Absolute dose measurements by means of a small cylindrical ionization chamber for very high dose per pulse high energy electron beams

被引:12
|
作者
Karaj, E. [1 ]
Righi, S. [1 ]
Di Martino, F. [1 ]
机构
[1] Azienda Osped Univ Pisana, UO Fis Sanitaria, I-56126 Pisa, Italy
关键词
IORT; dose per pulse; k(sat); inclined applicators; small cylindrical ionization chamber;
D O I
10.1118/1.2436979
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Very high dose per pulse (3-13 cGy/pulse) high energy electron beams are currently produced by special linear accelerators (linac) dedicated to Intra Operative Radiation Therapy (IORT). The electron beams produced by such linacs are collimated by special Perspex applicators of various size and cylindrically shaped. The biggest problems from the dosimetric point of view are caused by the high dose-per-pulse values and the use of inclined applicators. In this work measurements of absolute dose for the inclined applicators were done by using a small cylindrical ionization chamber, type CC01 (Wellhofer), a parallel plane ionization chamber type Markus (PTW 23343) and radiochromic films type EBT. We show a method which allows calculating the quality correction factors for CC01 chamber with an uncertainty of 1% and the absolute dose value for the inclined applicators using CC01 with an uncertainty of 3.1% for electron beams of energy of 6 and 7 MeV produced by the linac dedicated to IORT Novac7. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:952 / 958
页数:7
相关论文
共 50 条
  • [31] The Advanced Markus ionization chamber is useable for measurements at ultra high dose rates
    Petersson, K.
    Jaccard, M.
    Buchillier, T.
    Bailat, C.
    Germond, J.
    Vozenin, M.
    Bourhis, J.
    Bochud, F.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S373 - S373
  • [32] Comparison of film measurements and Monte Carlo simulations of dose delivered with very high-energy electron beams in a polystyrene phantom
    Bazalova-Carter, Magdalena
    Liu, Michael
    Palma, Bianey
    Dunning, Michael
    McCormick, Doug
    Hemsing, Erik
    Nelson, Janice
    Jobe, Keith
    Colby, Eric
    Koong, Albert C.
    Tantawi, Sami
    Dolgashev, Valery
    Maxim, Peter G.
    Loo, Billy W., Jr.
    MEDICAL PHYSICS, 2015, 42 (04) : 1606 - 1613
  • [33] Experimental investigation of the collection efficiency of a PTW Roos ionization chamber irradiated with pulsed beams at high pulse dose with different pulse lengths
    Karsch, Leonhard
    Richter, Christian
    Pawelke, Joerg
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2011, 21 (01): : 4 - 10
  • [34] DOSE CALCULATIONS FOR HIGH-ENERGY ELECTRON AND PHOTON BEAMS
    CUNNINGHAM, JR
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1989, 13 (03) : 241 - 250
  • [35] DOSE DISTRIBUTIONS IN HIGH-ENERGY ELECTRON-BEAMS
    BOAG, JW
    PHYSICS IN MEDICINE AND BIOLOGY, 1972, 17 (05): : 671 - &
  • [36] Dose rate dependence of the PTW 60019 microDiamond detector in high dose-per-pulse pulsed beams
    Pardo-Montero, J.
    Brualla-Gonzalez, L.
    Gomez, F.
    Pombar, M.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S375 - S375
  • [37] Dose rate dependence of the PTW 60019 microDiamond detector in high dose-per-pulse pulsed beams
    Brualla-Gonzalez, Luis
    Gomez, Faustino
    Pombar, Miguel
    Pardo-Montero, Juan
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (01): : N11 - N19
  • [38] Correction for volume recombination in liquid ionization chambers at high dose-per-pulse
    Gotz, Malte
    Ka, Leonhard
    Tolli, Heikki
    Pawelke, Joerg
    MEDICAL PHYSICS, 2019, 46 (08) : 3692 - 3699
  • [39] The saturation loss for plane parallel ionization chambers at high dose per pulse values
    Piermattei, A
    delle Canne, S
    Azario, L
    Russo, A
    Fidanzio, A
    Miceli, R
    Soriani, A
    Orvieto, A
    Fantini, M
    PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (07): : 1869 - 1883
  • [40] Possibility of using cylindrical ionization chambers for percent depth-dose measurements in clinical electron beams
    Ono, Takeshi
    Araki, Fujio
    Yoshiyama, Fumiaki
    MEDICAL PHYSICS, 2011, 38 (08) : 4647 - 4654