Sensitivity and Selectivity of Ion-Selective Electrodes Interpreted Using the Nernst-Planck-Poisson Model

被引:8
|
作者
Jasielec, Jerzy J. [1 ]
Mousavi, Zekra [2 ]
Granholm, Kim [2 ]
Sokolski, Tomasz [2 ]
Lewenstam, Andrzej [1 ,2 ]
机构
[1] AGH Univ Sci & Technol, Fac Mat Sci & Ceram, Phys Chem & Modeling Dept, Al Mickiewicza 30, PL-30059 Krakow, Poland
[2] Abo Akad Univ, Ctr Proc Analyt Chem & Sensor Technol ProSens, Johan Gadolin Proc Chem Ctr, Biskopsgatan 8, Turku 20500, Finland
关键词
LIQUID-MEMBRANE ELECTRODES; LOWER DETECTION LIMIT; TIME-DEPENDENT PHENOMENA; SEPARATE SOLUTION METHOD; TECHNICAL REPORT; NUMERICAL-SIMULATION; POTENTIAL RESPONSE; BULK OPTODES; COEFFICIENTS; SENSORS;
D O I
10.1021/acs.analchem.8b02659
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The Nernst-Planck-Poisson model is used for modeling the sensitivity and selectivity of ion-selective electrodes (ISEs) with plastic membranes. Two pivotal parameters characterizing ISE response are in focus: sensitivity and selectivity. An interpretation of sensitivity, which considers the concurrent influence of anions and cations on the ISE slope, is presented. The interpretation of selectivity shows the validity and limits of approaches hitherto taken to measure the true (unbiased) selectivity coefficient. The validity of more idealized interpretations by the diffusion-layer model is conceived.
引用
收藏
页码:9644 / 9649
页数:6
相关论文
共 50 条
  • [1] Neutral-Carrier Ion-Selective Electrodes Assessed by the Nernst-Planck-Poisson Model
    Jasielec, Jerzy J.
    Sokalski, Tomasz
    Filipek, Robert
    Lewenstam, Andrzej
    ANALYTICAL CHEMISTRY, 2015, 87 (17) : 8665 - 8672
  • [2] Nernst-Planck-Poisson Model for the Description of Behaviour of Solid-Contact Ion-Selective Electrodes at Low Analyte Concentration
    Jasielec, Jerzy J.
    Lisak, Grzegorz
    Wagner, Michal
    Sokalski, Tomasz
    Lewenstam, Andrzej
    ELECTROANALYSIS, 2013, 25 (01) : 133 - 140
  • [3] Time-dependent phenomena in the potential response of ion-selective electrodes treated by the Nernst-Planck-Poisson model. 1. Intramembrane processes and selectivity
    Lingenfelter, Peter
    Bedlechowicz-Sliwakowska, Iwona
    Sokalski, Tomasz
    Maj-Zurawska, Magdalena
    Lewenstam, Andrzej
    ANALYTICAL CHEMISTRY, 2006, 78 (19) : 6783 - 6791
  • [4] Electro-diffusion in multicomponent ion-selective membranes; numerical solution of the coupled nernst-planck-poisson equations
    Grysakowski, Bartosz
    Bozek, Boguslaw
    Danielewski, Marek
    DIFFUSION IN SOLIDS AND LIQUIDS III, 2008, 273-276 : 113 - +
  • [5] Time-Dependent Phenomena in the Potential Response of Ion-Selective Electrodes Treated by the Nernst-Planck-Poisson Model. Part 2: Transmembrane Processes and Detection Limit
    Sokalski, Tomasz
    Kucza, Witold
    Danielewski, Marek
    Lewenstam, Andrzej
    ANALYTICAL CHEMISTRY, 2009, 81 (12) : 5016 - 5022
  • [6] EIS simulations for ion-selective site-based membranes by a numerical solution of the coupled Nernst-Planck-Poisson equations
    Kucza, W
    Danielewski, M
    Lewenstam, A
    ELECTROCHEMISTRY COMMUNICATIONS, 2006, 8 (03) : 416 - 420
  • [7] Electrochemistry of Symmetrical Ion Channel: A Three-Dimensional Nernst-Planck-Poisson Model
    Bozek, B.
    Lewenstam, A.
    Tkacz-Smiech, K.
    Danielewski, M.
    SYMPOSIUM IN HONOR OF RICHARD BUCK, 2014, 61 (15): : 11 - 20
  • [8] A Finite Volume Scheme for Nernst-Planck-Poisson Systems with Ion Size and Solvation Effects
    Fuhrmann, Juergen
    Guhlke, Clemens
    FINITE VOLUMES FOR COMPLEX APPLICATIONS VIII-HYPERBOLIC, ELLIPTIC AND PARABOLIC PROBLEMS, 2017, 200 : 497 - 505
  • [9] Numerical solution of the coupled Nernst-Planck and poisson equations for ion-selective membrane potentials
    Lingenfelter, P
    Sokalski, T
    Lewenstam, A
    MEMBRANES-PREPARATION, PROPERTIES AND APPLICATIONS, 2003, 752 : 193 - 198
  • [10] ASSESSMENT OF THE SELECTIVITY OF ION-SELECTIVE ELECTRODES
    OKUNEV, MS
    KHITROVA, NV
    KORNIENKO, OI
    JOURNAL OF ANALYTICAL CHEMISTRY OF THE USSR, 1982, 37 (01): : 1 - 8