Fixation and Wear With a Contemporary Acetabular Component and Cross-Linked Polyethylene at Minimum 10-Year Follow-Up

被引:24
作者
Bedard, Nicholas A. [1 ]
Callaghan, John J. [1 ,2 ]
Stefl, Michael D. [3 ]
Willman, Tyler J. [1 ]
Liu, Steve S. [1 ]
Goetz, Devon D. [4 ]
机构
[1] Univ Iowa, Iowa City, IA USA
[2] VA Med Ctr, Iowa City, IA USA
[3] Univ So Calif, Los Angeles, CA USA
[4] Des Moines Orthopaed Surg, W Des Moines, IA USA
关键词
cementless acetabular component; wear; cross-linked polyethylene; outcomes; activity level; TOTAL HIP-ARTHROPLASTY; FEMORAL COMPONENTS; PATIENTS YOUNGER; REPLACEMENT; THA; CHARNLEY; CEMENT; OSTEOARTHRITIS; STEM; AGE;
D O I
10.1016/j.arth.2014.05.008
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Long term total hip arthroplasty follow-up has demonstrated bearing surface wear and failure of fixation as the major modes of failure. The purpose of this study was to evaluate the minimum 10 year results of a third generation cementless acetabular component using moderately cross-linked polyethylene liners. 150 primary THAs were evaluated clinically for need for revision and activity evaluated using accelerometers, WOMAC, Tegner and UCLA activity-level scores. Radiographs were evaluated for wear, osteolysis and loosening. We compared this cohort to three of our previously reported cohorts. The only revision was unrelated to liner wear and all acetabular components were bone ingrown. There were no cases of acetabular osteolysis and one case of femoral osteolysis. Average steps per year were 1.59 million. Mean steady state wear rate averaged 0.04 mm/year. At 10-years, excellent fixation and low wear were demonstrated using a cementless acetabular construct and moderately cross-linked polyethylene liner. Level of evidence: Therapeutic study, Level III (retrospective comparative study-prospective cohort with historical control). (c) 2014 Elsevier Inc. All rights reserved.
引用
收藏
页码:1961 / 1969
页数:9
相关论文
共 43 条
[11]   A Prospective, Randomized Study of Cross-Linked and Non-Cross-Linked Polyethylene for Total Hip Arthroplasty at 10-Year Follow-Up [J].
Engh, Charles A., Jr. ;
Hopper, Robert H., Jr. ;
Huynh, Cathy ;
Ho, Henry ;
Sritulanondha, Supatra ;
Engh, Charles A., Sr. .
JOURNAL OF ARTHROPLASTY, 2012, 27 (08) :2-7
[12]   Wear Comparison Between Conventional and Highly Cross-Linked Polyethylene Against a Zirconia Head A Concise Follow-Up, at an Average 10 Years, of a Previous Report [J].
Fukui, Kiyokazu ;
Kaneuji, Ayumi ;
Sugimori, Tanzo ;
Ichiseki, Toru ;
Matsumoto, Tadami .
JOURNAL OF ARTHROPLASTY, 2013, 28 (09) :1654-1658
[13]   Cementless acetabular fixation at fifteen years - A comparison with the same surgeon's results following acetabular fixation with cement [J].
Gaffey, JL ;
Callaghan, JJ ;
Pedersen, DR ;
Goetz, DD ;
Sullivan, PM ;
Johnston, RC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (02) :257-261
[14]   New polyethylenes in total hip replacement A TEN- TO 12-YEAR FOLLOW-UP STUDY [J].
Garcia-Rey, E. ;
Garcia-Cimbrelo, E. ;
Cruz-Pardos, A. .
BONE & JOINT JOURNAL, 2013, 95B (03) :326-332
[15]   Cross-linked Compared with Historical Polyethylene in THA: An 8-year Clinical Study [J].
Geerdink, Carel H. ;
Grimm, Bernd ;
Vencken, Wendy ;
Heyligers, Ide C. ;
Tonino, Alphons J. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2009, 467 (04) :979-984
[16]  
GRUEN TA, 1979, CLIN ORTHOP RELAT R, P17
[17]   LOOSENING OF THE FEMORAL COMPONENT AFTER USE OF THE MEDULLARY-PLUG CEMENTING TECHNIQUE - FOLLOW-UP NOTE WITH A MINIMUM 5-YEAR FOLLOW-UP [J].
HARRIS, WH ;
MCGANN, WA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1064-1066
[18]   Second-generation Extensively Porous-coated THA Stems at Minimum 10-year Followup [J].
Hennessy, David W. ;
Callaghan, John J. ;
Liu, Steve S. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2009, 467 (09) :2290-2296
[19]   Highly Crosslinked Polyethylene Does Not Reduce Aseptic Loosening in Cemented THA 10-year Findings of a Randomized Study [J].
Johanson, Per-Erik ;
Digas, Georgios ;
Herberts, Peter ;
Thanner, Jonas ;
Karrholm, Johan .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (11) :3083-3093
[20]   CLINICAL AND RADIOGRAPHIC EVALUATION OF TOTAL HIP-REPLACEMENT - A STANDARD SYSTEM OF TERMINOLOGY FOR REPORTING RESULTS [J].
JOHNSTON, RC ;
FITZGERALD, RH ;
HARRIS, WH ;
POSS, R ;
MULLER, ME ;
SLEDGE, CB .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (02) :161-168