High speed, wide velocity dynamic range Doppler optical coherence tomography (Part I): System design, signal processing, and performance

被引:154
作者
Yang, VXD [1 ]
Gordon, ML
Qi, B
Pekar, J
Lo, S
Seng-Yue, E
Mok, A
Wilson, BC
Vitkin, IA
机构
[1] Univ Toronto, Dept Med Biophys, Toronto, ON M5G 2M9, Canada
[2] Univ Toronto, Dept Radiat Oncol, Toronto, ON M5G 2M9, Canada
[3] Univ Hlth Network, Ontario Canc Inst, Toronto, ON M5G 2M9, Canada
[4] Photon Res Ontario, Biophoton Facil, Toronto, ON M5G 2M9, Canada
来源
OPTICS EXPRESS | 2003年 / 11卷 / 07期
关键词
D O I
10.1364/OE.11.000794
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Improvements in real-time Doppler optical coherence tomography (DOCT), acquiring up to 32 frames per second at 250 x 512 pixels per image, are reported using signal processing techniques commonly employed in Doppler ultrasound imaging. The ability to measure a wide range of flow velocities, ranging from less than 20 mum/s to more than 10 cm/s, is demonstrated using an 1.3 mum DOCT system with flow phantoms in steady and pulsatile flow conditions. Based on full implementation of a coherent demodulator, four different modes of flow visualization are demonstrated: color Doppler, velocity variance, Doppler spectrum, and power Doppler. The performance of the former two, which are computationally suitable for real-time imaging, are analyzed in detail under various signal-to-noise and frame-rate conditions. The results serve as a guideline for choosing appropriate imaging parameters for detecting in vivo blood flow. (C) 2003 Optical Society of America.
引用
收藏
页码:794 / 809
页数:16
相关论文
共 13 条
  • [1] Bouma B.E., 2002, Handbook of Optical Coherence Tomography
  • [2] Stable carrier generation and phase-resolved digital data processing in optical coherence tomography
    de Boer, JF
    Saxer, CE
    Nelson, JS
    [J]. APPLIED OPTICS, 2001, 40 (31) : 5787 - 5790
  • [3] Real-time phase-resolved optical coherence tomography and optical Doppler tomography
    Ding, ZH
    Zhao, YH
    Ren, HW
    Nelson, JS
    Chen, ZP
    [J]. OPTICS EXPRESS, 2002, 10 (05): : 236 - 245
  • [4] FOX SIF, 1987, HUMAN PHYSL
  • [5] OPTICAL COHERENCE TOMOGRAPHY
    HUANG, D
    SWANSON, EA
    LIN, CP
    SCHUMAN, JS
    STINSON, WG
    CHANG, W
    HEE, MR
    FLOTTE, T
    GREGORY, K
    PULIAFITO, CA
    FUJIMOTO, JG
    [J]. SCIENCE, 1991, 254 (5035) : 1178 - 1181
  • [6] Jensen J. A., 1996, Estimation of Blood Velocities using Ultrasound-A Signal Processing Approach
  • [7] Video-rate three-dimensional optical coherence tomography
    Laubscher, M
    Ducros, M
    Karamata, B
    Lasser, T
    Salathé, R
    [J]. OPTICS EXPRESS, 2002, 10 (09): : 429 - 435
  • [8] In vivo video rate optical coherence tomography
    Rollins, AM
    Kulkarni, MD
    Yazdanfar, S
    Ung-arunyawee, R
    Izatt, JA
    [J]. OPTICS EXPRESS, 1998, 3 (06): : 219 - 229
  • [9] High-speed phase- and group-delay scanning with a grating-based phase control delay line
    Tearney, GJ
    Bouma, BE
    Fujimoto, JG
    [J]. OPTICS LETTERS, 1997, 22 (23) : 1811 - 1813
  • [10] Real-time, high velocity-resolution color Doppler optical coherence tomography
    Westphal, V
    Yazdanfar, S
    Rollins, AM
    Izatt, JA
    [J]. OPTICS LETTERS, 2002, 27 (01) : 34 - 36