Knee simulator wear of polyethylene tibias articulating against explanted rough femoral components

被引:38
作者
Muratoglu, OK [1 ]
Burroughs, BR [1 ]
Bragdon, CR [1 ]
Christensen, S [1 ]
Lozynsky, A [1 ]
Harris, WH [1 ]
机构
[1] Massachusetts Gen Hosp, Orthopaed Biomech & Biomat Lab, Boston, MA 02114 USA
关键词
D O I
10.1097/01.blo.0000143801.41885.8b
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Highly cross-linked and melted polyethylene tibial inserts have recently been introduced for clinical use to reduce fatigue damage and adhesive wear in tibial inserts. Other authors have studied the effect of counterface roughness on the wear behavior of polyethylene tibial inserts in knee simulators using femoral components that were roughened artificially. They reported a higher wear rate with highly crosslinked polyethylene than with unirradiated polyethylene tibial inserts. Artificial roughening of femoral components may not be clinically relevant. To evaluate this concern, we studied the wear behavior of highly cross-linked and conventional polyethylene tibial inserts articulating in vitro against surgically retrieved femoral components that had become roughened in vivo. The wear rate of the highly cross-linked polyethylene (5.9 and 6.8 mg/1 million cycles with 100 and 50% serum) was 80% lower than the wear rate of the conventional polyethylene (33.5 and 32.2 mg/1 million cycles with 100 and 50% serum) tibial inserts after 2 million cycles of simulated gait. This study suggests that during in vivo use, scratches that are generated on the femoral components are likely to produce a higher wear rate with both cross-linked and conventional polyethylene than a smooth femoral component, but that this wear rate is likely to be higher with conventional polyethylene than with highly cross-linked polyethylene tibial inserts.
引用
收藏
页码:108 / 113
页数:6
相关论文
共 13 条
[1]  
*ASTM, 2002, 171496 ASTM F
[2]   Third-body wear of highly cross-linked polyethylene in a hip simulator [J].
Bragdon, CR ;
Jasty, M ;
Muratoglu, OK ;
O'Connor, DO ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2003, 18 (05) :553-561
[3]   Impact of gamma sterilization on clinical performance of polyethylene in the knee [J].
Collier, JP ;
Sperling, DK ;
Currier, JH ;
Sutula, LC ;
Saum, KA ;
Mayor, MB .
JOURNAL OF ARTHROPLASTY, 1996, 11 (04) :377-389
[4]   Comparison of wear of mobile and fixed bearing knees tested in a knee simulator [J].
Johnson, TS ;
Laurent, MP ;
Yao, JQ ;
Blanchard, CR .
WEAR, 2003, 255 (7-12) :1107-1112
[5]   Wear of gamma-crosslinked polyethylene acetabular cups against roughened femoral balls [J].
McKellop, H ;
Shen, FW ;
DiMaio, W ;
Lancaster, JG .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (369) :73-82
[6]   A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties - Recipient of the 1999 HAP Paul Award [J].
Muratoglu, OK ;
Bragdon, CR ;
O'Connor, DO ;
Jasty, M ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2001, 16 (02) :149-160
[7]   Knee-simulator testing of conventional and cross-linked polyethylene tibial inserts [J].
Muratoglu, OK ;
Bragdon, CR ;
Jasty, M ;
O'Connor, DO ;
Von Knoch, RS ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2004, 19 (07) :887-897
[8]   Effect of radiation, heat, and aging on in vitro wear resistance of polyethylene [J].
Muratoglu, OK ;
Merrill, EW ;
Bragdon, CR ;
O'Connor, D ;
Hoeffel, D ;
Burroughs, B ;
Jasty, M ;
Harris, WH .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (417) :253-262
[9]   Metrology to quantify wear and creep of polyethylene tibial knee inserts [J].
Muratoglu, OK ;
Perinchief, RS ;
Bragdon, CR ;
O'Connor, DO ;
Konrad, R ;
Harris, WH .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (410) :155-164
[10]   Unified wear model for highly crosslinked ultra-high molecular weight polyethylenes (UHMWPE) (vol 20, pg 1463, 1999) [J].
Muratoglu, OK ;
Bragdon, CR ;
O'Connor, DO ;
Jasty, M ;
Harris, WH ;
Gul, R ;
McGarry, F .
BIOMATERIALS, 2003, 24 (08) :1527-1527