Closed-loop glucose control in critically ill patients using continuous glucose monitoring system (CGMS) in real time

被引:81
作者
Chee, F [1 ]
Fernando, T
van Heerden, PV
机构
[1] Univ Western Australia, Sch Elect Elect & Comp Engn, Crawley, WA 6009, Australia
[2] Sir Charles Gairdner Hosp, Intens Care Unit, Nedlands, WA 6009, Australia
来源
IEEE TRANSACTIONS ON INFORMATION TECHNOLOGY IN BIOMEDICINE | 2003年 / 7卷 / 01期
关键词
artificial pancreas; closed-loop glucose control; intravenous insulin infusion; sliding scale; subcutaneous blood glucose monitoring;
D O I
10.1109/TITB.2003.808509
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A study was conducted to determine if continuous subcutaneous glucose monitoring (from MiniMed CGMS) could be used in real-time to control blood sugar level (BSL) in patients with critical illness. A closed-loop control system was constructed to use CGMS in a real-time manner, coupled with. a proportional integral (PI) control algorithm based on a sliding scale approach, for automatic intravenous infusion of insulin to patients. A total of five subjects with high BSL (> 10 mmol/L) participated in formal studies of the closed-loop control system. Subjects were recruited from critically ill patients in the intensive care unit (ICU) after informed consent was obtained. Error grid analysis showed that 64.6% of the BSL readings as determined in real time using CGMS sensor, when compared to conventional BSL measurements on blood drawn from an arterial line, was clinically accurate (i.e., < 20% deviation from glucometer value). In the five patients who underwent closed-loop control, the controller managed to control only one patient's glycaemia without any manual intervention. Manual intervention was required due to the real-time sensor reading deviating more than 20% from the glucometer value, and also as a safety mechanism. Test on equality of mean and variance for BSL attained prior to, during, and post trial showed that the controller's performance was comparable to manual control. We conclude that the automatic sliding scale approach of closed-loop BSL control is feasible in patients in intensive care. More work is needed in the refinement of the algorithm and the improvement of real-time sensor accuracy.
引用
收藏
页码:43 / 53
页数:11
相关论文
共 29 条
[1]   Continuous glucose monitoring used to adjust diabetes therapy improves glycosylated hemoglobin: a pilot study [J].
Bode, BW ;
Gross, TM ;
Thornton, KR ;
Mastrototaro, JJ .
DIABETES RESEARCH AND CLINICAL PRACTICE, 1999, 46 (03) :183-190
[2]   COMPARISON OF ALGORITHMS FOR THE CLOSED-LOOP CONTROL OF BLOOD-GLUCOSE USING THE ARTIFICIAL BETA-CELL [J].
BROEKHUYSE, HM ;
NELSON, JD ;
ZINMAN, B ;
ALBISSER, AM .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1981, 28 (10) :678-687
[3]   A SEMI-CLOSED LOOP COMPUTER-ASSISTED INSULIN INFUSION SYSTEM - HOSPITAL USE FOR CONTROL OF DIABETES IN PATIENTS [J].
CHISHOLM, DJ ;
KRAEGEN, EW ;
BELL, DJ ;
CHIPPS, DR .
MEDICAL JOURNAL OF AUSTRALIA, 1984, 141 (12-1) :784-789
[4]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[5]   CONTINUOUS IN-VIVO MONITORING IN DIABETES - THE SUBCUTANEOUS GLUCOSE-CONCENTRATION [J].
FISCHER, U .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1995, 39 :21-29
[6]   BLOOD-GLUCOSE CONTROL BY INTERMITTENT LOOP CLOSURE IN THE BASAL MODE - COMPUTER-SIMULATION STUDIES WITH A DIABETIC MODEL [J].
FURLER, SM ;
KRAEGEN, EW ;
SMALLWOOD, RH ;
CHISHOLM, DJ .
DIABETES CARE, 1985, 8 (06) :553-561
[7]  
GABOR H, 2000, BIOMEDICAL APPL FUND
[8]   DEVELOPMENT OF A MINIATURIZED GLUCOSE MONITORING-SYSTEM BY COMBINING A NEEDLE-TYPE GLUCOSE SENSOR WITH MICRODIALYSIS SAMPLING METHOD - LONG-TERM SUBCUTANEOUS TISSUE GLUCOSE MONITORING IN AMBULATORY DIABETIC-PATIENTS [J].
HASHIGUCHI, Y ;
UEMURA, T ;
SAKAKIDA, M ;
KAJIWARA, K ;
NISHIDA, K ;
SHICHIRI, M .
DIABETES CARE, 1994, 17 (05) :387-396
[9]   GLUCOSE CLAMPS WITH THE BIOSTATOR - A CRITICAL REAPPRAISAL [J].
HEINEMANN, L ;
AMPUDIABLASCO, FJ .
HORMONE AND METABOLIC RESEARCH, 1994, 26 (12) :579-583
[10]   Advances toward the implantable artificial pancreas for treatment of diabetes [J].
Jaremko, J ;
Rorstad, O .
DIABETES CARE, 1998, 21 (03) :444-450