Diffusion Tensor Imaging (DTI) of the Kidney at 3 Tesla-Feasibility, Protocol Evaluation and Comparison to 1.5 Tesla

被引:91
作者
Notohamiprodjo, Mike [1 ]
Dietrich, Olaf [2 ]
Horger, Wihelm [3 ]
Horng, Annie [1 ]
Helck, Andreas D. [1 ]
Herrmann, Karin A. [1 ]
Reiser, Maximilian F. [1 ]
Glaser, Christian [1 ]
机构
[1] Univ Hosp Munich, Dept Clin Radiol, Inst Clin Radiol, D-81377 Munich, Germany
[2] Univ Hosp Munich, Dept Clin Radiol, Josef Lissner Lab Biomed Imaging, D-81377 Munich, Germany
[3] Siemens Healthcare Sector, Erlangen, Germany
关键词
diffusion tensor imaging; kidney; 3T; DTI; DWI; reproducibility; 3.0; T; MR; ANISOTROPY; COEFFICIENT; NOISE; REPRODUCIBILITY; STRATEGIES; SCHEMES; ABDOMEN; EXPERIENCE;
D O I
10.1097/RLI.0b013e3181d83abc
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to evaluate the feasibility of diffusion tensor imaging of the kidney at a field strength of 3T. We assessed fractional anisotropy (FA) and apparent diffusion coefficients (ADC) of various acquisition protocols and determined the reproducibility of these measurements. FA, ADC, signal-to-noise ratios (SNR), and contrast-to-noise ratios (CNR) were compared with those acquired at 1.5T. Material and Methods: Ten healthy volunteers were examined with a respiratory-triggered echo-planar imaging sequence (TR: 1800 ms, TE: 58 ms, b = 0, 300 s/mm(2)) on a 3-Tesla whole-body MR scanner. Protocol variations included diffusion measurements during free-breathing, in 6 or 12 directions, and an additional b-value of 50 s/mm(2). A breath-hold protocol was also integrated (TR: 820 ms, TE: 58 ms, b = 0, 300 s/mm(2)). Measurements with 2 b-values and 6 diffusion directions were also acquired at 1.5 T. SNR was calculated with the difference-image method. Statistical analysis was performed with Wilcoxon signed-rank tests. Intrareader correlation was assessed with weighted kappa coefficients and reproducibility with the root-mean-square-average and the Bland-Altman-method. Results: At 3T, SNR of cortex and medulla and CNR of cortex/medulla were significantly higher than at 1.5T, leading to improved corticomedullary discrimination. There were no significant FA-and ADC differences with 2 b-values and 6 diffusion directions between measurements at 1.5T and 3T. FA of the medulla was significantly higher than that of the cortex in all measurements. Tractography visualized a typical radial diffusion direction in the medulla. Best image quality was achieved with a respiratory triggered protocol with 12 acquisition directions. Measurements with 3 b-values led to decreased ADCs. Acquisition in 12 directions resulted in decreased cortical FA. FA and ADC of breath-hold and free-breathing acquisitions were significantly higher than that of the respiratory-triggered protocol. Intrareader correlation ranged from kappa 0.60 to 0.96. Variance of the respiratory-triggered protocol was smaller than that of breath-hold and free-breathing protocols. Variance was highest for medullary FA in all protocols with reproducibility coefficients ranging from 0.36 to 0.46. Conclusion: Diffusion tensor imaging of the kidney at 3T is feasible and yields significantly higher SNR and CNR. FA and ADCs do not significantly differ from 1.5T. Number of b-values influences ADC-values. Acquisitions in 12 directions provide lower cortical FA-values. We recommend a respiratory-triggered protocol because of improved image quality and reproducibility.
