Diffusion Tensor Imaging (DTI) of the Kidney at 3 Tesla-Feasibility, Protocol Evaluation and Comparison to 1.5 Tesla

被引:91
|
作者
Notohamiprodjo, Mike [1 ]
Dietrich, Olaf [2 ]
Horger, Wihelm [3 ]
Horng, Annie [1 ]
Helck, Andreas D. [1 ]
Herrmann, Karin A. [1 ]
Reiser, Maximilian F. [1 ]
Glaser, Christian [1 ]
机构
[1] Univ Hosp Munich, Dept Clin Radiol, Inst Clin Radiol, D-81377 Munich, Germany
[2] Univ Hosp Munich, Dept Clin Radiol, Josef Lissner Lab Biomed Imaging, D-81377 Munich, Germany
[3] Siemens Healthcare Sector, Erlangen, Germany
关键词
diffusion tensor imaging; kidney; 3T; DTI; DWI; reproducibility; 3.0; T; MR; ANISOTROPY; COEFFICIENT; NOISE; REPRODUCIBILITY; STRATEGIES; SCHEMES; ABDOMEN; EXPERIENCE;
D O I
10.1097/RLI.0b013e3181d83abc
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to evaluate the feasibility of diffusion tensor imaging of the kidney at a field strength of 3T. We assessed fractional anisotropy (FA) and apparent diffusion coefficients (ADC) of various acquisition protocols and determined the reproducibility of these measurements. FA, ADC, signal-to-noise ratios (SNR), and contrast-to-noise ratios (CNR) were compared with those acquired at 1.5T. Material and Methods: Ten healthy volunteers were examined with a respiratory-triggered echo-planar imaging sequence (TR: 1800 ms, TE: 58 ms, b = 0, 300 s/mm(2)) on a 3-Tesla whole-body MR scanner. Protocol variations included diffusion measurements during free-breathing, in 6 or 12 directions, and an additional b-value of 50 s/mm(2). A breath-hold protocol was also integrated (TR: 820 ms, TE: 58 ms, b = 0, 300 s/mm(2)). Measurements with 2 b-values and 6 diffusion directions were also acquired at 1.5 T. SNR was calculated with the difference-image method. Statistical analysis was performed with Wilcoxon signed-rank tests. Intrareader correlation was assessed with weighted kappa coefficients and reproducibility with the root-mean-square-average and the Bland-Altman-method. Results: At 3T, SNR of cortex and medulla and CNR of cortex/medulla were significantly higher than at 1.5T, leading to improved corticomedullary discrimination. There were no significant FA-and ADC differences with 2 b-values and 6 diffusion directions between measurements at 1.5T and 3T. FA of the medulla was significantly higher than that of the cortex in all measurements. Tractography visualized a typical radial diffusion direction in the medulla. Best image quality was achieved with a respiratory triggered protocol with 12 acquisition directions. Measurements with 3 b-values led to decreased ADCs. Acquisition in 12 directions resulted in decreased cortical FA. FA and ADC of breath-hold and free-breathing acquisitions were significantly higher than that of the respiratory-triggered protocol. Intrareader correlation ranged from kappa 0.60 to 0.96. Variance of the respiratory-triggered protocol was smaller than that of breath-hold and free-breathing protocols. Variance was highest for medullary FA in all protocols with reproducibility coefficients ranging from 0.36 to 0.46. Conclusion: Diffusion tensor imaging of the kidney at 3T is feasible and yields significantly higher SNR and CNR. FA and ADCs do not significantly differ from 1.5T. Number of b-values influences ADC-values. Acquisitions in 12 directions provide lower cortical FA-values. We recommend a respiratory-triggered protocol because of improved image quality and reproducibility.
引用
收藏
页码:245 / 254
页数:10
相关论文
共 50 条
  • [21] Fiber Architecture Mapping of the Renal Medulla Using Respiratory Triggered Diffusion Tensor Imaging at 3 Tesla
    Martirosian, P.
