Measurement of petroleum fuel contamination in water by solid-phase microextraction with direct Raman spectroscopic detection

被引:19
作者
Jager, MJ [1 ]
McClintic, DP [1 ]
Tilotta, DC [1 ]
机构
[1] Univ N Dakota, Dept Chem, Grand Forks, ND 58201 USA
关键词
Raman spectroscopy; solid-phase microextraction; environmental analysis; petroleum hydrocarbons; petroleum fuels;
D O I
10.1366/0003702001948637
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A method is described for determining petroleum fuel contamination in water based on solid-phase microextraction and Raman spectroscopy (SPME/Raman). In this method, contaminants are extracted from aqueous solutions into a solid phase and then detected directly by using spontaneous Raman spectroscopy. The solid phase consists of a small volume (similar to 20 muL) of poly(dimethylsiloxane) that has an optical window in the 830-1600 cm(-1) Raman shift region. This region is suitable for determining both aromatic and aliphatic fuel components. Four "real world" fuels-aviation gasoline, unleaded gasoline, jet fuel A, and #1 fuel oil-were used to test this new method. As shown, the SPME/Raman determinations performed in the,visible spectral region were not adversely affected from the fluorescence of the marker dyes in these fuels. The linear dynamic ranges for the determinations exceeded the water solubility limits and spanned 1-3 orders of magnitude. The limits of detection with the use of the most sensitive Raman bands In each fuel were in the 8-12 ppm range with the exception of #1 fuel oil (25 ppm), and relative standard deviations typically were 2-10%. Finally, the SPME/Raman method was applied to the determination of total petroleum hydrocarbons in natural water matrices. With the use of hexane and undecane as standards, the results of the determinations compared favorably to the standard methods of purge-and-trap gas chromatography and liquid-liquid extraction followed by infrared detection.
引用
收藏
页码:1617 / 1623
页数:7
相关论文
共 13 条
[1]   Solid phase microextraction for quantitative analysis in nonequilibrium situations [J].
Ai, J .
ANALYTICAL CHEMISTRY, 1997, 69 (06) :1230-1236
[2]  
*AM PETR I, 1990, 911 API
[3]   FEASIBILITY STUDIES FOR THE DETECTION OF ORGANIC-SURFACE AND SUBSURFACE WATER CONTAMINANTS BY SURFACE-ENHANCED RAMAN-SPECTROSCOPY ON SILVER ELECTRODES [J].
CARRABBA, MM ;
EDMONDS, RB ;
RAUH, RD .
ANALYTICAL CHEMISTRY, 1987, 59 (21) :2559-2563
[4]   OCTADECYLTHIOL-MODIFIED SURFACE-ENHANCED RAMAN-SPECTROSCOPY SUBSTRATES - A NEW METHOD FOR THE DETECTION OF AROMATIC-COMPOUNDS [J].
CARRON, K ;
PEITERSEN, L ;
LEWIS, M .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1992, 26 (10) :1950-1954
[5]  
DOLLISH FR, 1974, CHARACTERISTIC RAMAN, P3
[6]  
Franson M.A., 1995, STANDARD METHODS EXA
[7]   DYNAMICS OF ORGANIC-COMPOUND EXTRACTION FROM WATER USING LIQUID-COATED FUSED-SILICA FIBERS [J].
LOUCH, D ;
MOTLAGH, S ;
PAWLISZYN, J .
ANALYTICAL CHEMISTRY, 1992, 64 (10) :1187-1199
[8]  
PAWILISYN J, 1997, SOLID PHASE MICROEXT
[9]   Partition infrared method for total gasoline range organics in water based on solid phase microextraction [J].
Stahl, DC ;
Tilotta, DC .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1999, 33 (05) :814-819
[10]   Determination of BTEX contaminants in water via a long-pathlength fiber-optic Raman 'dip-stick' [J].
Stellman, CM ;
Ewing, KJ ;
Bucholtz, F ;
Aggarwal, ID .
SENSORS AND ACTUATORS B-CHEMICAL, 1998, 53 (03) :173-178