Sequential injection system in flame atomic absorption spectrometry for the determination of calcium and magnesium in mineral waters

被引:55
作者
Araújo, AN
Costa, RCC
Lima, JLFC
Reis, BF
机构
[1] Fac Farm UP, Dept Quim Fis, CEQUP, P-4050 Porto, Portugal
[2] USP, Ctr Energia Nucl Agr, BR-13418260 Piracicaba, SP, Brazil
关键词
SIA; flame atomic absorption spectrometry; Ca and Mg in mineral waters;
D O I
10.1016/S0003-2670(97)00583-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow set-up based on sequential injection analysis (SIA) is proposed to increase the versatility and robustness of routine flame atomic absorption analytical methods, Four procedures were selected to allow the determinations of calcium within the ranges of 0.03-10.00 and 0.58-200.00 mg l(-1), and magnesium within 0.03-1.5 and 0.80-50.0 mg l(-1), in mineral waters. The sample aliquots inserted were 70.8 and 12 mu l for calcium and 13 and 12 mu l for magnesium. Moreover the system developed enabled to restrict the consumption of the releasing agent to respectively 142 and 97 mu l in the determinations of calcium and to 26 and 12 mu l in the determinations of magnesium. The results obtained in the analysis of 15 mineral waters were compared with those obtained by the corresponding conventional procedures yielding the equations: [Ca mg l(-1)](SIA)=1.008(+/-0.004)*[Ca mg l(-1)](conv.) -0.3 (+/-0.3) and [Mg mg l(-1)](SIA)=0.992(+/-0.006)*[Mg mg l(-1)](conv.) +0.0 (+/-0.1). In all cases the sampling rates achieved were higher than 110 h(-1), and reproducibility being better than 3%. (C) 1998 Elsevier Science B.V.
引用
收藏
页码:111 / 119
页数:9
相关论文
共 20 条
[1]   MULTICOMMUTATION IN FLOW-ANALYSIS .3. SPECTROPHOTOMETRIC KINETIC DETERMINATION OF CREATININE IN URINE EXPLOITING A NOVEL ZONE SAMPLING APPROACH [J].
ARAUJO, AN ;
LIMA, JLFC ;
REIS, BF ;
ZAGATTO, EAG .
ANALYTICA CHIMICA ACTA, 1995, 310 (03) :447-452
[2]  
Burguera J. L., 1989, FLOW INJECTION ATOMI
[3]  
EBON L, 1991, J AUTOM CHEM, V13, P281
[4]   INDIRECT ATOMIC-ABSORPTION DETERMINATION OF ANIONIC SURFACTANTS IN WASTEWATERS BY FLOW-INJECTION CONTINUOUS LIQUID-LIQUID-EXTRACTION [J].
GALLEGO, M ;
SILVA, M ;
VALCARCEL, M .
ANALYTICAL CHEMISTRY, 1986, 58 (11) :2265-2269
[5]   USE OF SUBMICROLITER-VOLUME SAMPLES FOR EXTENDING THE DYNAMIC-RANGE OF FLOW-INJECTION FLAME ATOMIC-ABSORPTION SPECTROMETRY [J].
GARCIA, IL ;
VINAS, P ;
CAMPILLO, N ;
CORDOBA, MH .
ANALYTICA CHIMICA ACTA, 1995, 308 (1-3) :85-95
[6]   FLOW-INJECTION DILUTION SYSTEM FOR THE ANALYSIS OF HIGHLY CONCENTRATED SAMPLES USING FLAME ATOMIC-ABSORPTION SPECTROMETRY [J].
GARCIA, IL ;
VINAS, P ;
CORDOBA, MH .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (10) :1167-1172
[7]   FROM FLOW-INJECTION TO SEQUENTIAL INJECTION - COMPARISON OF METHODOLOGIES AND SELECTION OF LIQUID DRIVES [J].
IVASKA, A ;
RUZICKA, J .
ANALYST, 1993, 118 (07) :885-889
[8]   Determination of calcium, magnesium, sodium and potassium in wines by FIA using an automatic zone sampling system [J].
Lapa, RAS ;
Lima, JLFC ;
Santos, JLM .
FOOD CHEMISTRY, 1996, 55 (04) :397-402
[9]  
Lima JLFC, 1996, ATOM SPECTROSC, V17, P196
[10]   MULTICOMMUTATION IN FLOW-ANALYSIS .1. BINARY SAMPLING - CONCEPTS, INSTRUMENTATION AND SPECTROPHOTOMETRIC DETERMINATION OF IRON IN PLANT DIGESTS [J].
REIS, BF ;
GINE, MF ;
ZAGATTO, EAG ;
LIMA, JLFC ;
LAPA, RA .
ANALYTICA CHIMICA ACTA, 1994, 293 (1-2) :129-138