Minimally Invasive Percutaneous TightRope® System Fixation for an Unstable Posterior Pelvic Ring: Clinical Follow-up and Biomechanical Studies

被引:3
作者
Gu, Feng [1 ]
Zhang, Jiting [1 ]
Sui, Zhenjiang [1 ]
Zhang, Ke [1 ]
Xie, Xiaoping [1 ]
Yu, Tiecheng [1 ]
机构
[1] First Hosp Jilin Univ, Dept Orthopaed, Changchun 130021, Peoples R China
基金
中国国家自然科学基金;
关键词
Biomechanics; Clinical study; Percutaneous; TightRope (R) system; Unstable pelvic fracture; ACROMIOCLAVICULAR JOINT RECONSTRUCTION; ILIOSACRAL SCREW FIXATION; INTERNAL-FIXATION; LIGAMENT RECONSTRUCTION; SACROILIAC SCREWS; SACRAL FRACTURE; SUTURE-BUTTON; REDUCTION; INJURIES; PLATE;
D O I
10.1111/os.13261
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: To evaluate the mechanical stability and clinical efficacy of minimally invasive percutaneous TightRope (R) systems applied via gun-shaped reduction forceps for unstable posterior pelvic ring fractures. Materials and methods: This study consists of two parts: a clinical retrospective study and a randomized controlled biomechanical test. For the clinical study, a retrospective analysis of posterior pelvic ring fractures was performed between June 2015 and May 2020. Eighteen patients underwent surgery using two TightRope (R) systems to fix a broken posterior pelvic ring because of unstable AO type C1 and C2 pelvic ring fractures. The patients were followed up for at least 2 years, and all patients were evaluated using the Majeed scoring system and vertical displacement. In the biomechanical tests, six embalmed adult pelvic specimens were used. The fractures were subjected to TightRope (R), IS screw, and TBP fixation in a randomized block design. The specimens were placed in a biomechanical testing machine in a standing neutral posture. A cyclic vertical load of up to 500 N was applied, and the displacement of the specimens was recorded by the testing machine. The ultimate load in each group of specimens was recorded. The displacement and ultimate load were compared and analyzed by statistical methods. Results: At a mean follow-up of 38.89 +/- 8.72 months, the functional Majeed score was excellent in 14 patients and good in four patients. The final radiological examinations showed that the outcome was excellent in 14 patients and good in four patients. In these patients, no serious clinical complications were found. Weight-bearing was delayed in four patients. In biomechanical tests, the displacement of the specimens fixed with TightRope (R) was significantly lower than that of the specimens fixed with TBP (P < 0.05) when the load ranged from 300 to 500 N. The displacement in the IS screw group was significantly lower than that in either the TBP or TightRope (R) group (P < 0.05) when the load ranged from 0 to 500 N. The ultimate load in the IS screw group (1798 +/- 83.53 N) was significantly greater than that in the TBP group (1352 +/- 74.41 N) (t = 9.78, P < 0.0001) and the TightRope (R) group (1347 +/- 54.28 N) (t = 11.11, P < 0.0001). However, no significant difference was observed between the TightRope (R) and TBP groups (t = 0.13, P = 0.90). Conclusion: Percutaneous posterior TightRope (R) system shows strong stability in mechanical experiments and shows good results in clinical follow-up while this system has certain advantages in lower surgical requirements and lower risk of related nerve and vascular structural damage.
引用
收藏
页码:1078 / 1092
页数:15
相关论文
共 55 条
[1]   Vertically unstable sacral fractures with neurological insult: outcomes of surgical decompression and reconstruction plate internal fixation [J].
Ayoub, Mostafa A. .
INTERNATIONAL ORTHOPAEDICS, 2009, 33 (01) :261-267
[2]   Tape suture for stabilization of incomplete posterior pelvic ring fractures-biomechanical analysis of a new minimally invasive treatment for incomplete lateral compression pelvic ring fractures [J].
Becker, Christopher Alexander ;
Kussmaul, Adrian Cavalcanti ;
Suero, Eduardo Manuel ;
Regauer, Markus ;
Woiczinski, Matthias ;
Braun, Christian ;
Flatz, Wilhelm ;
Pieske, Oliver ;
Kammerlander, Christian ;
Boecker, Wolfgang ;
Greiner, Axel .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2019, 14 (01)
[3]   Treatment of Unstable Posterior Pelvic Ring Fracture with Pedicle Screw-Rod Fixator Versus Locking Compression Plate: A Comparative Study [J].
Bi, Chun ;
Wang, Qiugen ;
Nagelli, Christopher ;
Wu, Jianhong ;
Wang, Qian ;
Wang, Jiandong .
MEDICAL SCIENCE MONITOR, 2016, 22 :3764-3770
[4]   Does adjustable-loop femoral cortical suspension loosen after anterior cruciate ligament reconstruction? A retrospective comparative study [J].
Boyle, Matthew J. ;
Vovos, Tyler J. ;
Walker, Cameron G. ;
Stabile, Kathryne J. ;
Roth, Jonathan M. ;
Garrett, William E., Jr. .
KNEE, 2015, 22 (04) :304-308
[5]   Risks to the superior gluteal neurovascular bundle during percutaneous iliosacral screw insertion - An anatomical cadaver study [J].
Collinge, C ;
Coons, D ;
Aschenbrenner, J .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2005, 19 (02) :96-101
[6]   Biomechanical comparison of posterior internal fixation techniques for unstable pelvic fractures [J].
Comstock, CP ;
vanderMeulen, MCH ;
Goodman, SB .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1996, 10 (08) :517-522
[7]  
Culemann U, 2007, UNFALLCHIRURG, V110, P528, DOI 10.1007/s00113-007-1236-4
[8]   Mini TightRope® suture button indications for thumb basal joint arthritis [J].
Dreant, N. .
HAND SURGERY & REHABILITATION, 2021, 40 :S77-S82
[9]   Syndesmotic Stabilization: Syndesmotic Screw Versus Flexible Fixation: A Systematic Review [J].
Elabd, Ahmed ;
Abdullah, Samir ;
Kandel, Wael ;
Hegazy, Mohamed .
JOURNAL OF FOOT & ANKLE SURGERY, 2021, 60 (05) :998-1007
[10]   Biomechanical comparison of different fixation techniques for reconstruction of tibial avulsion fractures of the anterior cruciate ligament [J].
Ezechieli, Marco ;
Schaefer, Madeline ;
Becher, Christoph ;
Dratzidis, Antonios ;
Glaab, Richard ;
Ryf, Christian ;
Hurschler, Christof ;
Ettinger, Max .
INTERNATIONAL ORTHOPAEDICS, 2013, 37 (05) :919-923