Comparison of chemical modifiers in electrothermal atomization atomic absorption spectrometry with boron nitride tube and platform furnaces for mercury

被引:6
作者
Imai, S [1 ]
Sugimoto, A
Hayashi, M
Yonetani, A
Iwamoto, E
Hayashi, Y
机构
[1] Univ Tokushima, Fac Integrated Arts & Sci, Dept Chem, Tokushima 7708502, Japan
[2] Hitachi Sci Syst Ltd, Hitachinaka, Ibaraki 3120033, Japan
[3] Hiroshima Womens Univ, Dept Hlth Sci, Hiroshima 7348558, Japan
[4] Joetsu Univ Educ, Dept Chem, Niigata 9430815, Japan
关键词
Certified reference materials - Certified values - Chemical modifiers - Electrothermal atomization atomic absorption spectrometries - Graphite furnaces - Human hair - Maximum absorbance - Pyrolytic boron nitrides;
D O I
10.2116/analsci.16.739
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The effect of various chemical modifiers (KMno(4), AgNo(3)+KMnO(4), Pd, Rh, Ir and Pd+Rh) on ETAAS of Hg with laboratory-made boron nitride tube and platform furnaces was studied. The effectiveness of the chemical modifier was improved by the use of the ceramic furnaces instead of the graphite furnace. The sensitivity, which gives a maximum absorbance of 0.0044 abs for Hg, is found to be 0.11 ng, and the LOD, based on the variability of the blank (3 sigma), is 27 pg using 10 mu g of Pd with the non-pyrolytic boron nitride platform furnace. The good recovery of 102+/-3% was observed for Hg spiked into a hair sample digested by microwave with nitric acid. The observed Hg content in a certified reference material of human hair was in agreement with its certified value.