10.1146/annurev.fl.24.010192.000321
  • [7] Ghiaasiaan SM, 2008, TWO-PHASE FLOW, BOILING AND CONDENSATION: IN CONVENTIONAL AND MINIATURE SYSTEMS, P1
  • [8] High resolution gamma ray tomography scanner for flow measurement and non-destructive testing applications
    Hampel, U.
    Bieberle, A.
    Hoppe, D.
    Kronenberg, J.
    Schleicher, E.
    Suehnel, T.
    Zimmermann, F.
    Zippe, C.
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2007, 78 (10)
  • [9] Measurement of the liquid film thickness in micro tube slug flow
    Han, Youngbae
    Shikazono, Naoki
    [J]. INTERNATIONAL JOURNAL OF HEAT AND FLUID FLOW, 2009, 30 (05) : 842 - 853
  • [10] Development and validation of an X-ray tomograph for two-phase flow
    Hervieu, E
    Jouet, E
    Desbat, L
    [J]. VISUALIZATION AND IMAGING IN TRANSPORT PHENOMENA, 2002, 972 : 87 - 94