Absolute dose measurements by means of a small cylindrical ionization chamber for very high dose per pulse high energy electron beams

被引:12
作者
Karaj, E. [1 ]
Righi, S. [1 ]
Di Martino, F. [1 ]
机构
[1] Azienda Osped Univ Pisana, UO Fis Sanitaria, I-56126 Pisa, Italy
关键词
IORT; dose per pulse; k(sat); inclined applicators; small cylindrical ionization chamber;
D O I
10.1118/1.2436979
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Very high dose per pulse (3-13 cGy/pulse) high energy electron beams are currently produced by special linear accelerators (linac) dedicated to Intra Operative Radiation Therapy (IORT). The electron beams produced by such linacs are collimated by special Perspex applicators of various size and cylindrically shaped. The biggest problems from the dosimetric point of view are caused by the high dose-per-pulse values and the use of inclined applicators. In this work measurements of absolute dose for the inclined applicators were done by using a small cylindrical ionization chamber, type CC01 (Wellhofer), a parallel plane ionization chamber type Markus (PTW 23343) and radiochromic films type EBT. We show a method which allows calculating the quality correction factors for CC01 chamber with an uncertainty of 1% and the absolute dose value for the inclined applicators using CC01 with an uncertainty of 3.1% for electron beams of energy of 6 and 7 MeV produced by the linac dedicated to IORT Novac7. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:952 / 958
页数:7
相关论文
共 15 条
[1]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[2]   Intraoperative radiation therapy using mobile electron linear accelerators: Report of AAPM radiation therapy committee task group no. 72 [J].
Beddar, A. Sam ;
Biggs, Peter J. ;
Chang, Sha ;
Ezzell, Gary A. ;
Faddegon, Bruce A. ;
Hensley, Frank W. ;
Mills, Michael D. .
MEDICAL PHYSICS, 2006, 33 (05) :1476-1489
[3]   Calibration of low-energy electron beams from a mobile linear accelerator with plane-parallel chambers using both TG-51 and TG-21 protocols [J].
Beddar, AS ;
Tailor, RC .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (08) :N105-N110
[4]   Measurements of output factors with different detector types and Monte Carlo calculations of stopping-power ratios for degraded electron beams [J].
Björk, P ;
Knöös, T ;
Nilsson, P .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (19) :4493-4506
[5]   Dosimetry characteristics of degraded electron beams investigated by Monte Carlo calculations in a setup for intraoperative radiation therapy [J].
Björk, P ;
Nilsson, P ;
Knöös, T .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (02) :239-256
[7]   The effect of free-electron collection on the recombination correction to ionization measurements of pulsed radiation [J].
Boag, JW ;
Hochhauser, E ;
Balk, OA .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (05) :885-897
[8]   CURRENT COLLECTION AND IONIC RECOMBINATION IN SMALL CYLINDRICAL IONIZATION CHAMBERS EXPOSED TO PULSED RADIATION [J].
BOAG, JW ;
CURRANT, J .
BRITISH JOURNAL OF RADIOLOGY, 1980, 53 (629) :471-478
[9]   Weak energy dependence of EBT gafchromic film dose response in the 50 kVp-10 MVp X-ray range [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN .
APPLIED RADIATION AND ISOTOPES, 2006, 64 (01) :60-62
[10]   Absorption spectra variations of EBT radiochromic film from radiation exposure [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (13) :N135-N140