引用
收藏
页码:245 / 254
页数:10
相关论文
共 50 条
  • [31] Diffusion Tensor Imaging of the Uterine Zones Related to the Menstrual Cycle and Menopausal Status at 3 Tesla MRI
    Kilickesmez, Ozgur
    Firat, Zeynep
    Oygen, Aysegul
    Bozkurt, Duygu Kara
    Guzelbey, Tevfik
    Gurses, Bengi
    Tasdelen, Neslihan
    BALKAN MEDICAL JOURNAL, 2016, 33 (06) : 607 - 613
  • [32] Fetal magnetic resonance imaging: jumping from 1.5 to 3 tesla (preliminary experience)
    Victoria, Teresa
    Jaramillo, Diego
    Roberts, Timothy Paul Leslie
    Zarnow, Deborah
    Johnson, Ann Michelle
    Delgado, Jorge
    Rubesova, Erika
    Vossough, Arastoo
    PEDIATRIC RADIOLOGY, 2014, 44 (04) : 376 - 386
  • [33] Reproducibility of tract-specific magnetization transfer and diffusion tensor imaging in the cervical spinal cord at 3 tesla
    Smith, Seth A.
    Jones, Craig K.
    Gifford, Aliya
    Belegu, Visar
    Chodkowski, BettyAnn
    Farrell, Jonathan A. D.
    Landman, Bennett A.
    Reich, Daniel S.
    Calabresi, Peter A.
    McDonald, John W.
    van Zijl, Peter C. M.
    NMR IN BIOMEDICINE, 2010, 23 (02) : 207 - 217
  • [34] Transient global amnesia: 7 Tesla MRI reveals more hippocampal lesions with diffusion restriction compared to 1.5 and 3 Tesla MRI
    Unsgard, Runa Geirmundsdatter
    Doan, Thanh P.
    Nordlid, Knut Kristian
    Kvistad, Kjell Arne
    Goa, Pal Erik
    Berntsen, Erik Magnus
    NEURORADIOLOGY, 2022, 64 (12) : 2217 - 2226
  • [35] Evaluation of Measurement Uncertainties in Human Diffusion Tensor Imaging (DTI)-Derived Parameters and Optimization of Clinical DTI Protocols With a Wild Bootstrap Analysis
    Zhu, Tong
    Liu, Xiaoxu
    Gaugh, Michelle D.
    Connelly, Patrick R.
    Ni, Hongyan
    Ekholm, Sven
    Schifitto, Giovanni
    Zhong, Jianhui
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2009, 29 (02) : 422 - 435
  • [36] Diffusion tensor imaging and tractography of the kidney in children: feasibility and preliminary experience
    Camilo Jaimes
    Kassa Darge
    Dmitry Khrichenko
    Robert H. Carson
    Jeffrey I. Berman
    Pediatric Radiology, 2014, 44 : 30 - 41
  • [37] Higher field reduced FOV diffusion-weighted imaging for abdominal imaging at 5.0 Tesla: image quality evaluation compared with 3.0 Tesla
    Zhang, Yunfei
    Sheng, Ruofan
    Yang, Chun
    Dai, Yongming
    Zeng, Mengsu
    INSIGHTS INTO IMAGING, 2023, 14 (01)
  • [38] Diffusion MRI of normal pancreas: variations and reproducibility of the measurement of apparent diffusion coefficient at 1.5 and 3 Tesla
    Barral, M.
    Soyer, P.
    Ben Hassen, W.
    Gayat, E.
    Aout, M.
    Chiaradia, M.
    Rahmouni, A.
    Luciani, A.
    JOURNAL DE RADIOLOGIE DIAGNOSTIQUE ET INTERVENTIONNELLE, 2013, 94 (04) : 433 - 443
  • [39] Evaluation of Fractional Anisotropy and Apparent Diffusion Coefficient of Broca's Area in Patients with Parkinson's Disease: Quantitative MR Diffusion Tensor Imaging Study at 3 Tesla
    Lee, Jung-Hoon
    Lee, Kyung-Bae
    Kim, Sang-Young
    Lee, Do-Wan
    Choe, Bo-Young
    Choi, Youn-Bong
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2011, 58 (02) : 343 - 348
  • [40] Intraindividual Comparison of Image Quality in MR Urography at 1.5 and 3 Tesla in an Animal Model
    Regier, M.
    Nolte-Ernsting, C.
    Adam, G.
    Kemper, J.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2008, 180 (10): : 915 - 921