    Schraml, C.
    Schwenzer, N. F.
    Steidle, G.
    Rossi, C.
    Boss, A.
    Kumar, V.
    Erb, M.
    Klose, U.
    Feiweier, T.
    Claussen, C. D.
    Schick, F.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 346 - 349
  • [22] 3 Tesla-Diffusion Tensor Imaging in Autosomal Dominant Polycystic Kidney Disease: The Nephrologist's Point of View
    Lupica, Rosaria
    Mormina, Enricomaria
    Lacquaniti, Antonio
    Trimboli, Domenico
    Bianchimano, Biagio
    Marino, Silvia
    Bramanti, Placido
    Longo, Marcello
    Buemi, Michele
    Granata, Francesca
    NEPHRON, 2016, 134 (02) : 73 - 80
  • [23] Diffusion tensor imaging and diffusion kurtosis imaging of the pancreas - feasibility, robustness and protocol comparison in a healthy population
    Bilreiro, Carlos
    Andrade, Luisa
    Henriques, Rafael
    Loucao, Nuno
    Matos, Celso
    Shemesh, Noam
    ABDOMINAL RADIOLOGY, 2025,
  • [24] Early detection of cervical spondylotic myelopathy using diffusion tensor imaging: Experiences in 1.5-tesla magnetic resonance imaging
    Ahmadli, Uzeyir
    Ulrich, Nils H.
    Yao Yuqiang
    Nanz, Daniel
    Sarnthein, Johannes
    Kollias, Spyros S.
    NEURORADIOLOGY JOURNAL, 2015, 28 (05) : 508 - 514
  • [25] Can 3.0 Tesla diffusion tensor Imaging parameters be prognostic indicators in breast cancer?
    Ozal, Safiye Tokgoz
    Inci, Ercan
    Gemici, Aysegul Akdogan
    Turgut, Hurriyet
    Cikot, Murat
    Karabulut, Mehmet
    CLINICAL IMAGING, 2018, 51 : 240 - 247
  • [26] The comparison of diffusion tensor imaging in human hearts between 1.5 T and 3.0 T
    Li, Xiaodan
    Chen, Rui
    Xu, Xi
    Xiao, Zebin
    Wei, Xiaoyu
    Yang, Yuelong
    Zhang, Zhongping
    Wu, Zhigang
    Zhu, Yanjie
    Liu, Hui
    BMC MEDICAL IMAGING, 2023, 23 (01)
  • [27] Diffusion kurtosis imaging of the liver at 3 Tesla: invivo comparison to standard diffusion-weighted imaging
    Budjan, Johannes
    Sauter, Elke A.
    Zoellner, Frank G.
    Lemke, Andreas
    Wambsganss, Jens
    Schoenberg, Stefan O.
    Attenberger, Ulrike I.
    ACTA RADIOLOGICA, 2018, 59 (01) : 18 - 25
  • [28] Diffusion tensor imaging and tractography of the kidney in children: feasibility and preliminary experience
    Jaimes, Camilo
    Darge, Kassa
    Khrichenko, Dmitry
    Carson, Robert H.
    Berman, Jeffrey I.
    PEDIATRIC RADIOLOGY, 2014, 44 (01) : 30 - 41
  • [29] Quantitative assessment of parallel acquisition techniques in diffusion tensor imaging at 3.0 tesla
    Ardekani, S
    Sinha, U
    PROCEEDINGS OF THE 26TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2004, 26 : 1072 - 1075
  • [30] Diffusion-weighted MR-imaging for the detection of pulmonary nodules at 1.5 Tesla: Intraindividual comparison with multidetector computed tomography
    Regier, Marc
    Schwarz, Dorothee
    Henes, Frank Oliver
    Groth, Michael
    Kooijman, Hendrik
    Begemann, Philipp G. C.
    Adam, Gerhard
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2011, 55 (03) : 266 - 274