Evaluation of the A&D UA-767 and Welch Allyn Spot Vital Signs noninvasive blood pressure monitors using a blood pressure simulator

被引:11
作者
Davis, PD
Dennis, JL
Railton, R
机构
[1] So Gen Hosp, Dept Clin Phys & Bioengn, Glasgow G51 4TF, Lanark, Scotland
[2] Monklands Hosp, Dept Med Phys, Airdrie, Scotland
关键词
noninvasive; blood pressure; monitor; protocol;
D O I
10.1038/sj.jhh.1001804
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
The performance of five units of the A&D UA-767 NIBP monitor and five units of the Welch Allyn Spot Vital Signs noninvasive blood pressure monitor was evaluated with the Biotek BP Pump blood pressure simulator under a variety of conditions. Using the simulator to provide a normal blood pressure waveform at 80 bpm over a range of pressures, it was found that the mean bias for the combined results from the A&D monitors was 1.9+/-2.8 mmHg and from the Welch Allyn monitors was 0.7+/-2.4 mmHg. No individual measurement showed a bias greater than 10 mmHg. A bias of greater than 5 mmHg was present in 28 out of 150 measurements for the A&D monitor and 10 out of 150 measurements for the Welch Allyn monitor. These results are comparable with ratings achieved by the instruments when tested previously according to the British Hypertension Society protocol, but testing with a simulator allowed assessment of aspects of performance which were not included in the British Hypertension Society protocol.
引用
收藏
页码:197 / 203
页数:7
相关论文
共 11 条
[1]   Evaluation of the Critikon 8100 and Spacelabs 90207 non-invasive blood pressure monitors using a test simulator [J].
Amoore, JN ;
Geake, WB .
JOURNAL OF HUMAN HYPERTENSION, 1997, 11 (03) :163-169
[2]  
[Anonymous], 1987, NAT STAND EL AUT SPH
[3]  
Association for the Advancement of Medical Instrumentation, 1993, AM NAT STAND EL AUT
[4]   Validation of the Welch Allyn 'Vital Signs' oscillometric blood pressure monitor [J].
Jones, CR ;
Taylor, K ;
Poston, L ;
Shennan, AH .
JOURNAL OF HUMAN HYPERTENSION, 2001, 15 (03) :191-195
[5]   Toxicity of mercury [J].
Langford, NJ ;
Ferner, RE .
JOURNAL OF HUMAN HYPERTENSION, 1999, 13 (10) :651-656
[6]   Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults [J].
O'Brien, E ;
Pickering, T ;
Asmar, R ;
Myers, M ;
Parati, G ;
Staessen, J ;
Mengden, T ;
Imai, Y ;
Waeber, B ;
Palatini, P ;
Atkins, N ;
Gerin, W .
BLOOD PRESSURE MONITORING, 2002, 7 (01) :3-17
[7]   Blood pressure measuring devices: recommendations of the European Society of Hypertension [J].
O'Brien, E ;
Waeber, B ;
Parati, G ;
Staessen, J ;
Myers, MG .
BMJ-BRITISH MEDICAL JOURNAL, 2001, 322 (7285) :531-536
[8]   THE BRITISH-HYPERTENSION-SOCIETY PROTOCOL FOR THE EVALUATION OF AUTOMATED AND SEMIAUTOMATED BLOOD-PRESSURE MEASURING DEVICES WITH SPECIAL REFERENCE TO AMBULATORY SYSTEMS [J].
OBRIEN, E ;
PETRIE, J ;
LITTLER, W ;
DESWIET, M ;
PADFIELD, PL ;
OMALLEY, K ;
JAMIESON, M ;
ALTMAN, D ;
BLAND, M ;
ATKINS, N .
JOURNAL OF HYPERTENSION, 1990, 8 (07) :607-619
[9]  
OBRIEN E, 1994, J HYPERTENS, V12, P1089
[10]   THE BRITISH-HYPERTENSION-SOCIETY PROTOCOL FOR THE EVALUATION OF AUTOMATED AND SEMIAUTOMATED BLOOD-PRESSURE MEASURING DEVICES WITH SPECIAL REFERENCE TO AMBULATORY SYSTEMS [J].
OBRIEN, E ;
PETRIE, J ;
LITTLER, W ;
DESWIET, M ;
PADFIELD, PL ;
OMALLEY, K ;
JAMIESON, M ;
ALTMAN, D ;
BLAND, M ;
ATKINS, N .
JOURNAL OF HYPERTENSION, 1990, 8 (07) :607-619